Designing the phase grating for Talbot-Lau phase-contrast imaging systems: a simulation and experiment study.

Jens Rieger, Pascal Meyer, Georg Pelzer, Thomas Weber, Thilo Michel, Jürgen Mohr, Gisela Anton.   

Abstract

The performance of a Talbot-Lau interferometer depends to a great extent on its visibility. This means, to obtain high quality phase-contrast and dark-field images a high visibility is mandatory. Several parameters influence the visibility of such a system, like for example the x-ray spectrum, the inter-grating distances or the parameters of the three gratings. In this multidimensional space, wave field simulations help to find the optimal combination of the grating specifications to construct a setup with a high visibility while retaining a fixed angular sensitivity. In this work we specifically analyzed the influence of the G1 grating duty cycle in simulations and experiments. We show that there is a lot of room for improvement by varying the duty cycle of the phase-shifting grating G1. As a result, by employing a third-integer duty cycle we can increase the visibility to up to 53 % in a laboratory setup with a polychromatic spectrum. The achieved visibility is more than two times higher compared to the result with a standard-type setup. This visibility gain allows a dose reduction by a factor of 5 preserving the same image quality.

Year:  2016        PMID: 27410353     DOI: 10.1364/OE.24.013357

Source DB:  PubMed          Journal:  Opt Express        ISSN: 1094-4087            Impact factor:   3.894


  2 in total

1.  Simulation study on X-ray phase contrast imaging with dual-phase gratings.

Authors:  Johannes Bopp; Veronika Ludwig; Maria Seifert; Georg Pelzer; Andreas Maier; Gisela Anton; Christian Riess
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-10-22       Impact factor: 2.924

2.  Improved reconstruction of phase-stepping data for Talbot-Lau x-ray imaging.

Authors:  Sebastian Kaeppler; Jens Rieger; Georg Pelzer; Florian Horn; Thilo Michel; Andreas Maier; Gisela Anton; Christian Riess
Journal:  J Med Imaging (Bellingham)       Date:  2017-09-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.