Literature DB >> 27299146

Proper Angle of Sono-guided Central Venous Line Insertion.

Hassan Barzegari1, Arash Forouzan1, Mohammad Ali Fahimi1, Behzad Zohrevandi2, Mandana Ghanavati1.   

Abstract

INTRODUCTION: Determining the proper angle for inserting central venous catheter (CV line) is of great importance for decreasing the complications and increasing success rate. The present study was designed to determine the proper angle of needle insertion for internal jugular vein catheterization.
METHODS: In the present case series study, candidate patients for catheterization of the right internal jugular vein under guidance of ultrasonography were studied. At the time of proper placing of the catheter, photograph was taken and Auto Cad 2014 software was used to measure the angles of the needle in the sagittal and axial planes, as well as patient's head rotation. RESULT: 114 patients with the mean age of 56.96 ± 14.71 years were evaluated (68.4% male). The most common indications of catheterization were hemodialysis (55.3%) and shock state (24.6%). The mean angles of needle insertion were 102.15 ± 6.80 for axial plane, 36.21 ± 3.12 for sagittal plane and the mean head rotation angle was 40.49 ± 5.09.
CONCLUSION: Based on the results of the present study it seems that CV line insertion under the angles 102.15 ± 6.80 degrees in the axial plane, 36.21 ± 3.12 in the sagittal plane and 40.49 ± 5.09 head rotation yield satisfactory results.

Entities:  

Keywords:  Central venous catheters; catheterization; emergencies; ultrasonography; vascular access devices

Year:  2016        PMID: 27299146      PMCID: PMC4902211     

Source DB:  PubMed          Journal:  Emerg (Tehran)        ISSN: 2345-4563


Introduction:

Insertion of central venous catheter (CV line) is a common procedure to reach the central veins with the aim of monitoring the patients regarding hemodynamic status, measuring central venous pressure (CVP), fluid therapy, infusion of vasoactive agents, etc. (1, 2). CV line insertion may be done either with the guidance of anatomic landmarks or ultrasonography. One of the problems caused by this procedure is the risk of pneumothorax and arterial puncture, which usually occur due to improper selection of angle and depth of needle insertion. The angle for needle insertion in both methods is suggested to be 30 - 45 degrees. Yet, determining the angle, which minimizes the mentioned complications, is still needed (-). In a study, the proper needle tilt angle for internal jugular vein catheterization was shown to be 75 degrees and higher (6). The present study aimed to determine the proper angle of needle insertion for right internal jugular vein catheterization.

Methods:

Study design and setting The present case series study was designed to determine the proper angle of CV line insertion in internal jugular vein. Patients admitted to Imam Khomeini and Golestan Hospitals, Ahvaz, Iran were included. Based on the principles of Helsinki Declaration, researchers adhered to confidentiality of patient information and used the information only for research purposes. The study was approved by the Ethical Committee of Ahvaz Jondi Shapour University of Medical Sciences, Ahvaz, Iran. Written informed consent was obtained from all participants. Participants The participants were patients in need for catheterization of the right internal jugular vein, who were free of any evident deformity in chest and neck. Unsuccessful insertion of catheter, change in hemodynamic status or occurrence of dysrhythmia during procedure, and not permitting frequent photography and venipuncture, were among the most important exclusion criteria. Procedure After screening the cases in need of CV line insertion based on common standards (not being able to reach a peripheral vein, CVP monitoring, need to vasopressor agent, and …), patients underwent cardiopulmonary monitoring, pulse-oximetry, and nasal oxygen therapy. Then CV lines were inserted using the Seldinger method, under sterilized conditions, under ultrasonography guidance, and local anesthesia with lidocaine 1%. All patients were in supine or trendelenburg position. Sonography was done using a 7.5 MHz linear superficial probe under a sterilized nylon covering with a short axis approach on the right side of the neck in the anatomical position of internal jugular vein. Catheter needle was aimed at the apex of a triangle made from the crossing of the medial one-third of the clavicle bone in the base and 2 sternal and clavicular heads of sternocleidomastoid muscle towards the right nipple. After blood aspiration and ensuring the needle has entered the jugular vein, the needle and patient’s head position were kept fixed and photographs were taken from two perpendicular views, patient’s right side (sagittal plane) and patient’s down side (axial plane) (figure 1). The camera was equipped with a protractor application and the camera lens was in horizontal state.
Figure 1

