Mihai Rimbaş1, Denise Carmen Mihaela Zahiu2, Andrei Mihai Voiosu1, Theodor Alexandru Voiosu1, Alina Ana-Maria Zlate2, Roxana Dinu2, Domenico Galasso3, Leonardo Minelli Grazioli3, Mariachiara Campanale3, Federico Barbaro3, Bogdan Radu Mateescu1, Bogdan Busuioc2, Tiberiu Iordache2, Oana Dolofan2, Adelina Maria Popescu4, Vasile Daniel Balaban5, Mircea Mihai Raducan6, Cristiano Spada3, Cristian Răsvan Băicuş6, Guido Costamagna3.
Abstract
BACKGROUND AND STUDY AIMS: In videocapsule endoscopy examination (VCE), subtle variations in mucosal hue or pattern such as those seen in ulcerations can be difficult to detect, depending on the experience of the reader. Our aim was to test whether virtual chromoendoscopy (VC) techniques, designed to enhance the contrast between the lesion and the normal mucosa, could improve the characterization of ulcerative mucosal lesions. PATIENTS AND METHODS: Fifteen trainees or young gastroenterologists with no experience in VCE were randomly assigned to evaluate 250 true ulcerative and 100 false ulcerative, difficult-to-interpret small bowel lesions, initially as white light images (WLI) and then, in a second round, with the addition of one VC setting or again as WLI, labeling them as real lesions or artifacts.Entities:
Year: 2016 PMID: 27227106 PMCID: PMC4874791 DOI: 10.1055/s-0042-106206
Source DB: PubMed Journal: Endosc Int Open ISSN: 2196-9736
Fig. 1Evaluation of a succession of frames depicting a small erosion (right lower quadrant) in the context of dirty luminal content; only the last image (e) was selected for the purpose of the study, being the least representative, but still suggestive for an ulcerative lesion.
Fig. 2Images representing a true ulcerative lesion (upper set, left upper quadrant) and an artifact mimicking an ulcerative one (lower set, left upper quadrant) and their 4 chromoendoscopic correspondents.
Performance characteristics in the OVERALL evaluation of the study images.
| Randomized VC setting | Evaluator | Accuracy for the first reading | Accuracy for the second reading | Difference in accuracy (%) [95 % CI] |
| WLI | 1 | 61.7 [56.5, 66.7] | 54.3 [49.0, 59.4] | – 7.4 [– 12.5, – 2.3] |
| 2 | 49.1 [43.9, 54.4] | 44.6 [39.5, 49.8] | – 4.6 [– 8.8, – 0.3] | |
| 3 | 49.1 [43.9, 54.4] | 52.6 [47.3, 57.7] | 3.4 [0.0, 6.8] | |
| Globally | 53.3 | 50.5 | – 2.9 [– 9.2, 3.5] | |
| FICE 1 | 4 | 63.4 [58.3, 68.3] | 52.3 [47.1, 57.5] | – 11.1 [– 16.2, – 6.0] |
| 5 | 61.4 [56.2, 66.4] | 63.4 [58.3, 68.3] | 2.0 [– 3.3, 7.3] | |
| 6 | 44.7 [39.6, 49.9] | 75.4 [70.6, 79.6] | 30.7 [24.3, 36.7] | |
| Globally | 56.5 | 63.7 | 7.1 [– 17.0, 31.3] | |
| FICE 2 | 7 | 63.1 [58.0, 68.0] | 73.4 [68.6, 77.8] | 10.3 [5.3, 15.2] |
| 8 | 65.4 [60.3, 70.2] | 58.9 [53.6, 63.9] | – 6.6 [– 12.0, – 1.1] | |
| 9 | 75.7 [71.0, 79.9] | 79.4 [74.9, 83.3] | 3.7 [– 0.5, 7.9] | |
| Globally | 68.1 | 70.6 | 2.5 [– 7.1, 12.1] | |
