Livia Angeloni1,2, Daniele Passeri1, Melania Reggente1, Diego Mantovani2, Marco Rossi1,3. 1. Department of Basic and Applied Sciences for Engineering, SAPIENZA University of Rome, Via A. Scarpa 16, 00161 Rome, Italy. 2. Lab. for Biomaterials and Bioengineering (CRC-I), Dept. Min-Met-Materials Eng. &University Hospital Research Center, Laval University, Quebec City, Canada. 3. Research Center for Nanotechnology applied to Engineering of SAPIENZA University of Rome (CNIS), Piazzale A. Moro 5, 00185 Rome, Italy.
Abstract
Magnetic force microscopy (MFM) has been demonstrated as valuable technique for the characterization of magnetic nanomaterials. To be analyzed by MFM techniques, nanomaterials are generally deposited on flat substrates, resulting in an additional contrast in MFM images due to unavoidable heterogeneous electrostatic tip-sample interactions, which cannot be easily distinguished from the magnetic one. In order to correctly interpret MFM data, a method to remove the electrostatic contributions from MFM images is needed. In this work, we propose a new MFM technique, called controlled magnetization MFM (CM-MFM), based on the in situ control of the probe magnetization state, which allows the evaluation and the elimination of electrostatic contribution in MFM images. The effectiveness of the technique is demonstrated through a challenging case study, i.e., the analysis of superparamagnetic nanoparticles in absence of applied external magnetic field. Our CM-MFM technique allowed us to acquire magnetic images depurated of the electrostatic contributions, which revealed that the magnetic field generated by the tip is sufficient to completely orient the superparamagnetic nanoparticles and that the magnetic tip-sample interaction is describable through simple models once the electrostatic artifacts are removed.
Magnetic force microscopy (MFM) has been demonstrated as valuable technique for the characterization of magnetic nanomaterials. To be analyzed by MFM techniques, nanomaterials are generally deposited on flat substrates, resulting in an additional contrast in MFM images due to unavoidable heterogeneous electrostatic tip-sample interactions, which cannot be easily distinguished from the magnetic one. In order to correctly interpret MFM data, a method to remove the electrostatic contributions from MFM images is needed. In this work, we propose a new MFM technique, called controlled magnetization MFM (CM-MFM), based on the in situ control of the probe magnetization state, which allows the evaluation and the elimination of electrostatic contribution in MFM images. The effectiveness of the technique is demonstrated through a challenging case study, i.e., the analysis of superparamagnetic nanoparticles in absence of applied external magnetic field. Our CM-MFM technique allowed us to acquire magnetic images depurated of the electrostatic contributions, which revealed that the magnetic field generated by the tip is sufficient to completely orient the superparamagnetic nanoparticles and that the magnetic tip-sample interaction is describable through simple models once the electrostatic artifacts are removed.
The increasing interest in the study and the development of magnetic nanomaterials for
different technological applications123 has highlighted the need of
new tools and procedures for the characterization of magnetic properties at the
nanometer scale. Conventional techniques, such as superconducting quantum interference
devices (SQUID) magnetometry4 or vibrating sample magnetometry (VSM)5, are widely used for the characterization of magnetic nanomaterials, but
the magnetic characterization of single nanomaterials can be achieved only through the
use of techniques which combine the capability of positioning and imaging at the
nanometer scale with the probing of ultra-low magnetic fields. Accurate mapping of
ultra-low magnetic field distribution has been demonstrated through the use of scanning
magnetometry with Nitrogen-vacancy (N-V) color centers in diamond6789, which however requires a quite complex experimental setup. Therefore, despite the
growing interest in these methods, the use of techniques based on simpler setups is
still more widespread. Among them, magnetic force microscopy (MFM) is considered a
promising technique thanks to its lateral resolution comparable to transmission
techniques (10–20 nm), its applicability to all kinds of nanomaterials
without any particular sample preparation, both in air and in liquid, its high magnetic
sensitivity and its capability to map the magnetic evolution of a sample with respect to
an applied field10111213. Nevertheless, despite the wide employment
of MFM technique for the qualitative characterization of magnetic nanomaterials, only a
few studies have been performed using MFM for the measurements of their magnetic
properties1415161718. The difficulty of obtaining reliable
results is ascribable to a certain inconsistency between the experimental data and the
theoretical models describing the magnetic tip-sample interactions. This incongruence
has been mainly attributed to the not satisfactory description of the probe, generally
assumed as a single magnetic point dipole1920, and several attempts
have been made to find more accurate mathematical approaches21. Such
proposed models generally take into account only the magnetic tip-sample interactions,
but different authors2223 recently demonstrated that the signal
detected by MFM contains also a significant contribution due to long-range electrostatic
phenomena, which include the effect of fixed electric charges on the sample as well as
of topography-modulated tip-sample capacitive coupling. Therefore, it should be more
realistically described as the sum of a magnetic and an electrostatic contribution.