Sample photographs taken

Taking a photo from the sagittal plane: While the patient was in supine position, the protractor was set to be parallel with the sagittal plane and when the needle entered the jugular vein and blood was aspirated, the center mark of the protractor was set at the place of needle insertion and the angle of needle with the horizontal line was measured. Taking a photo from the axial plane: While the patient was in supine position, the protractor was set to be parallel with the axial plane on the thoracic area and the center mark of the protractor was set at the place of needle insertion. Finally, when the needle entered the jugular vein and blood was aspirated, the angle of needle with horizontal line was measured. In addition, head rotation angle was measured in this plane by setting the zero point of the protractor on the mid sagittal line and measuring the angle of the line crossing the middle of the patient’s chin and the center mark of the protractor with vertical line. Chest imaging was done to verify the place of the catheter and evaluate complications such as pneumothorax and hemothorax. All Imaging and photography was performed by a trained emergency resident and 2 nurses. The angles were measured using Auto Cad 2014 software. Demographic data of participants as well as duration of procedure and measured angles were collected via a predesigned checklist. In the end, data were presented as mean ± standard deviation and frequency and percentage using SPSS version 20. Sample photographs taken Baseline characteristics of studied patients CVP: central venous pressure. Measured angles of head and needle in sagittal and axial planes SD: standard deviation.

Results:

120 patients were evaluated, 6 of which were excluded due to complications such as hematoma (4 cases) and carotid artery perforation (2 cases). The mean age of participants was 56.96 ± 14.71 years (68.4% male). The most common indications of catheterization were dialysis (55.3%) and shock state (24.4%). Table 1 depicts the baseline characteristics of studied patients. Table 2 shows the measured angles of head and needle in sagittal and axial planes.
Table 1

Baseline characteristics of studied patients

Variable
Age (year)
   Mean 56.96 ± 14.71
   Range 19 - 88
Sex number (%)
   Male 78 (68.4)
   Female 36 (31.6)
Body mass index
   > 30 102 (89.5)
   < 30 12 (10.5)
Cause of catheterization
   Shock 28 (24.6)
   Post resuscitation 13 (11.4)
   Hemodialysis 63 (55.3)
   Inaccessible peripheral vein 8 (7)
   Catheter change 1 (0.9)
   CVP measurement 1 (0.9)
Number of performers
   1 103 (90.4)
   2 11 (9.6)
Duration of catheterization (minutes)
   Mean 18.73 ± 1.51
   Range 15 - 23

CVP: central venous pressure.

Table 2

Measured angles of head and needle in sagittal and axial planes

Angles Mean ± SD (range)
Needle tilt (degree)
   In sagittal plane36.21 ± 3.12 (28-45)
   In axial plane102.15 ± 6.80 (90-115)
Head rotation (degree)
   In axial plane40.49 ± 5.09 (29-52)

SD: standard deviation.

Discussion:

The findings of the present study showed that CV line insertion under the angles 102.15 ± 6.80 degrees in the axial plane, 36.21 ± 3.12 in the sagittal plane, and 40.49 ± 5.09 head rotation may yield good results. In recent years, the number of patients in need of CV line insertion has increased due to various reasons (7). Arterial puncture, hematoma, pneumothorax, hemothorax, nerve injury, arrhythmia, air embolism, cardiac tear, tamponade, and probability of infection after surgery, are among the most important complications of catheterization (8). Catheterization, based on superficial landmarks is done in 1 attempt in about 20% of the time, while with the help of bedside ultrasonography, catheterization can be done with fewer complications and higher success rate (8). In a study aiming to determine the proper angle of needle insertion in the sagittal plane on the internal jugular vein, it was determined that a 16 degree insertion angle from the sagittal plane results in an increase in the probability of successful catheterization (9). In our study, mean sagittal angle of needle tilt from the horizontal line was about 36 degrees, which is vastly different from the mentioned study. In another study, the best angle for head rotation in internal jugular vein catheterization was reported to be ≥ 75 degrees (6). Head rotation in the present study was found to be 40 degrees from vertical line. To date, few studies have been carried out for determining the accurate axial angle of the needle. Considering the importance of this matter, it is suggested to design studies with proper sample size calculated based on existing data in the future.

