Dongsheng Jiang1,2, Yonghong Shi1,2, Demin Yao1,2, Manning Wang3,4, Zhijian Song5,6.
Abstract
PURPOSE: Computer-assisted intervention often depends on multimodal deformable registration to provide complementary information. However, multimodal deformable registration remains a challenging task.Entities:
Keywords: Deformable registration; Multimodal image; Similarity measure; Ultrasound
Mesh:
Year: 2016 PMID: 27250854 PMCID: PMC4893381 DOI: 10.1007/s11548-016-1407-2
Source DB: PubMed Journal: Int J Comput Assist Radiol Surg ISSN: 1861-6410 Impact factor: 2.924
Fig. 1The miLBP and SSC descriptor images for clean and degraded T1 and T2 images. The first row from left to right: the original T1 image, T1 image with 3 % noise and 40 %, the original T2 image, T2 image with 3 % noise and 40 % intensity nonuniformity. The second and third rows are the corresponding descriptor images for 2D miLBP and 2D SSC. For 2D miLBP, we used 8 neighbors for miLBP clacluation and map the 8-bit binary strings to integer values in the range of [0, 255]. For the 4D vector-valued descriptor of 2D SSC, we only show the first-dimensional for displaying
Fig. 2Illustration of obtaining the LBP (a) and the miLBP (b) in 2D
Fig. 3Artificial images representing CT and MRI with different levels of noise and their corresponding descriptor images of miLBP and SSC. The left four images are the CT images. For 2D SSC, we only show the first-dimensional values of the 4D vector-valued descriptor
Fig. 4Illustration of the influence of the noise on the two similarity measures. Dissimilarity values at each noise level are plotted with respect to the translation of the CT image in the x, y direction in a range of ±5 pixels. The images were aligned at 0 displacements, where the dissimilarity value between the fixed MRI image and the moving CT image should be smallest
Fig. 5The warping index for different pairwise registrations using miLBP and SSC. For all the registration tests, the p value are lower than 0.05, which indicate the statistical significance. a BrainWeb. b RIRE
Fig. 6Average of the results obtained over all 20 registrations after displacement field distortion. a Original. b Average warped. c Average miLBP. d Average SSC
A list of currently published results in [4, 5, 8], and results using miLBP and SSC
| Case# | Initial | miLBP | SSC | SeSaMI | CoCoMI | LC2 |
|---|---|---|---|---|---|---|
| 1 | 6.30 | 1.79 |
| 1.82 | 3.22 | 1.64 |
| 2 | 9.38 | 2.42 |
| 2.54 | 3.03 | 2.43 |
| 3 | 3.93 |
| 1.65 | 1.96 | 2.17 | 1.91 |
| 4 | 2.62 |
| 1.98 | 2.59 | 2.18 | 2.26 |
| 5 | 2.30 | 1.85 | 1.88 |
| 2.20 | 2.20 |
| 6 | 3.04 | 2.40 | 2.59 |
| 2.13 | 2.52 |
| 7 | 3.75 |
| 2.10 | 2.91 | 2.00 | 3.64 |
| 8 | 5.09 | 2.40 | 2.42 | 2.52 |
| 2.65 |
| 9 | 2.99 |
| 1.83 | 2.74 | 2.04 | 2.09 |
| 10 | 1.52 | 1.88 | 2.05 |
| 2.48 | 1.76 |
| 11 | 3.70 | 2.42 | 2.80 | 2.78 |
| 2.45 |
| 12 | 5.15 | 3.20 | 3.55 | 2.91 |
| 3.71 |
| 13 | 3.78 | 2.52 | 2.49 | 2.16 |
| 2.76 |
| Avg. |
|
|
|
|
|
|
| Avg. time |
|
| 35.2 s |
|
| 270 s |
The last two rows show the average mTRE and time of all cases
The minimum error in each row is highlighted with bold letters
Fig. 7Deformable MRI-US registration results of case 1 using miLBP. a MRI slice. b The corresponding US slice before registration. The corresponding US slices before and after deformable registration shown in pseudo-color are overlaid on the MRI image in c and d, respectively