Consequently, the evaluation of the effects of electrostatic forces appears essential to
obtain accurate magnetic measurements by MFM. Only a few studies have been carried out
with the aim of distinguishing or eliminating the electrostatic signal in MFM images and
a few methodologies have been proposed. For example, in the case of homogenous samples,
the electrostatic contribution can be eliminated compensating the tip-sample contact
potential difference by the application of an appropriate bias voltage24.
Nevertheless, if the analyzed sample is heterogeneous as in the case of magnetic
nanomaterials deposited on flat substrates, the contact potential difference depends on
the actual position of the probe on the sample surface and the electrostatic
contribution cannot be removed by the application of a single, fixed, bias voltage
value. In order to evaluate and eliminate the electrostatic contribution also in the
case of heterogeneous samples, Jafaar et al.22 proposed the
combined use of the Kelvin probe force microscopy (KPFM) and MFM techniques, the former
allowing the measurement of the contact potential difference in each point of the
scanned area and its compensation by opportunely adjusting the applied bias voltage
during the scan. In switching magnetization MFM (SM-MFM), proposed by Cambel et
al.2526, the analyzed surface is scanned twice in tapping mode,
with opposite tip magnetization orientations, obtained by applying an opportune magnetic
field before each scan. If the magnetization state of the sample is not affected by the
external field applied to invert the probe magnetic moment and by the magnetic field
induced by the tip during the measurements, reversing the probe magnetization results in
the inversion of the detected magnetic contrast while the atomic and electrostatic
contributions remain unchanged. Thus, adding the traces obtained with opposite tip
moments the magnetic signal is annulled and only the contrast due to the atomic and
electrostatic tip-sample interactions is visible. On the contrary, subtracting the same
two traces the electrostatic signal is nullified and only the magnetic tip-sample
interactions give rise to the contrast in the image. Differential MFM is an analogous
method recently proposed by Wang et al.27, in which the two MFM
images with reversed polarization are acquired subsequently to the topography with the
tip maintained at a fixed distance (lift height) from the surface (lift mode). The
applicability of these techniques for the evaluation of the electrostatic and magnetic
signal is limited to hard ferromagnetic materials, having significant remanent magnetic
moment and coercivity sufficiently high to ensure a constant magnetization of the sample
even after the application of the external magnetic field necessary to invert the probe
magnetization and under the magnetic stray field induced by the tip during the scan. Due
to this limitation, SM-MFM and differential MFM are not applicable to the most of
magnetic nanomaterials, which exhibit magnetic characteristics near to the
superparamagnetic limit (i.e., having low or zero coercivity).In this work, we propose a new MFM technique, which we refer to controlled magnetization
MFM (CM-MFM), which allows the evaluation and the elimination of electrostatic
contribution in MFM images by controlling the tip magnetization state. The effectiveness
of the technique is verified through its application to the investigation of
superparamagnetic nanoparticles (NPs) in absence of external field, which represents one
of the most challenging targets of MFM. Magnetic images were acquired and rationalized
describing both the tip and the NP as magnetic dipoles, demonstrating the suitability of
this simple model when MFM data are depurated of the electrostatic contributions.