| FICE 3 | 10 | 62.0 [56.8, 66.9 | 77.7 [73.1, 81.8] | 15.7 [10.1, 21.2] |
| 11 | 58.9 [53.6, 63.9] | 50.6 [45.4, 55.8] | – 8.3 [– 14.2, – 2.2] | |
| 12 | 52.6 [47.3, 57.7] | 59.7 [54.5, 64.7] | 7.1 [1.3, 12.9] | |
| Globally | 57.8 | 62.7 | 4.9 [– 8.9, 18.6] | |
| Blue Mode | 13 | 36.9 [32.0, 42.0] | 62.3 [57.1, 67.2] | 25.4 [19.1, 31.4] |
| 14 | 71.7 [66.8, 76.2] | 77.1 [72.5, 81.2] | 5.4 [1.0, 9.8] | |
| 15 | 68.0 [62.9, 72.7] | 76.3 [71.6, 80.4] | 8.3 [3.3, 13.3] | |
| Globally | 58.9 | 71.9 | 13.0 [0.8, 25.3] |
VC, virtual chromoendoscopy; WLI, white light image; CI, confidence interval
statistically significant
stratified McNemar test
Fig. 3Comparison of accuracy in correctly identifying the ulcerative images for all readers at their first and second evaluations. The investigators were randomized for the second reading to evaluate the study images in WLI (O1 – 3), FICE 1 (O4 – 6), FICE 2 (O7 – 9), FICE 3 (O10 – 12) and Blue mode (O13 – 15). Of the overall evaluations, statistical significance was obtained only for Blue mode (*); WLI, white light imaging; O, observer.
Performance characteristics in the evaluation of the TRUE ULCERATIVE image set.
| Randomized VC setting | Evaluator | Accuracy for the first reading | Accuracy for the second reading | Difference in accuracy (%) [95 % CI] |
| WLI | 1 | 56.4 [50.2, 62.4] | 44.0 [38.0, 50.2] | – 12.4 [– 18.4, – 6.2] |
| 2 | 34.8 [29.2, 40.9] | 29.2 [23.9, 35.1] | – 5.6 [– 11.0, – 0.2] | |
| 3 | 38.8 [33.0, 45.0] | 38.0 [32.2, 44.2] | – 0.8 [– 4.4, 2.8] | |
| Globally | 43.3 | 37.1 | – 6.3 [– 12.9, 0.3] | |
| FICE 1 | 4 | 49.6 [43.5, 55.8] | 33.6 [28.0, 39.7] | – 16.0 [– 22.8, – 9.0] |
| 5 | 52.8 [46.6, 58.9] | 59.2 [53.0, 65.1] | 6.4 [0.0, 12.7] | |
| 6 | 32.1 [26.6, 38.2] | 80.7 [75.4, 85.1] | 48.6 [41.4, 54.9] | |
| Globally | 44.9 | 57.8 | 12.9 [– 24.1, 50.0] | |
| FICE 2 | 7 | 54.4 [48.2, 60.5] | 69.2 [63.2, 74.6] | 14.8 [8.7, 20.7] |
| 8 | 54.8 [48.6, 60.9] | 44.4 [38.4, 50.6] | – 10.4 [– 17.5, – 3.1] | |
| 9 | 90.0 [85.7, 93.1] | 91.6 [87.5, 94.4] | 1.6 [– 2.6, 5.9] | |
| Globally | 66.4 | 68.4 | 2.0 [– 12.3, 16.3] | |
| FICE 3 | 10 | 53.2 [47.0, 59.3] | 79.2 [73.7, 83.8] | 26.0 [19.4, 32.2] |
| 11 | 51.2 [45.0, 57.3] | 41.6 [35.7, 47.8] | – 9.6 [– 16.7, – 2.4] | |
| 12 | 44.8 [38.8, 51.0] | 46.4 [40.3, 52.6] | 1.6 [– 5.5, 8.7] | |
| Globally | 49.7 | 55.7 | 6.0 [– 14.6, 26.6] | |
| Blue Mode | 13 | 12.8 [9.2, 17.5] | 58.4 [52.2, 64.3] | 45.6 [39.0, 51.6] |
| 14 | 62.4 [56.3, 68.2] | 77.2 [71.6, 82.0] | 14.8 [9.9, 19.7] | |
| 15 | 68.0 [62.0, 73.5] | 75.2 [69.5, 80.1] | 7.2 [1.5, 12.8] | |
| Globally | 47.7 | 70.3 | 22.5 [– 0.5, 45.5] |
VC, virtual chromoendoscopy; WLI, white light image; CI, confidence interval
statistically significant
stratified McNemar test
p = 0.07
Performance characteristics in the evaluation of the FALSE ULCERATIVE image set.
