ECG predictors of T wave oversensing in subcutaneous implantable cardioverter defibrillators.

D G Wilson1, G Leventigiannis2, C Barr2, J M Morgan3.   

Abstract

BACKGROUND: T wave oversensing (TWOS) is the commonest cause of inappropriate shocks in subcutaneous implantable cardioverter defibrillators (S-ICDs). We hypothesise that predictors of TWOS can be derived from surface ECG parameters.
METHODS: In a cohort of SICD recipients in two UK centres, all patients who had TWOS (study group) were compared to all those who had not (control group). The pre-implant screen was scanned and the R wave, T wave amplitudes, QRS interval, time to peak T wave, QT interval and R:T ratio was measured using digital callipers. Logistic regression was performed to identify ECG predictors of TWOS.
RESULTS: One hundred one patients were studied. Six (5.9%) had TWOS. The mean age of the population was 58.6±18years and the median follow-up was 19.5months. By univariate analysis, the predictors of TWOS are QRS duration (140.7±28.7 vs. 105.9±24.6, P=0.007), time to peak T wave (corrected for heart rate, pTc) (403.9±22.6 vs. 347.8±41.4, P=0.006), QTc interval (500.4±41.2 vs. 446.8±49.7, P=0.021), and R:T ratio (3.5±1.1 vs. 9.5±13.2, P=0.034). By multivariate analysis, time to pTc is the most predictive of TWOS. A time to pTc of 390ms cut-off point provided a sensitivity 38.5%, a specificity of 98.9%, a positive predictive value for TWOS of 83.3%, and a negative predictive value of 91.6% (AUC=0.687).
CONCLUSION: In this study, time to pTc is the most powerful ECG predictor of TWOS.
Copyright © 2016. Published by Elsevier Ireland Ltd.

Entities:  

Keywords:  Electrocardiogram; Inappropriate shock; Predictors; Subcutaneous implantable cardioverter defibrillator; T wave oversensing

Mesh:

Year:  2016        PMID: 27372039     DOI: 10.1016/j.ijcard.2016.06.128

Source DB:  PubMed          Journal:  Int J Cardiol        ISSN: 0167-5273            Impact factor:   4.164


  2 in total

1.  T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing.

Authors:  Ya-Xun Sun; Jing Gao; Chen-Yang Jiang; Yu-Mei Xue; Yi-Zhou Xu; Gang Liu; Ji-Hong Guo; Xia Sheng; Yang Ye; Hong He; Yun-Tao Zhao; Hector Barajas-Martinez; Guo-Sheng Fu; Dan Hu
Journal:  Front Physiol       Date:  2017-09-01       Impact factor: 4.566

Review 2.  State-of-the-art consensus on non-transvenous implantable cardioverter-defibrillator therapy.

Authors:  Christoph Schukro; David Santer; Günther Prenner; Markus Stühlinger; Martin Martinek; Alexander Teubl; Deddo Moertl; Stefan Schwarz; Michael Nürnberg; Lukas Fiedler; Robert Hatala; Cesar Khazen
Journal:  Clin Cardiol       Date:  2020-08-14       Impact factor: 2.882

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.