Magnetic force microscopy
Magnetic force microscopy (MFM) is a particular scanning probe microscopy (SPM)
technique, which allows one to detect tip-sample magnetostatic interaction forces
and to image them on the sample surface. This is obtained using an atomic force
microscopy (AFM) setup equipped with a magnetic tip, generally consisting in a
standard AFM Si probe coated with a magnetic layer with thickness of a few tens of
nanometers. The cantilever is set into oscillation at a frequency f close to
its first free resonance frequency f0. When the probe is brought
sufficiently close to the sample surface, the gradient along the z axis of
the tip-sample force F produces a variation in the dynamic behavior of the
cantilever, which can be described as a change in the phase shift:where Q and k are the cantilever
quality factor and spring constant in air and with the tip not interacting with the
sample, respectively, and is the gradient along the
vertical axis z of the vertical component of the (magnetostatic) tip-sample
interaction force (F)28. The actual
expression of F depends on both the geometry and the
magnetic domains configuration of both the tip and the sample. For example, on the
basis of experimental data the former has been modeled using either a single point
magnetic dipole, a cone with uniformly magnetized magnetic surface, or more exotic
magnetic structures212930. Also, different analytical expression
for F are obtained in case the sample is a single magnetic
dipole or a more complex structure with periodic magnetic domains293132. In particular, let us suppose that the sample is
represented by a single small magnetic NP with diameter d uniformly
magnetized with magnetization M, possibly coated with a
nonmagnetic layer with thickness c, which we describe as a
single point magnetic dipole with magnetic momentum
m = 1/6πMd3. If the tip can be modeled by a punctiform magnetic dipole with
momentum m, the magnetic phase shift
Δφ observed when the tip is placed on the top of
the NP, i.e., the symmetry axes of the tip and the NP coincide, can be described
using a one dimensional analytical model which is justified by the symmetry of the
problem. In this case, Δφ is given bywhere μ0 is the permeability of free space,
Δz is the lift height (the vertical distance between the tip
apex and sample surface, i.e., the top of the NP), A is
the set-point amplitude during the first pass in tapping mode, and
δ is the position of the equivalent
momentum m evaluated from the tip apex2933. MFM images are generally acquired in the so-called
‘lift height mode’. In this two-pass modality, each line is
scanned twice and two different images of the selected sample area are recorded. The
first scan is performed in standard tapping mode to acquire and record the
topography, while the second scan is performed in non-contact mode, in order to
detect only long-range interaction forces (e.g., magnetic and electric forces) and
obtain a magnetic map of the sample. During this second scan, the probe follows the
trajectory of the previously recorded sample profile at a selected distance
Δz (the lift height), which is maintained constant at each
point (x, y) of the scanned area, in order to eliminate any possible artifact
in the magnetic signal due to the variation in the long-range the tip-sample
interaction forces produced by the modulation of the actual tip-sample distance.
Controlled magnetization MFM
The experimental setup to perform controlled magnetization MFM (CM-MFM) consists in a
standard MFM apparatus equipped with an electromagnet placed under the sample, which
allows one to apply controllable out-of-plane static magnetic fields H in the
range
−480 Oe < H < +480 Oe
to the tip-sample system, without moving the probe from the scan area. Similar
systems have been already applied to vary the magnetization state of the sample in
order to study its magnetic evolution in response to magnetic fields3435363738. Conversely, here the system is used to in
situ control the magnetization state of the probe. The measurement procedure
consists in two different phases: (i) the calibration of the remanent magnetic
behavior of the MFM tip and (ii) the measurement of the ‘only
magnetic’ MFM contrast by recording and opportunely post-processing two
MFM images of the same sample area, acquired with two different magnetization states
of the probe.
Step I: Probe calibration
The calibration phase consists in the individuation of the characteristic parameters
of the remanent hysteresis loop of the probe. This hysteresis curve is the plot, as
a function of the value of magnetic field applied and then switched off, of the
remanence corresponding to the in-field minor hysteresis loop where the maximum
magnetization corresponds to the actual value of the magnetic field applied and then
switched off39404142. The parameters to be determined to
calibrate the tip are the remanent saturation magnetic field
H and the remanent coercivity
H. Different calibration methods have
already been developed for the characterization of the in-field magnetic
characteristics of the MFM probes434445. Here, we propose a
simple method to measure the remanent hysteresis loop of the MFM probe. The
procedure consists in the measurement of the MFM contrast detected on a high
coercivity reference sample after the application and the subsequent switching-off
of out-of-plane magnetic fields with different intensities. We use a commercial
floppy disk as a reference sample. The out-of-plane coercivity of this kind of
samples, significantly higher than the probe coercivity, allows us to ascribe the
changes in the MFM contrast exclusively to the variation of the magnetization state
of the sensor. At the same time, the well defined magnetic structure, consisting in
a periodic pattern of in-plane domains alternatively oriented in opposite direction,
allows us to easily measure the variations of the phase contrast in response to the
changes in the magnetization state of the tip. When the tip magnetization is
directed perpendicularly to the sample surface, the MFM contrast is maximum in
correspondence of the domains transitions (positive or negative depending on the
mutual direction of the involved tip and sample magnetic domains), where the
magnetic field generated by the floppy has only vertical component32.
Conversely, the MFM contrast is zero in correspondence of the internal domains
regions, where the magnetic field generated by the floppy has only horizontal
component32. Therefore, the (out-of-plane) remanent hysteresis
loop of the probe can be obtained plotting the MFM phase difference
Δφ between two adjacent
transition regions (i.e., the image contrast) as a function of the previously
applied (and eventually switched-off) external magnetic field H.
