Noninvasive hemoglobin measurement in pediatric trauma patients.

Mark Leo Ryan1, Angela C Maxwell, Lisa Manning, Jonathan D Jacobs, Marielena Bachier-Rodriguez, Alexander Feliz, Regan F Williams.   

Abstract

INTRODUCTION: Hemorrhage is a major cause of preventable death secondary to traumatic injury. Diagnosis often requires multiple blood draws, which are psychologically stressful in pediatric patients. The Pronto device is a pulse co-oximeter that measures the total hemoglobin level using multiple wavelengths of light. The purpose of this study was to evaluate the accuracy of the noninvasive hemoglobin measurements relative to current invasive and point of care testing methods in pediatric trauma patients.
METHODS: We performed a prospective observational trial involving patients younger than 17 years presenting to a Level I pediatric trauma center. Following admission, blood was sampled from each patient for testing using an i-Stat device (point-of-care hemoglobin) and a complete blood count within our core laboratory (invasive hemoglobin). Noninvasive hemoglobin analysis was performed within 15 minutes of phlebotomy. Data were evaluated using Spearman correlation and Bland-Altman analysis.
RESULTS: Over 2 years, 114 patients had attempted noninvasive hemoglobin measurements, with a success rate of 89%. Mean ± SD age was 9.2 ± 5.1 years. Ninety percent of admissions were for blunt injury, 3% penetrating, 5% near drowning, and 1% burns. Mean invasive hemoglobin was 12.6 ± 1.9 g/dL, mean point-of-care hemoglobin was 12.2 ± 2.0 g/dL, and mean noninvasive hemoglobin was 12.3 ± 1.6 g/dL. Noninvasive hemoglobin values were strongly correlated with both invasive and point of care measurements (R = 0.672 and R = 0.645, respectively; p < 0.001). Bland-Altman analysis comparing noninvasive to point-of-care and invasive hemoglobin levels resulted in an estimated bias of -0.39 and -0.49, respectively.
CONCLUSION: Noninvasive hemoglobin values had excellent correlation with both invasive and point-of-care hemoglobin measurements, although the device was not successful for all patients. Given the rapid availability of results and the lack of requirement of venipuncture, noninvasive hemoglobin monitoring may be a valuable adjunct in the initial evaluation and monitoring of pediatric trauma patients. LEVEL OF EVIDENCE: Diagnostic test study, level II.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27389126     DOI: 10.1097/TA.0000000000001160

Source DB:  PubMed          Journal:  J Trauma Acute Care Surg        ISSN: 2163-0755            Impact factor:   3.313


  2 in total

1.  The European guideline on management of major bleeding and coagulopathy following trauma: fifth edition.

Authors:  Donat R Spahn; Bertil Bouillon; Vladimir Cerny; Jacques Duranteau; Daniela Filipescu; Beverley J Hunt; Radko Komadina; Marc Maegele; Giuseppe Nardi; Louis Riddez; Charles-Marc Samama; Jean-Louis Vincent; Rolf Rossaint
Journal:  Crit Care       Date:  2019-03-27       Impact factor: 9.097

2.  Reliability of transcutaneous hemoglobin measurements in an outpatient pediatric clinic.

Authors:  Shae Margulies; Molly Posa; Melissa Fitzgerald; Stephanie Filipp; Diane Howell; Puneet Tung; Maria Kelly
Journal:  Postgrad Med       Date:  2021-12-27       Impact factor: 3.840

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.