Intervendor consistency and reproducibility of left ventricular 2D global and regional strain with two different high-end ultrasound systems.

Kenji Shiino1, Akira Yamada1, Matthew Ischenko2, Bijoy K Khandheria3, Mahala Hudaverdi2, Vicki Speranza2, Mary Harten2, Anthony Benjamin2, Christian R Hamilton-Craig2,4, David G Platts2, Darryl J Burstow2, Gregory M Scalia2, Jonathan Chan1,2.   

Abstract

AIMS: We aimed to assess intervendor agreement of global (GLS) and regional longitudinal strain by vendor-specific software after EACVI/ASE Industry Task Force Standardization Initiatives for Deformation Imaging. METHODS AND
RESULTS: Fifty-five patients underwent prospective dataset acquisitions on the same day by the same operator using two commercially available cardiac ultrasound systems (GE Vivid E9 and Philips iE33). GLS and regional peak longitudinal strain were analyzed offline using corresponding vendor-specific software (EchoPAC BT13 and QLAB version 10.3). Absolute mean GLS measurements were similar between the two vendors (GE -17.5 ± 5.2% vs. Philips -18.9 ± 5.1%, P = 0.15). There was excellent intervendor correlation of GLS by the same observer [r = 0.94, P < 0.0001; bias -1.3%, 95% CI limits of agreement (LOA) -4.8 to 2.2%). Intervendor comparison for regional longitudinal strain by coronary artery territories distribution were: LAD: r = 0.85, P < 0.0001; bias 0.5%, LOA -5.3 to 6.4%; RCA: r = 0.88, P < 0.0001; bias -2.4%, LOA -8.6 to 3.7%; LCX: r = 0.76, P < 0.0001; bias -5.3%, LOA -10.6 to 2.0%. Intervendor comparison for regional longitudinal strain by LV levels were: basal: r = 0.86, P < 0.0001; bias -3.6%, LOA -9.9 to 2.0%; mid: r = 0.90, P < 0.0001; bias -2.6%, LOA -7.8 to 2.6%; apical: r = 0.74; P < 0.0001; bias -1.3%, LOA -9.4 to 6.8%.
CONCLUSIONS: Intervendor agreement in GLS and regional strain measurements have significantly improved after the EACVI/ASE Task Force Strain Standardization Initiatives. However, significant wide LOA still exist, especially for regional strain measurements, which remains relevant when considering vendor-specific software for serial measurements. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author 2016. For permissions please email: journals.permissions@oup.com.

Entities:  

Keywords:  speckle-tracking echocardiography; strain

Mesh:

Year:  2017        PMID: 27330151     DOI: 10.1093/ehjci/jew120

Source DB:  PubMed          Journal:  Eur Heart J Cardiovasc Imaging        ISSN: 2047-2404            Impact factor:   6.875


  10 in total

1.  Quantitative evaluation of longitudinal strain in layer-specific myocardium in patients with preeclampsia.

Authors:  Juan Cong; Yong Lee; Xiuxiu Fu; Zhibin Wang; Wugang Wang; Junfang Lee
Journal:  Int J Cardiovasc Imaging       Date:  2017-08-03       Impact factor: 2.357

2.  Two-Dimensional Strain is more Precise than Conventional Measures of Left Ventricular Systolic Function in Pediatric Patients.

Authors:  Meghna D Patel; Craig Myers; Kazuaki Negishi; Gautam K Singh; Shafkat Anwar
Journal:  Pediatr Cardiol       Date:  2019-11-09       Impact factor: 1.655

3.  Advantage and validation of vendor-independent software for myocardial strain analysis compared to vendor-specific software.

Authors:  Robert Chamberlain; Kenji Shiino; Gregory M Scalia; Surendran Sabapathy; Jonathan Chan
Journal:  Australas J Ultrasound Med       Date:  2020-10-14

4.  Novel functional advanced echocardiography for the assessment of myocardial mechanics in children with neurocardiogenic syncope - a blinded prospective speckle tracking head-up tilt-table challenge study.

Authors:  Kai O Hensel; Markus Roskopf; Francisca Abellan Schneyder; Andreas Heusch
Journal:  BMC Cardiovasc Disord       Date:  2018-05-08       Impact factor: 2.298

5.  Myocardial strain imaging: review of general principles, validation, and sources of discrepancies.

Authors:  M S Amzulescu; M De Craene; H Langet; A Pasquet; D Vancraeynest; A C Pouleur; J L Vanoverschelde; B L Gerber
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2019-06-01       Impact factor: 6.875

6.  Fast and Automated Segmentation for the Three-Directional Multi-Slice Cine Myocardial Velocity Mapping.

Authors:  Yinzhe Wu; Suzan Hatipoglu; Diego Alonso-Álvarez; Peter Gatehouse; Binghuan Li; Yikai Gao; David Firmin; Jennifer Keegan; Guang Yang
Journal:  Diagnostics (Basel)       Date:  2021-02-19

7.  British Society for Echocardiography and British Cardio-Oncology Society guideline for transthoracic echocardiographic assessment of adult cancer patients receiving anthracyclines and/or trastuzumab.

Authors:  Rebecca Dobson; Arjun K Ghosh; Bonnie Ky; Tom Marwick; Martin Stout; Allan Harkness; Rick Steeds; Shaun Robinson; David Oxborough; David Adlam; Susannah Stanway; Bushra Rana; Thomas Ingram; Liam Ring; Stuart Rosen; Chris Plummer; Charlotte Manisty; Mark Harbinson; Vishal Sharma; Keith Pearce; Alexander R Lyon; Daniel X Augustine
Journal:  Echo Res Pract       Date:  2021-03-31

8.  Global longitudinal strain as a prognostic marker in cardiac resynchronisation therapy: A systematic review.

Authors:  Vinesh Appadurai; Nicholas D'Elia; Thomas Mew; Stephen Tomlinson; Jonathan Chan; Christian Hamilton-Craig; Gregory M Scalia
Journal:  Int J Cardiol Heart Vasc       Date:  2021-07-31

Review 9.  A manifesto for cardiovascular imaging: addressing the human factor.

Authors:  Alan G Fraser
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2017-12-01       Impact factor: 6.875

10.  Effect of tafamidis on global longitudinal strain and myocardial work in transthyretin cardiac amyloidosis.

Authors:  Gerard T Giblin; Sarah A M Cuddy; Esther González-López; Alanna Sewell; Amanda Murphy; Sharmila Dorbala; Rodney H Falk
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2022-07-21       Impact factor: 9.130

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.