Alexander Rodriguez-Palacios1, Henry R Stämpfli2, Yung-Fu Chang3.
Abstract
Aberrant false-positive reactions in negative-controls during ELISA testing for Clostridium difficile indicated the potential for false-diagnoses. Experiments with 96-well products showed a maximum peak of false-positive immunoassay reactions with the provided negative-control reagents after 5 refrigeration-to-room temperature cycles (P < 0.001), decreasing thereafter with additional refrigeration cycles. Because repetitive refrigeration causes a peak of false-positives, the use of single negative-controls per ELISA run might be insufficient to monitor aberrant preanalytical false-positives if immunoassays are subject to repetitive refrigeration.Entities:
Year: 2014 PMID: 27350964 PMCID: PMC4897431 DOI: 10.1155/2014/128120
Source DB: PubMed Journal: Int Sch Res Notices ISSN: 2356-7872
Figure 1Refrigeration cycles increased the OD450 and false-positivity (convex correlation) in the ELISA wells of a valid nonexpired ELISA kit. The product insert of this ELISA kit instructed to bring all reagents to room temperature each time before testing. Positive-controls, not shown, performed as expected. The expired products yielded no positive or negative reactions as expected.