Literature DB >> 27225549

Walking Intensity Estimation with a Portable Pedobarography System.

Per Anders Rickard Hellstrom1, Anna Åkerberg1, Martin Ekström1, Mia Folke1.   

Abstract

The aim of this pilot study was to investigate the possibility to find a correlation between the output from a portable pedobarography system and the walking intensity expressed as walking speed. The system uses shoe insoles with force sensing resistors and wireless transmission of the data via Bluetooth. The force-time integral, at the toe-off phase of the step, for the force sensors in the forward part of the right foot was used to measure impulse data for 10 subjects performing walks in three different walking speeds. This data was then corrected by multiplication with the step frequency. This pilot study indicates that the portable pedobarography system output shows a linear relationship with the walking intensity expressed as walking speed on an individual level.

Mesh:

Year:  2016        PMID: 27225549

Source DB:  PubMed          Journal:  Stud Health Technol Inform        ISSN: 0926-9630


  3 in total

1.  Classifying sitting, standing, and walking using plantar force data.

Authors:  Kohle J Merry; Evan Macdonald; Megan MacPherson; Omar Aziz; Edward Park; Michael Ryan; Carolyn J Sparrey
Journal:  Med Biol Eng Comput       Date:  2021-01-08       Impact factor: 2.602

2.  Investigation of Impact of Walking Speed on Forces Acting on a Foot-Ground Unit.

Authors:  Barbara Jasiewicz; Ewa Klimiec; Piotr Guzdek; Grzegorz Kołaszczyński; Jacek Piekarski; Krzysztof Zaraska; Tomasz Potaczek
Journal:  Sensors (Basel)       Date:  2022-04-18       Impact factor: 3.847

3.  Evaluation of the IngVaL Pedobarography System for Monitoring of Walking Speed.

Authors:  Per Anders Rickard Hellstrom; Anna Åkerberg; Martin Ekström; Mia Folke
Journal:  Healthc Inform Res       Date:  2018-04-30
  3 in total