| Literature DB >> 27247557 |
Henry Kleiner, Matthew B Podgorsak.
Abstract
BACKGROUND: The purpose of the study was to analyse the dosimetric differences when using 10 MV instead of 6 MV for VMAT treatment plans for post-prostatectomy irradiation of the prostate bed. METHODS AND MATERIALS: Ten post-prostatectomy prostate bed irradiation cases previously treated using 6 MV with volumetric modulated arc therapy (VMAT) were re-planned using 10 MV with VMAT. Prescription dose was 66.6 Gy with 1.8 Gy per fraction for 37 daily fractions. The same structure set, number of arcs, field sizes, and minimum dose to the Planning Target Volume (PTV) were used for both 6 MV and 10 MV plans. Results were collected for dose to Organs at Risk (OAR) constraints, dose to the target structures, number of monitor units for each arc, Body V5, Conformity Index, and Integral Dose. The mean values were used to compare the 6 MV and 10 MV results. To determine the statistical significance of the results, a paired Student t test and power analysis was performed.Entities:
Keywords: 10 MV; 6 MV; prostate bed; radiation therapy; volumetric modulated arc therapy (VMAT)
Year: 2016 PMID: 27247557 PMCID: PMC4852958 DOI: 10.1515/raon-2016-0012
Source DB: PubMed Journal: Radiol Oncol ISSN: 1318-2099 Impact factor: 2.991
Dose constraint limits adapted from RTOG protocol 0534
| Category | Dose Constraint |
|---|---|
| Bladder-CTV D50 | 65 Gy |
| Bladder-CTV D70 | 40 Gy |
| Rectum D35 | 65 Gy |
| Rectum D55 | 40 Gy |
| Right Femoral Head D10 | 50 Gy |
| Left Femoral Head D10 | 50 Gy |
CTV = Clinical Target Volume
Mean, standard deviation of the mean, percent increase, p-value, and power for both 6 MV and 10 MV are displayed
| Category | 6 MV Mean ± SDOM | 10 MV Mean ± SDOM | Percent Increase | Power | |
|---|---|---|---|---|---|
| Bladder-CTV D50 | 32.5 ± 4.3 Gy | 29.7 ± 3.9 Gy | −8.62% | 0.013 | 0.79 |
| Bladder-CTV D70 | 18.5 ± 3.7 Gy | 16.2 ± 3.2 Gy | −12.4% | 0.011 | 0.81 |
| Rectum D35 | 49.5 ± 3.3 Gy | 46.8 ± 3.9 Gy | −5.45% | 6.6 × 10-3 | 0.88 |
| Rectum D55 | 28.5 ± 2.7 Gy | 26.7 ± 2.7 Gy | −6.32% | 0.023 | 0.68 |
| Right Femoral Head D10 | 34.12 ± 0.86 Gy | 29.80 ± 0.99 Gy | −12.66% | 1.2 × 10-4 | 1.0 |
| Left Femoral Head D10 | 32.74 ± 0.94 Gy | 29.4 ± 1.1 Gy | −10.20% | 8.3 × 10-5 | 1.0 |
| CTV Min Dose | 65.53 ± 0.38 Gy | 65.29 ± 0.21 Gy | −0.3662% | 0.41 | 0.12 |
| CTV Max Dose | 71.01 ± 0.37 Gy | 70.53 ± 0.31 Gy | −0.6760% | 0.10 | 0.37 |
| CTV Mean Dose | 68.30 ± 0.29 Gy | 67.68 ± 0.27 Gy | −0.9078% | 0.019 | 0.72 |
| CTV Percent Volume Covered by the 95% Isodose Line | 100% | 100% | 0% | N/A | N/A |
| PTV Min Dose | 64.42 ± 0.29 Gy | 64.42 ± 0.29 Gy | 0% | 0.10 | 0.37 |
| PTV Max Dose | 71.78 ± 0.29 Gy | 71.76 ± 0.33 Gy | −0.02786% | 0.94 | 0.051 |
| PTV Mean Dose | 68.39 ± 0.38 Gy | 67.94 ± 0.28 Gy | −0.6580% | 0.063 | 0.47 |
| PTV Percent Volume Covered by the 95% Isodose Line | 100% | 100% | 0% | N/A | N/A |
| Body V5 | (27.0 ± 1.0)% | (26.5 ± 1.0)% | −1.85% | 2.2 × 10-3 | 0.96 |
| Global Max Dose | 71.80 ± 0.38 Gy | 71.76 ± 0.33 Gy | −0.05571% | 0.89 | 0.052 |
| Arc 1 Monitor Units | 325 ± 17 MU | 311.8 ± 9.8 MU | −4.06% | 0.41 | 0.12 |
| Arc 2 Monitor Units | 330 ± 15 MU | 312 ± 10 MU | −5.5% | 0.19 | 0.24 |
| Conformity Index | 1.127 ± 0.013 | 1.091 ± 0.015 | −3.194% | 6.8 × 10-4 | 0.99 |
| Integral Dose | 207 ± 12 Gy-L | 191 ± 11 Gy-L | −7.73% | 1.1 × 10-5 | 1.0 |
CTV = Clinical Target Volume; N/A = not applicable for that category; PTV = Planning Target Volume