Literature DB >> 2724949

Prosthetic hip failure: retrospective radiograph image analysis of the acetabular cup.

P R Jones1, C J Taylor, D W Hukins, M L Porter, K Hardinge.   

Abstract

A method has been developed for quantifying movement and wear of the acetabular component (cup) of total hip replacements (THR) from routine postoperative and review radiographs. The method uses both interactive and automatic computer image analysis techniques. Dimensions of the prosthesis are used to scale the measurements and so overcome variation in radiographic alignment. The application of the method is illustrated by retrospective investigations of cup migration and wear using review radiographs taken over a follow-up of at least 12 years.

Entities:  

Mesh:

Year:  1989        PMID: 2724949     DOI: 10.1016/0141-5425(89)90153-2

Source DB:  PubMed          Journal:  J Biomed Eng        ISSN: 0141-5425


  1 in total

1.  A new method to make 2-D wear measurements less sensitive to projection differences of cemented THAs.

Authors:  Bertram The; Gunnar Flivik; Ron L Diercks; Nico Verdonschot
Journal:  Clin Orthop Relat Res       Date:  2008-02-10       Impact factor: 4.176