Diaphragmatic Myopotential Oversensing Caused by Change in Implantable Cardioverter Defibrillator Sensing Bandpass Filter.

Sylvain Ploux1, Charles D Swerdlow2, Romain Eschalier3, Benjamin Monteil1, Sana Ouali4, Michel Haïssaguerre1, Pierre Bordachar1.   

Abstract

Diaphragmatic myopotential oversensing (DMO) causes inhibition of pacing and inappropriate detection of ventricular fibrillation in implantable cardioverter defibrillators (ICDs). It occurs almost exclusively with integrated bipolar leads and is extremely rare with dedicated bipolar leads. If DMO cannot be corrected by reducing programmed sensitivity, ventricular lead revision is often required. The new Low Frequency Attenuation (LFA) filter in St. Jude Medical ICDs (St. Jude Medical, Sylmar, CA, USA) alters the sensing bandpass to reduce T-wave oversensing. This paper aims to present the LFA filter as a reversible cause of DMO. Unnecessary lead revision can be avoided by the simple programming solution of deactivating this LFA filter. ©2016 Wiley Periodicals, Inc.

Entities:  

Keywords:  defibrillation - ICD; electrocardiogram; filter; oversensing

Year:  2016        PMID: 27378562     DOI: 10.1111/pace.12837

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

Review 1.  Sensing and detection in Medtronic implantable cardioverter defibrillators.

Authors:  Mark L Brown; Charles D Swerdlow
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-09-08

2.  SMART pass will prevent inappropriate operation of S-ICD.

Authors:  Motomi Tachibana; Nobuhiro Nishii; Kimikazu Banba; Shinpei Fujita; Etsuko Ikeda; Keisuke Okawa; Hiroshi Morita; Hiroshi Ito
Journal:  J Arrhythm       Date:  2018-11-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.