Literature DB >> 27304083

Evaluation of the SAVI SCOUT Surgical Guidance System for Localization and Excision of Nonpalpable Breast Lesions: A Feasibility Study.

Victoria Mango1, Richard Ha1, Ameer Gomberawalla2, Ralph Wynn1, Sheldon Feldman2.   

Abstract

OBJECTIVE: The purpose of this study is to evaluate the feasibility of the SAVI SCOUT surgical guidance system, which uses a nonradioactive infrared-activated electromagnetic wave reflector, to localize and excise nonpalpable breast lesions. We evaluated the system's use in 15 nonpalpable breast lesions in 13 patients.
CONCLUSION: Image-guided placement was successful for 15 of 15 (100%) reflectors. The final pathologic analysis found that lesion excision was successful, including five malignancies with negative margins. No patients required reexcision or experienced complications. SAVI SCOUT is a feasible method for breast lesion localization and excision.

Entities:  

Keywords:  breast cancer; electromagnetic wave; infrared activated; localization

Year:  2016        PMID: 27304083     DOI: 10.2214/AJR.15.15962

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  8 in total

Review 1.  Innovations in image-guided preoperative breast lesion localization.

Authors:  Ellen Cheang; Richard Ha; Cynthia M Thornton; Victoria L Mango
Journal:  Br J Radiol       Date:  2018-02-06       Impact factor: 3.039

2.  Preoperative radioactive seed localization of nonpalpable soft tissue masses: an established localization technique with a new application.

Authors:  Hillary W Garner; Joseph M Bestic; Jeffrey J Peterson; Steven Attia; Daniel E Wessell
Journal:  Skeletal Radiol       Date:  2016-11-24       Impact factor: 2.199

3.  Impact of deformation on a supine-positioned image-guided breast surgery approach.

Authors:  Winona L Richey; Jon S Heiselman; Ma Luo; Ingrid M Meszoely; Michael I Miga
Journal:  Int J Comput Assist Radiol Surg       Date:  2021-08-12       Impact factor: 3.421

4.  Textual fiducial detection in breast conserving surgery for a near-real time image guidance system.

Authors:  Winona L Richey; Jon Heiselman; Ma Luo; Ingrid M Meszoely; Michael I Miga
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2020-03-16

5.  Preoperative Localization of Breast MRI Lesions: MRI-guided Marker Placement With Radioactive Seed Localization as an Alternative to MRI-guided Wire Localization.

Authors:  Nicole Saphier; Jessica Kondraciuk; Elizabeth Morris; Blanca Bernard-Davila; Victoria Mango
Journal:  J Breast Imaging       Date:  2020-04-23

Review 6.  Image-Guided Localization Techniques for Surgical Excision of Non-Palpable Breast Lesions: An Overview of Current Literature and Our Experience with Preoperative Skin Tattoo.

Authors:  Gianluca Franceschini; Elena Jane Mason; Cristina Grippo; Sabatino D'Archi; Anna D'Angelo; Lorenzo Scardina; Alejandro Martin Sanchez; Marco Conti; Charlotte Trombadori; Daniela Andreina Terribile; Alba Di Leone; Beatrice Carnassale; Paolo Belli; Riccardo Manfredi; Riccardo Masetti
Journal:  J Pers Med       Date:  2021-02-04

7.  Evaluation of the nonradioactive inducible magnetic seed system Magseed for preoperative localization of nonpalpable breast lesions - initial clinical experience.

Authors:  Karolina Pieszko; Mateusz Wichtowski; Marcin Cieciorowski; Robert Jamont; Dawid Murawa
Journal:  Contemp Oncol (Pozn)       Date:  2020-03-13

8.  Patient and clinician satisfaction and clinical outcomes of Magseed compared with wire-guided localisation for impalpable breast lesions.

Authors:  Aikaterini E Micha; Victoria Sinnett; Kate Downey; Steve Allen; Briony Bishop; Lauren R Hector; Elaine P Patrick; Ruth Edmonds; Peter A Barry; Katherine D C Krupa; Jennifer E Rusby
Journal:  Breast Cancer       Date:  2020-09-24       Impact factor: 4.239

  8 in total