On-demand calibration and evaluation for electromagnetically tracked laparoscope in augmented reality visualization.

Xinyang Liu1, William Plishker2, George Zaki2, Sukryool Kang1, Timothy D Kane1, Raj Shekhar3.   

Abstract

PURPOSE: Common camera calibration methods employed in current laparoscopic augmented reality systems require the acquisition of multiple images of an entire checkerboard pattern from various poses. This lengthy procedure prevents performing laparoscope calibration in the operating room (OR). The purpose of this work was to develop a fast calibration method for electromagnetically (EM) tracked laparoscopes, such that the calibration can be performed in the OR on demand.
METHODS: We designed a mechanical tracking mount to uniquely and snugly position an EM sensor to an appropriate location on a conventional laparoscope. A tool named fCalib was developed to calibrate intrinsic camera parameters, distortion coefficients, and extrinsic parameters (transformation between the scope lens coordinate system and the EM sensor coordinate system) using a single image that shows an arbitrary portion of a special target pattern. For quick evaluation of calibration results in the OR, we integrated a tube phantom with fCalib prototype and overlaid a virtual representation of the tube on the live video scene.
RESULTS: We compared spatial target registration error between the common OpenCV method and the fCalib method in a laboratory setting. In addition, we compared the calibration re-projection error between the EM tracking-based fCalib and the optical tracking-based fCalib in a clinical setting. Our results suggest that the proposed method is comparable to the OpenCV method. However, changing the environment, e.g., inserting or removing surgical tools, might affect re-projection accuracy for the EM tracking-based approach. Computational time of the fCalib method averaged 14.0 s (range 3.5 s-22.7 s).
CONCLUSIONS: We developed and validated a prototype for fast calibration and evaluation of EM tracked conventional (forward viewing) laparoscopes. The calibration method achieved acceptable accuracy and was relatively fast and easy to be performed in the OR on demand.

Entities:  

Keywords:  Augmented reality; Camera calibration; Electromagnetic tracking; Laparoscopic procedure; Laparoscopic visualization

Mesh:

Year:  2016        PMID: 27250853      PMCID: PMC5290310          DOI: 10.1007/s11548-016-1406-3

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  11 in total

1.  A new solution for camera calibration and real-time image distortion correction in medical endoscopy-initial technical evaluation.

Authors:  Rui Melo; João P Barreto; Gabriel Falcão
Journal:  IEEE Trans Biomed Eng       Date:  2011-11-23       Impact factor: 4.538

2.  Electromagnetic tracking for US-guided interventions: standardized assessment of a new compact field generator.

Authors:  A M Franz; K März; J Hummel; W Birkfellner; R Bendl; S Delorme; H-P Schlemmer; H-P Meinzer; L Maier-Hein
Journal:  Int J Comput Assist Radiol Surg       Date:  2012-05-24       Impact factor: 2.924

3.  Standardized assessment of new electromagnetic field generators in an interventional radiology setting.

Authors:  L Maier-Hein; A M Franz; W Birkfellner; J Hummel; I Gergel; I Wegner; H-P Meinzer
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

4.  Fused video and ultrasound images for minimally invasive partial nephrectomy: a phantom study.

Authors:  Carling L Cheung; Chris Wedlake; John Moore; Stephen E Pautler; Terry M Peters
Journal:  Med Image Comput Comput Assist Interv       Date:  2010

5.  DaVinci canvas: a telerobotic surgical system with integrated, robot-assisted, laparoscopic ultrasound capability.

Authors:  Joshua Leven; Darius Burschka; Rajesh Kumar; Gary Zhang; Steve Blumenkranz; Xiangtian Donald Dai; Mike Awad; Gregory D Hager; Mike Marohn; Mike Choti; Chris Hasser; Russell H Taylor
Journal:  Med Image Comput Comput Assist Interv       Date:  2005

6.  Intraoperative laparoscope augmentation for port placement and resection planning in minimally invasive liver resection.

Authors:  Marco Feuerstein; Thomas Mussack; Sandro M Heining; Nassir Navab
Journal:  IEEE Trans Med Imaging       Date:  2008-03       Impact factor: 10.048

7.  Magneto-optical tracking of flexible laparoscopic ultrasound: model-based online detection and correction of magnetic tracking errors.

