Sarah J Andrewartha1, Suzanne L Munns2, Ashley Edwards3.
Abstract
Haemoglobin concentration ([Hb]) is measured for a wide variety of animal studies. The use of point-of-care devices, such as the HemoCue, is becoming increasingly common because of their portability, relative ease of use and low cost. In this study, we aimed to determine whether the [Hb] of blue-tongued skink (Tiliqua nigrolutea) blood can be determined accurately using the HemoCue and whether the HemoCue overestimates the [Hb] of reptile blood in a similar manner to fish blood. Additionally, we aimed to test whether ploidy affected [Hb] determined by the HemoCue using blood from diploid and triploid Atlantic salmon (Salmo salar). The HemoCue Hb 201(+) systematically overestimated [Hb] in both blue-tongued skinks and Atlantic salmon, and there was no difference between calibration equations determined for diploid or triploid salmon. The overestimation was systematic in both species and, as such, [Hb] determined by the HemoCue can be corrected using appropriate calibration equations.Entities:
Keywords: Diploid; HemoCue; fish; haemoglobin concentration; reptile; triploid
Year: 2016 PMID: 27293758 PMCID: PMC4804733 DOI: 10.1093/conphys/cow006
Source DB: PubMed Journal: Conserv Physiol ISSN: 2051-1434 Impact factor: 3.079
Summary of studies in a range of species that have validated the use of HemoCue for determination of haemoglobin concentration ([Hb])
| Species | Model | Comparison with standard techniques | Validation equation | Reference | |
|---|---|---|---|---|---|
| Mammals | Cats | HemoCue B | Comparable | Not required | |
| Rabbits | Not provided | Comparable | Not required | ||
| Pigs | Not provided | Comparable | Not required | ||
| Birds | Bar-headed geese | HemoCue201+ | Overestimation | (HemoCue[Hb]) = 1.408 (Drabkin’s’[Hb]) − 0.272 | |
| Fish | Sockeye salmon, Chinook salmon, Pacific bluefin tuna, Chub mackerel | HemoCue201+ | Overestimation | [Hb] = 0.815 (HemoCue [Hb]) − 2.198 Combined regression | |
| Atlantic salmon | HemoCue201+ | Overestimation | [Hb] = 0.820 (HemoCue [Hb]) − 8.883 | Present study | |
| Reptiles | Blue-tongued skink | HemoCue201+ | Overestimation | [Hb] = 0.885 (HemoCue [Hb]) + 7.498 | Present study |
Figure 1:(A) Comparison of haemoglobin concentration ([Hb]) determined using HemoCue system and Drabkin’s method in blue-tongued skinks (grey squares), triploid (open circles) and diploid (filled circles) Atlantic salmon. Continuous lines are species-specific regressions for skinks (y = 0.885x + 7.498, r2 = 0.985) and salmon (y = 0.820x − 5.831, r2 = 0.975) with 95% confidence intervals (dotted). The long dashed line represents the line of equality and the short dashed line is fish blood taken from Clark ). (B) [Hb] measured using Drabkin’s method against the difference in [Hb] determined by Drabkin’s method and the Hemocue {Δ[Hb], equation (3)}. The dashed line is the line of equality.