Literature DB >> 27419615

Design and verification of a shape memory polymer peripheral occlusion device.

Todd L Landsman1, Ruth L Bush2, Alan Glowczwski3, John Horn1, Staci L Jessen1, Ethan Ungchusri2, Katelin Diguette1, Harrison R Smith1, Sayyeda M Hasan1, Daniel Nash4, Fred J Clubb5, Duncan J Maitland6.   

Abstract

Shape memory polymer foams have been previously investigated for their safety and efficacy in treating a porcine aneurysm model. Their biocompatibility, rapid thrombus formation, and ability for endovascular catheter-based delivery to a variety of vascular beds makes these foams ideal candidates for use in numerous embolic applications, particularly within the peripheral vasculature. This study sought to investigate the material properties, safety, and efficacy of a shape memory polymer peripheral embolization device in vitro. The material characteristics of the device were analyzed to show tunability of the glass transition temperature (Tg) and the expansion rate of the polymer to ensure adequate time to deliver the device through a catheter prior to excessive foam expansion. Mechanical analysis and flow migration studies were performed to ensure minimal risk of vessel perforation and undesired thromboembolism upon device deployment. The efficacy of the device was verified by performing blood flow studies that established affinity for thrombus formation and blood penetration throughout the foam and by delivery of the device in an ultrasound phantom that demonstrated flow stagnation and diversion of flow to collateral pathways.
Copyright © 2016 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Embolization; Pelvic congestion syndrome; Shape memory polymer; Varicocele; Varicose veins

Mesh:

Substances:

Year:  2016        PMID: 27419615      PMCID: PMC5508979          DOI: 10.1016/j.jmbbm.2016.06.019

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  34 in total

1.  Effect of 3.2% vs 3.8% sodium citrate concentration on routine coagulation testing.

Authors:  D M Adcock; D C Kressin; R A Marlar
Journal:  Am J Clin Pathol       Date:  1997-01       Impact factor: 2.493

2.  In vivo response to an implanted shape memory polyurethane foam in a porcine aneurysm model.

Authors:  Jennifer N Rodriguez; Fred J Clubb; Thomas S Wilson; Matthew W Miller; Theresa W Fossum; Jonathan Hartman; Egemen Tuzun; Pooja Singhal; Duncan J Maitland
Journal:  J Biomed Mater Res A       Date:  2013-05-23       Impact factor: 4.396

3.  The effect of moisture absorption on the physical properties of polyurethane shape memory polymer foams.

Authors:  Ya-Jen Yu; Keith Hearon; Thomas S Wilson; Duncan J Maitland
Journal:  Smart Mater Struct       Date:  2011-08-01       Impact factor: 3.585

Review 4.  The epidemiology of chronic venous insufficiency and varicose veins.

Authors:  Jennifer L Beebe-Dimmer; John R Pfeifer; Jennifer S Engle; David Schottenfeld
Journal:  Ann Epidemiol       Date:  2005-03       Impact factor: 3.797

5.  Gross and microscopic histopathological findings in aneurysms of the human brain treated with Guglielmi detachable coils.

Authors:  G Bavinzski; V Talazoglu; M Killer; B Richling; A Gruber; C E Gross; H Plenk
Journal:  J Neurosurg       Date:  1999-08       Impact factor: 5.115

6.  Virtual treatment of basilar aneurysms using shape memory polymer foam.

Authors:  J M Ortega; J Hartman; J N Rodriguez; D J Maitland
Journal:  Ann Biomed Eng       Date:  2013-01-18       Impact factor: 3.934

7.  Percutaneous embolization of a postnephrectomy arteriovenous fistula with intervening pseudoaneurysm using the Amplatzer vascular plug 2.

Authors:  Ahmed Kamel Abdel-Aal; Ahmed Elsabbagh; Hesham Soliman; Maysoon Hamed; Edgar Underwood; Souheil Saddekni
Journal:  Vasc Endovascular Surg       Date:  2014-12-08       Impact factor: 1.089

8.  The endovascular occlusion system for safe and immediate peripheral vessel occlusion during vascular interventions.

Authors:  Maximilian Y Emmert; Anthony Venbrux; Leon Rudakov; Nikola Cesarovic; Martin G Radvany; Philippe Gailloud; Volkmar Falk; Andre Plass
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-07-17

9.  Mechanical properties of completely autologous human tissue engineered blood vessels compared to human saphenous vein and mammary artery.