Authors:  Marco Feuerstein; Tobias Reichl; Jakob Vogel; Joerg Traub; Nassir Navab
Journal:  IEEE Trans Med Imaging       Date:  2009-02-10       Impact factor: 10.048

8.  Electromagnetic tracking in the clinical environment.

Authors:  Ziv Yaniv; Emmanuel Wilson; David Lindisch; Kevin Cleary
Journal:  Med Phys       Date:  2009-03       Impact factor: 4.071

9.  Live augmented reality: a new visualization method for laparoscopic surgery using continuous volumetric computed tomography.

Authors:  Raj Shekhar; Omkar Dandekar; Venkatesh Bhat; Mathew Philip; Peng Lei; Carlos Godinez; Erica Sutton; Ivan George; Steven Kavic; Reuben Mezrich; Adrian Park
Journal:  Surg Endosc       Date:  2010-02-21       Impact factor: 4.584

10.  Stereoscopic augmented reality for laparoscopic surgery.

Authors:  Xin Kang; Mahdi Azizian; Emmanuel Wilson; Kyle Wu; Aaron D Martin; Timothy D Kane; Craig A Peters; Kevin Cleary; Raj Shekhar
Journal:  Surg Endosc       Date:  2014-02-01       Impact factor: 4.584

View more
  8 in total

1.  Preclinical evaluation of ultrasound-augmented needle navigation for laparoscopic liver ablation.

Authors:  Xinyang Liu; William Plishker; Timothy D Kane; David A Geller; Lung W Lau; Jun Tashiro; Karun Sharma; Raj Shekhar
Journal:  Int J Comput Assist Radiol Surg       Date:  2020-04-22       Impact factor: 2.924

2.  Anser EMT: the first open-source electromagnetic tracking platform for image-guided interventions.

Authors:  Herman Alexander Jaeger; Alfred Michael Franz; Kilian O'Donoghue; Alexander Seitel; Fabian Trauzettel; Lena Maier-Hein; Pádraig Cantillon-Murphy
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-03-29       Impact factor: 2.924

3.  Camera-Robot Calibration for the da Vinci® Robotic Surgery System.

Authors:  Orhan Özgüner; Thomas Shkurti; Siqi Huang; Ran Hao; Russell C Jackson; Wyatt S Newman; M Cenk Çavuşoğlu
Journal:  IEEE Trans Autom Sci Eng       Date:  2020-05-06       Impact factor: 5.083

4.  Laparoscopic stereoscopic augmented reality: toward a clinically viable electromagnetic tracking solution.

Authors:  Xinyang Liu; Sukryool Kang; William Plishker; George Zaki; Timothy D Kane; Raj Shekhar
Journal:  J Med Imaging (Bellingham)       Date:  2016-10-10

5.  Hybrid electromagnetic-ArUco tracking of laparoscopic ultrasound transducer in laparoscopic video.

Authors:  Xinyang Liu; William Plishker; Raj Shekhar
Journal:  J Med Imaging (Bellingham)       Date:  2021-02-03

6.  Measuring Respiratory Motion for Supporting the Minimally Invasive Destruction of Liver Tumors.

Authors:  Dominik Spinczyk
Journal:  Sensors (Basel)       Date:  2022-08-26       Impact factor: 3.847

7.  Fast calibration of electromagnetically tracked oblique-viewing rigid endoscopes.

Authors:  Xinyang Liu; Christina E Rice; Raj Shekhar
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-06-16       Impact factor: 3.421

8.  Electromagnetic tracking in image-guided laparoscopic surgery: Comparison with optical tracking and feasibility study of a combined laparoscope and laparoscopic ultrasound system.

Authors:  Guofang Xiao; Ester Bonmati; Stephen Thompson; Joe Evans; John Hipwell; Daniil Nikitichev; Kurinchi Gurusamy; Sébastien Ourselin; David J Hawkes; Brian Davidson; Matthew J Clarkson
Journal:  Med Phys       Date:  2018-10-19       Impact factor: 4.071

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.