Muhammet Gozdemir1, Ruveyda Irem Demircioglu2, Safinaz Karabayirli3, Huseyin Sert4, Bunyamin Muslu5, Burhanettin Usta6, Ummugulsum Yazici7.
Abstract
OBJECTIVES: Local anesthetic infiltration is also a process of a painful process itself. INJEX™ technology, known as "Needle-free" drug delivery system, was designed for reducing the pain associated with cutaneous procedures. We conducted a prospective, randomized trial to evaluate the application of lidocaine with INJEX™ system and 27-gauge needle.Entities:
Keywords: Injection; Local Anesthetic delivery; Needle Free; Technology
Year: 2016 PMID: 27375728 PMCID: PMC4928437 DOI: 10.12669/pjms.323.9174
Source DB: PubMed Journal: Pak J Med Sci ISSN: 1681-715X Impact factor: 1.088
Fig.1Needle-free injection system; INJEX™ model 210.
Demographic Data.
| Age (yr) | 31±5 | 30±4 | 0.311 |
| Height (cm) | 164±5 | 167±6 | 0.17 |
| Weight (kg) | 78±9 | 76±9 | 0.376 |
| Applied interval (L3-4/L4-5) | 18/12 | 23/7 | 0.165 |
| 1 | 26 | 27 | 0.431 |
| 2 | 4 | 2 | |
| >2 | 0 | 1 |
Values are means±SD and number of patients.
ENI: Epidural needle insertion, No significant differences were observed.
Patient’s VAS values and local injection site reaction.
| LA VAS values when implementing | 0 [0-3] | 2 [0-4] | 0.001 |
| EN VAS values when implementing | 1 [0-4] | 2 [0-5] | 0.074 |
| Bleeding | 9 | 7 | 0.559 |
| Erythema | 6 | 3 | 0.217 |
| Edema | 5 | 0 | 0.001 |
Values are median (range) (%25-%75) and number of patients. LA: Local Anesthesi, EN: Epidural Needle, VAS: Visual Analogue Scale
p<0 05 compared to the two groups.
Fig.2Pain ratings at the time of local anesthetic injection.
Patient Satisfaction.
| Excellent | 2 | 0 |
| Good | 18 | 16 |
| Fair | 10 | 11 |
| Poor | 0 | 3 |
| Preferred rate again (Yes/No) | 28/2 | 25/5 |
Values are number of patients. No significant differences were observed.
Fig.3Pain ratings during epidural needle insertion.