Δφ(H) is a
function of the remanent magnetization of the tip and the sample
(M and
M, respectively).
M depends on the applied magnetic field
H. Conversely, if a sample with high out-of-plane coercivity like the
floppy disk is used, M is independent of
H.Thus, the normalized remanent hysteresis curve of the MFM probe can be obtainedwhere is the MFM phase difference between two adjacent
transition regions detected when the remanent magnetization of the tip reaches its
saturation value (M).A typical remanent hysteresis loop of a standard MFM tip (MESP, Bruker Inc.) measured
with this method is reported in Fig. 1a. Experiments were
performed using a standard AFM apparatus (Icon, Bruker Inc.) provided with standard
MFM imaging technique and equipped with an in-house made CM-MFM setup. The latter is
an electromagnet constituted by a coil (with inner diameter 1 cm, outer
diameter 2.7 cm, and height 1.6 cm) supplied with direct
electric current through a dc power supply and placed under the sample holder. The
control of the power supply is external to and independent of the AFM electronics.
Therefore, our technique can be implemented on every AFM apparatus, providing that
enough room is available for the coil. At the beginning of the experiment, a
magnetic field H = 480 Oe was applied.
Switching off the magnetic field results in a partial demagnetization of the tip,
which reaches a near-saturation state corresponding to its maximum remanent
magnetization, i.e., its remanent saturation. An image of the magnetic domains of
the floppy was acquired (image A in Fig. 1b, from with the
value of Δφ marked with A
in Fig. 1a is determined). Then, several MFM images were
recorded after the application and the switching off of magnetic fields in the
opposite direction with increasing intensities (e.g., images B and C in Fig. 1b corresponding to the point B and C in Fig. 1a) down to a magnetic field
H = −480 Oe at which the
saturation of Δφ in the
opposite direction is observed (image D in Fig. 1b and point D
in Fig. 1a). Then, positive values of H are applied to
complete the hysteresis curve (e.g., images and points E and F). From the curve
reported in Fig. 1a, it is possible to individuate both the
saturation magnetic field H necessary to
obtain the saturation remanent magnetization of the probe, corresponding to its
maximum magnetic sensitivity, and the coercive magnetic field
H necessary to annul the remanent
magnetization of the probe46.
Figure 1
Experimental characterization of the remanent magnetic properties of a
standard MFM tip.
(a) Hysteresis curve of the MFM phase contrast as a function of the
magnetic field applied and subsequently switched off. (b) Examples of
MFM images of the periodic magnetic domains of a standard floppy disk, from
which the phase difference between two adjacent transition regions was
measured in order to determine the points of the hysteresis curve. Points
from A to F in panel (a) are obtained from images from A to F in
panel (b).
Step II: Determination of the magnetic signal
Once the remanent properties of the probe have been determined, the magnetic moment
of the tip can be in situ controlled through the application and the
subsequent switching off of appropriate magnetic fields. In particular,
+H must be applied to magnetize the
tip, while −H or
−H must be applied to annul or
invert the tip magnetization, respectively.In the CM-MFM procedure, a first scan of the area is performed at fixed
Δz with the tip magnetized in its saturation state (having
applied and then switched off a magnetic field
+H) and a ‘standard MFM
image’ is acquired. As previously discussed, such an image is affected
by both magnetic and electrostatic tip-sample interactions. Indeed, the
‘standard MFM signal’ obtained with the magnetized tip
(Δφ) is actually the
superimposition of both the ‘true’ magnetic signal
(Δφ) and the electrostatic
signal (Δφ), i.e.,which is schematically represented in Fig. 2a. After the first
scan, a magnetic field with intensity
−H is applied and switched
off. A second image is acquired with the same instrumental parameters and the same
lift height Δz, but with the probe having zero magnetization
(Fig. 2c). In this case, the signal detected with the
demagnetized tip Δφ is
represented by the only electrostatic contribution, i.e.,
Figure 2
Sketch of the tip-sample interactions in CM-MFM.
The sample is assumed constituted by magnetic domains with different
orientation (red arrows) and by distributed electric charges which are
responsible for a tip-sample electrostatic interaction not uniform on the
surface. The three configurations of the tip are characterized by different
magnetization of the tip: (a) tip with saturated
‘up’ magnetization; (b) tip with saturated
‘down’ magnetization; (c) demagnetized
tip.
