| Literature DB >> 27250434 |
Wei-Min Wang1, Chung-Hsiang Cheng1, Gabor Molnar2, Ing-Shouh Hwang1, Kuang-Yuh Huang3, Hans-Ulrich Danzebrink2, En-Te Hwu1.
Abstract
In this paper, an optical imaging module design for an astigmatic detection system (ADS) is presented. The module is based on a commercial optical pickup unit (OPU) and it contains a coaxial illuminant for illuminating a specimen. Furthermore, the imaging module facilitates viewing the specimen and the detection laser spot of the ADS with a lateral resolution of approximately 1 μm without requiring the removal of an element of the OPU. Two polarizers and one infrared filter are used to eliminate stray laser light in the OPU and stray light produced by the illuminant. Imaging modules designed for digital versatile disks (DVDs) and Blu-ray DVDs were demonstrated. Furthermore, the module can be used for imaging a small cantilever with approximate dimensions of 2 μm (width) × 5 μm (length), and therefore, it has the potential to be used in high-speed atomic force microscopy.Entities:
Year: 2016 PMID: 27250434 DOI: 10.1063/1.4952438
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523