Authors:  Gerhardt Konig; Todd N McAllister; Nathalie Dusserre; Sergio A Garrido; Corey Iyican; Alicia Marini; Alex Fiorillo; Hernan Avila; Wojciech Wystrychowski; Krzysztof Zagalski; Marcin Maruszewski; Alyce Linthurst Jones; Lech Cierpka; Luis M de la Fuente; Nicolas L'Heureux
Journal:  Biomaterials       Date:  2008-12-25       Impact factor: 12.479

10.  Occlusive thrombus formation on indwelling catheters: in vitro investigation and computational analysis.

Authors:  P Friedrich; A J Reininger
Journal:  Thromb Haemost       Date:  1995-01       Impact factor: 5.249

View more
  12 in total

1.  Microscopic Assessment of Healing and Effectiveness of a Foam-Based Peripheral Occlusion Device.

Authors:  Staci L Jessen; Molly C Friedemann; Anne-Marie Ginn-Hedman; Lance M Graul; Steven Jokerst; Cedric B Robinson; Todd L Landsman; Fred J Clubb; Duncan J Maitland
Journal:  ACS Biomater Sci Eng       Date:  2019-09-23

Review 2.  Advances in Biomaterials and Technologies for Vascular Embolization.

Authors:  Jingjie Hu; Hassan Albadawi; Brian W Chong; Amy R Deipolyi; Rahul A Sheth; Ali Khademhosseini; Rahmi Oklu
Journal:  Adv Mater       Date:  2019-06-06       Impact factor: 30.849

3.  Particulate Release From Nanoparticle-Loaded Shape Memory Polymer Foams.

Authors:  Adam L Nathan; Grace K Fletcher; Mary Beth B Monroe; Wonjun Hwang; Scott M Herting; Sayyeda M Hasan; Brandis K Keller; Duncan J Maitland
Journal:  J Med Device       Date:  2017-01-16       Impact factor: 0.582

4.  Biodegradable shape memory polymer foams with appropriate thermal properties for hemostatic applications.

Authors:  Lindy K Jang; Grace K Fletcher; Mary Beth B Monroe; Duncan J Maitland
Journal:  J Biomed Mater Res A       Date:  2020-02-21       Impact factor: 4.396

5.  Micro-CT and histopathology methods to assess host response of aneurysms treated with shape memory polymer foam-coated coils versus bare metal coil occlusion devices.

Authors:  Staci L Jessen; Molly C Friedemann; Annmarie E Mullen; Anne-Marie Ginn-Hedman; Scott M Herting; Duncan J Maitland; Fred J Clubb
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2020-01-21       Impact factor: 3.368

6.  Thermomechanical data of polyurethane shape memory polymer: Considering varying compositions.

Authors:  Hailey Fisher; Payton Woolard; Colton Ross; Robert Kunkel; Bradley N Bohnstedt; Yingtao Liu; Chung-Hao Lee
Journal:  Data Brief       Date:  2020-09-09

7.  Chemical Modifications of Porous Shape Memory Polymers for Enhanced X-ray and MRI Visibility.

Authors:  Grace K Fletcher; Landon D Nash; Lance M Graul; Lindy K Jang; Scott M Herting; Matthew D Wilcox; Tyler J Touchet; Ana Katarina Sweatt; Mary P McDougall; Steven M Wright; Duncan J Maitland
Journal:  Molecules       Date:  2020-10-13       Impact factor: 4.411

8.  Embolization of the false lumen using IMPEDE-FX embolization plugs as part of treatment of an infrarenal post-dissection aneurysm: a case report.

Authors:  Anne-Jet S Jansen; Paul M van Schaik; Jasper M Martens; Michel M P J Reijnen
Journal:  CVIR Endovasc       Date:  2020-12-04

9.  Clinical experience with a shape memory polymer peripheral vascular embolisation plug: a case series.

Authors:  Robert A Morgan; Ian Loftus; Lakshmi Ratnam; Raj Das; Leto Mailli; Mohamad S Hamady; Kyriakos Lobotesis
Journal:  CVIR Endovasc       Date:  2021-03-09

10.  Increased X-ray Visualization of Shape Memory Polymer Foams by Chemical Incorporation of Iodine Motifs.

Authors:  Landon D Nash; Mary Beth Browning Monroe; Yong-Hong Ding; Kendal P Ezell; Anthony J Boyle; Ramanathan Kadirvel; David F Kallmes; Duncan J Maitland
Journal:  Polymers (Basel)       Date:  2017-08-20       Impact factor: 4.329

View more