Conclusion:

Based on the results of the present study it seems that CV line insertion under the angles 102.15 ± 6.80 degrees in the axial plane, 36.21 ± 3.12 in the sagittal plane and 40.49 ± 5.09 head rotation yield satisfactory results.
  8 in total

1.  Guidelines for performing ultrasound guided vascular cannulation: recommendations of the American Society of Echocardiography and the Society of Cardiovascular Anesthesiologists.

Authors:  Christopher A Troianos; Gregg S Hartman; Kathryn E Glas; Nikolaos J Skubas; Robert T Eberhardt; Jennifer D Walker; Scott T Reeves
Journal:  J Am Soc Echocardiogr       Date:  2011-12       Impact factor: 5.251

2.  [Tesio twin catheter system for hemodialysis tunnelized using an echo-guided technique. Retrospective analysis of 210 catheters].

Authors:  O Ibrik; R Samon; R Roca; J Viladoms; J Mora
Journal:  Nefrologia       Date:  2006       Impact factor: 2.033

3.  Guidelines for the prevention of intravascular catheter-related infections.

Authors:  Naomi P O'Grady; Mary Alexander; Lillian A Burns; E Patchen Dellinger; Jeffrey Garland; Stephen O Heard; Pamela A Lipsett; Henry Masur; Leonard A Mermel; Michele L Pearson; Issam I Raad; Adrienne G Randolph; Mark E Rupp; Sanjay Saint
Journal:  Am J Infect Control       Date:  2011-05       Impact factor: 2.918

4.  Needle tip visualization during ultrasound-guided vascular access: short-axis vs long-axis approach.

Authors:  Michael B Stone; Cynthia Moon; Darrell Sutijono; Michael Blaivas
Journal:  Am J Emerg Med       Date:  2010-01-28       Impact factor: 2.469

5.  Optimal angle of needle entry for internal jugular vein catheterization with a neutral head position: a CT study.

Authors:  Seda Ozbek; Seza Apiliogullari; Cengiz Erol; Ali Sami Kivrak; Inci Kara; Emine Uysal; Mustafa Koplay; Ateş Duman
Journal:  Ren Fail       Date:  2013-03-12       Impact factor: 2.606

6.  The Optimal Angle of Head Rotation for Internal Jugular Cannulation as Determined by Ultrasound Evaluation.

Authors:  Vincent DeAngelis; John Denny; Darrick Chyu; Thomas Jan; Anthony Lemaire; Antonio Chiricolo; Al Solina
Journal:  J Cardiothorac Vasc Anesth       Date:  2015-02-09       Impact factor: 2.628

7.  The anatomic relationship of femoral vein to femoral artery in euvolemic pediatric patients by ultrasonography: implications for pediatric femoral central venous access.

Authors:  Fred H Warkentine; Mary Clyde Pierce; Doug Lorenz; In K Kim
Journal:  Acad Emerg Med       Date:  2008-05       Impact factor: 3.451

8.  Real-time ultrasound-guided catheterisation of the internal jugular vein: a prospective comparison with the landmark technique in critical care patients.

Authors:  Dimitrios Karakitsos; Nicolaos Labropoulos; Eric De Groot; Alexandros P Patrianakos; Gregorios Kouraklis; John Poularas; George Samonis; Dimosthenis A Tsoutsos; Manousos M Konstadoulakis; Andreas Karabinis
Journal:  Crit Care       Date:  2006       Impact factor: 9.097

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.