| Literature DB >> 27374705 |
Yoon Young Cho1, Sejong Chun2, Soo Youn Lee2, Jae Hoon Chung1, Hyung Doo Park3, Sun Wook Kim4.
Abstract
BACKGROUND: Measurement of postoperative serum thyroglobulin (Tg) is important for detecting persistent or recurrent differentiated thyroid cancer. We evaluated the analytic performance of the DxI 800 assay (Beckman Coulter, USA) for serum Tg and anti-thyroglobulin antibodies (TgAbs) in comparison with that of the GAMMA-10 assay (Shinjin Medics Inc., Korea) for serum Tg and RIA-MAT 280 assay (Stratec, Germany) for TgAb.Entities:
Keywords: DxI 800; Evaluation; Immunochemiluminometric assay; Performance; Thyroglobulin
Mesh:
Substances:
Year: 2016 PMID: 27374705 PMCID: PMC4940483 DOI: 10.3343/alm.2016.36.5.413
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
Precision profile of the DxI 800 assay for serum thyroglobulin and anti-thyroglobulin antibody
| Analyte | Calibration level | Mean | SD | CV (%) | |||
|---|---|---|---|---|---|---|---|
| Within run | Between run | Between day | Total | ||||
| Tg (ng/mL) | Pooled serum | 170.6 | 6.7 | 2.9 | 1.7 | 1.9 | 3.9 |
| Low level | 31.7 | 1.4 | 3.9 | 0 | 2.2 | 4.4 | |
| Middle level | 120.6 | 5.8 | 4.2 | 1.0 | 2.1 | 4.8 | |
| High level | 380.0 | 21.4 | 4.0 | 3.9 | 1.0 | 5.6 | |
| TgAb (IU/mL) | Pooled serum | 80.4 | 4.3 | 3.5 | 2.5 | 3.2 | 5.3 |
| Low level | 4.2 | 0.3 | 6.2 | 3.1 | 0 | 6.9 | |
| High level | 475.5 | 32.6 | 4.5 | 5.1 | 1.2 | 6.9 | |
For serum Tg, control materials were manufactured with three concentrations: 32.5 ng/mL (low level calibrator), 250 ng/mL (middle level calibrator), and 400 ng/mL (high level calibrator). For serum TgAb, control materials were manufactured with two concentrations: 4.5 IU/mL (low level calibrator) and 500 IU/mL (high level calibrator).
Abbreviations: Tg, thyroglobulin; TgAb, anti-thyroglobulin antibody.
Fig. 1Linearity profile of the DxI 800 assay for serum thyroglobulin (Tg) and anti-thyroglobulin antibody (TgAb). For the evaluation of linearity, five concentrations of control materials were used for both Tg and TgAb. The intended concentrations for serum Tg were 0, 94, 188, 282, 376, and 470 ng/mL (A), and 0, 480, 960, 1,440, 1,920, and 2,400 IU/mL for serum TgAb (B).
Fig. 2Correlation of serum thyroglobulin (A) and anti-thyroglobulin antibody (B) concentrations measured by the two assays (results within analytical measurement ranges).
Agreement of serum thyroglobulin and anti-thyroglobulin antibody between the two assays
| Tg (ng/mL)* | GAMMA-10 | Total | |||
|---|---|---|---|---|---|
| <1 | 1-10 | >10 | |||
| DxI 800 | <1 | 37 | 5 | 0 | 42 |
| 1-10 | 0 | 37 | 2 | 39 | |
| >10 | 0 | 1 | 17 | 18 | |
| Total | 37 | 43 | 19 | 99 | |
Values are presented with numbers of subjects.
*Weighted Cohen's Kappa statistics was 0.884 for serum Tg; †Cohen's Kappa statistics was 0.231 for TgAb, and the agreement rate was 68.7%; ‡The cut-offs for positivity of serum TgAb were 4.0 IU/mL for the DxI 800 and 60.0 IU/mL for the RIA-MAT 280.
Abbreviations: Tg, thyroglobulin; TgAb, anti-thyroglobulin antibody.
Functional sensitivity of low levels of serum thyroglobulin manufactured from calibrators and samples from healthy subjects
| Target (ng/mL) | Calibrators | Healthy subject samples | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 0.2 | 0.1 | 0.075 | 0.05 | 2 | 1.5 | 1 | 0.75 | 0.5 | 0.25 | 0.125 | 0.0625 | 0.0313 | |
| Mean (ng/mL) | 0.195 | 0.100 | 0.072 | 0.039 | 2.065 | 1.489 | 0.878 | 0.773 | 0.504 | 0.209 | 0.114 | 0.062 | 0.040 |
| SD (ng/mL) | 0.013 | 0.008 | 0.007 | 0.011 | 0.078 | 0.071 | 0.041 | 0.032 | 0.034 | 0.012 | 0.008 | 0.009 | 0.008 |
| CV (%) | 6.5 | 7.6 | 9.3 | 28.5 | 3.8 | 4.7 | 4.6 | 4.1 | 6.7 | 5.6 | 6.9 | 14.5 | 21.3 |