| Literature DB >> 27225549 |
Per Anders Rickard Hellstrom1, Anna Åkerberg1, Martin Ekström1, Mia Folke1.
Abstract
The aim of this pilot study was to investigate the possibility to find a correlation between the output from a portable pedobarography system and the walking intensity expressed as walking speed. The system uses shoe insoles with force sensing resistors and wireless transmission of the data via Bluetooth. The force-time integral, at the toe-off phase of the step, for the force sensors in the forward part of the right foot was used to measure impulse data for 10 subjects performing walks in three different walking speeds. This data was then corrected by multiplication with the step frequency. This pilot study indicates that the portable pedobarography system output shows a linear relationship with the walking intensity expressed as walking speed on an individual level.Mesh:
Year: 2016 PMID: 27225549