| Randomized VC setting | Evaluator | Accuracy for the first reading | Accuracy for the second reading | Difference in accuracy (%) [95 % CI] |
| WLI | 1 | 75 [65.7, 82.5] | 80 [71.1, 86.7] | 5 [– 3.8, 13.8] |
| 2 | 85 [76.7, 90.7] | 83 [74.5, 89.1] | – 2 [– 8.9, 4.8] | |
| 3 | 75 [65.7, 82.5] | 89 [81.4, 93.7] | 14 [6.4, 22.1] | |
| Globally | 78.3 | 84.0 | 5.7 [– 3.4, 14.7] | |
| FICE 1 | 4 | 98 [93.0, 99.4] | 99 [94.6, 99.8] | 1 [– 3.7, 6.1] |
| 5 | 83 [74.5, 89.1] | 74 [64.6, 81.6] | – 9 [– 18.5, 5.0] | |
| 6 | 76 [66.8, 83.3] | 62 [52.2, 70.9] | – 14 [– 23.2, – 14.5] | |
| Globally | 85.7 | 78.3 | – 7.3 [– 16.0, 1.3] | |
| FICE 2 | 7 | 85 [76.7, 90.7] | 84 [75.6, 89.9] | – 1 [– 10.0, 8.0] |
| 8 | 92 [85.0, 95.9] | 95 [88.8, 97.8] | 3 [– 3.7, 10.0] | |
| 9 | 40 [30.9, 49.8] | 49 [39.4, 58.7] | 9 [– 1.6, 19.2] | |
| Globally | 72.3 | 76.0 | 3.7 [– 2.0, 9.4] | |
| FICE 3 | 10 | 84 [75.6, 89.9] | 74 [64.6, 81.6] | – 10 [– 19.3, – 7.0] |
| 11 | 78 [68.9, 85.0] | 73 [63.6, 80.7] | – 5 [– 15.9, 6.1] | |
| 12 | 72 [62.5, 79.9] | 93 [86.3, 96.6] | 21 [10.9, 31.0] | |
| Globally | 78.0 | 80.0 | 2.0 [– 16.8, 20.8] | |
| Blue Mode | 13 | 97 [91.5, 99.0] | 72 [62.5, 79.9] | – 25 [– 34.5, – 15.8] |
| 14 | 95 [88.8, 97.8] | 77 [67.8, 84.2] | – 18 [– 26.5, – 10.4] | |
| 15 | 68 [58.3, 76.3] | 79 [70.0, 85.8] | 11 [4.0, 21.3] | |
| Globally | 86.7 | 76.0 | – 10.7 [– 32.3, 10.9] |
VC, virtual chromoendoscopy; WLI, white light image; CI, confidence interval
statistically significant
stratified McNemar test
Interobserver agreement in the evaluation of the study images.
| Randomization |
|
|
| WLI | 0.395 [0.382 – 0.415] | 0.524 [0.477 – 0.565] |
| FICE 1 | 0.317 [0.267 – 0.384] | 0.097 [0.040 – 0.151] |
| FICE 2 | 0.150 [0.122 – 0.195] | 0.092 [0.086 – 0.119] |
| FICE 3 | 0.344 [0.329 – 0.383] | 0.235 [0.191 – 0.267] |
| Blue Mode | 0.205 [0.158 – 0.272] | 0.302 [0.280 – 0.380] |
WLI, white light endoscopic imaging; VC, virtual chromoendoscopy
For all 15 evaluators, for the initial WLI reading, the kappa value was 0.306 [0.293 – 0.321]