Sylvain Ploux1, Charles D Swerdlow2, Romain Eschalier3, Benjamin Monteil1, Sana Ouali4, Michel Haïssaguerre1, Pierre Bordachar1.
Abstract
Diaphragmatic myopotential oversensing (DMO) causes inhibition of pacing and inappropriate detection of ventricular fibrillation in implantable cardioverter defibrillators (ICDs). It occurs almost exclusively with integrated bipolar leads and is extremely rare with dedicated bipolar leads. If DMO cannot be corrected by reducing programmed sensitivity, ventricular lead revision is often required. The new Low Frequency Attenuation (LFA) filter in St. Jude Medical ICDs (St. Jude Medical, Sylmar, CA, USA) alters the sensing bandpass to reduce T-wave oversensing. This paper aims to present the LFA filter as a reversible cause of DMO. Unnecessary lead revision can be avoided by the simple programming solution of deactivating this LFA filter. ©2016 Wiley Periodicals, Inc.Entities:
Keywords: defibrillation - ICD; electrocardiogram; filter; oversensing
Year: 2016 PMID: 27378562 DOI: 10.1111/pace.12837
Source DB: PubMed Journal: Pacing Clin Electrophysiol ISSN: 0147-8389 Impact factor: 1.976