Accuracy and Trending of Continuous Noninvasive Hemoglobin Monitoring in Patients Undergoing Liver Transplantation.

P H Huang1, B F Shih1, Y-F Tsai2, P C H Chung2, F C Liu2, H P Yu2, W C Lee3, C J Chang4, C C Lin5.   

Abstract

BACKGROUND: Shift in large fluid volumes and massive blood loss during liver transplantation frequently leads to rapid changes in hemoglobin (Hb) concentration; thus, to ensure adequate tissue oxygenation, accurate and rapid determination of Hb concentration is essential in transplant recipients. The Radical-7 Pulse CO-Oximeter provides a noninvasive and continuous way to monitor Hb concentration (SpHb) in real time and is an ideal candidate for use during liver transplantation. In this study, we assessed the relationship between SpHb and total Hb (tHb) obtained from arterial blood samples during surgery.
METHODS: Forty patients undergoing liver transplantation were enrolled in this study. tHb and time-matched SpHb were measured at 5 different phases throughout surgery. Paired SpHb and tHb levels were assessed using linear regression, Bland-Altman analysis, and the Critchley polar plot method.
RESULTS: A total of 161 paired measurements with sufficient signal quality were analyzed. The correlation between SpHb and tHb was 0.59 (P < .001). Bland-Altman analysis revealed that a bias between SpHb and tHb was 2.28 g/dL, and limits of agreement (LoA) were from -0.78 to 5.34 g/dL. Trending analysis showed that 87% of data were located within the acceptable trending area, indicating that the trending ability was not satisfied.
CONCLUSIONS: The Radical-7 Pulse CO-Oximeter was not sufficient to monitor Hb levels and trends during liver transplantation surgery in our cohort. In particular, in critical patients and in those with low Hb levels, invasive Hb measurement should be used for assessment.
Copyright © 2016. Published by Elsevier Inc.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27320558     DOI: 10.1016/j.transproceed.2015.12.121

Source DB:  PubMed          Journal:  Transplant Proc        ISSN: 0041-1345            Impact factor:   1.066


  4 in total

1.  Comparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis.

Authors:  Hosein Shabaninejad; Nashmil Ghadimi; Kourosh Sayehmiri; Hossein Hosseinifard; Rasoul Azarfarin; Hasan Abolghasem Gorji
Journal:  J Anesth       Date:  2019-03-20       Impact factor: 2.078

2.  Validity of accuracy and trending ability of non-invasive continuous total hemoglobin measurement in complex spine surgery: a prospective cohort study.

Authors:  Feng-Cheng Chang; Jr-Rung Lin; Fu-Chao Liu
Journal:  BMC Anesthesiol       Date:  2019-07-04       Impact factor: 2.217

3.  Intraoperative continuous noninvasive hemoglobin monitoring in patients with placenta previa undergoing cesarean section: a prospective observational study.

Authors:  Hyunsung Kim; Sang-Hwan Do; Jung-Won Hwang; Hyo-Seok Na
Journal:  Anesth Pain Med (Seoul)       Date:  2019-10-31

4.  Accuracy of non-invasive haemoglobin measurements in patients undergoing transurethral resection of prostate surgery.

Authors:  Naveen Selvaraj; Gaurav Jain; Debendra Kumar Tripathy; Ankur Mittal; Haritha Indulekha
Journal:  Indian J Anaesth       Date:  2021-05-10
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.