Therefore, the magnetic contribution, which in the following we refer to as the
CM-MFM signal
Δφ,
can be obtained by subtracting the second image to the first one, i.e.,The use of this mode, which we called zero probe magnetization (ZM) mode,
allows the detection of the electrostatic and magnetic tip-sample interactions
independently of the magnetization state of the sample, enabling the analysis of
soft ferromagnetic, paramagnetic and superparamagnetic materials, to which the
SM-MFM and differential MFM are not applicable. The same CM-MFM instrumentation can
be also used to distinguish the electrostatic and magnetic signals in MFM images of
relatively hard ferromagnetic samples, the stray field of which could orient the
domains of the tip, thus reducing the effectiveness of the tip demagnetization
procedure. Indeed, a magnetic field with intensity
−H can be applied and
switched off after the first scan, following a procedure analogous to that used in
SM-MFM2526 or in differential MFM27. A second
MFM image is recorded with the probe magnetized along the opposite direction with
respect to the first one (Fig. 2b) and the electrostatic and
magnetic contributions can be evaluated by adding or subtracting the two images,
respectively.An example of application demonstrating the effectiveness of the method is shown in
Fig. 3, in which the characterization of a standard floppy
disk using CM-MFM in ZM mode is reported. The topography (Fig.
3a) shows the presence of a particle (likely dust) with height of some
hundreds of nanometers. The corresponding bright contrast
(Δφ = +1.42 deg)
in standard MFM image (Fig. 3b) would suggest a repulsive
tip-sample interaction, which however can hardly find a convincing physical
rationalization. The same contrast
Δφ = +1.42 deg is
observed in the electrostatic image acquired with the demagnetized tip (Fig. 3c), compatible with the presence of an electrostatic
interaction produced by the tip-sample capacitive coupling23. After
subtraction, the magnetic image is obtained (Fig. 3d), where
the magnetic domains of the floppy are correctly visualized but no contrast
(Δφ = 0 deg)
is observed in correspondence of the particle. This result experimentally
demonstrates the potential capability of our method to compensate electrostatic
phase shift signal resulting from the tip-sample capacitive coupling in magnetic
images.
Figure 3
CM-MFM characterization of a standard floppy disk.
(a) Topography of an area where a particle (likely dust) is observable
on the floppy surface and (b) corresponding standard MFM phase image
acquired with the magnetized tip. (c) Phase image acquired with the
demagnetized tip. (d) CM-MFM phase image obtained by subtracting
(c) from (b).
The value of H can be easily determined with
enough accuracy from the curve in Fig. 1a as the intersection
with the horizontal axis of the linear curve fitting of the two points immediately
below and above it. From a conceptual point of view, such a value can always be
applied to completely demagnetize the MFM probe. The sensitivity of the dc power
source used to generate the magnetic field, however, could prevent the application
of the exact value of H, as it is discussed in
details below. It is worth explicitly discussing the range of possible samples that
can be investigated using CM-MFM. As detailed above, CM-MFM in ZM mode is
particularly suitable for the analysis of superparamagnetic NPs, which cannot be
investigated by SM-MFM and differential MFM as the NP magnetization is reversed
together with that of the tip, preventing the inversion of contrast in the two
magnetic images. The technique is also effective on relatively hard magnetic
samples, like the floppy disk, the magnetic domains of which are not affected by the
magnetic fields applied during the magnetization and demagnetization procedures and
which generate magnetic fields not sufficient to (even partially) polarize the
demagnetized tip. In particular, the latter issue was explicitly verified in the
case of floppy disk. If the magnetic field it generates could orient the
demagnetized tip, no alternation between dark and bright stripes in correspondence
of domain transitions would be observed in the phase shift images, resulting in a
pattern of only dark (attractive) stripes with halved spatial period. Apart from
these experimental evidences, the weakness of the field generated by the floppy is
confirmed by its rough estimation from the phase contrast values it produces and
using analytical models present in literature32. We calculated this
field to be lower than 10−2 Oe at a lift
height of 100 nm and thus negligible with the respect to
H. The magnetization of the tip by the
magnetic stray field of the sample, which has never been observed with the
cantilevers we used in our experiments but it is likely to occur with low-coercive
tips, limits the application of our technique on magnetic materials much harder that
the floppy. In principle, these materials could be analyzed selecting different tip
with higher coercivity. The magnetic field generated by these materials, however,
would be so intense that electrostatic artifacts would be negligible. Really, even
in the case of the standard floppy disk, the topography induced electrostatic
artifacts modulate the standard MFM response but they are negligible with respect to
the magnetic signal, so that the latter is well described by analytical models which
consider only magnetic tip-sample interactions32. Also, it should be
observed that MFM mainly targets to nanomaterials, which are unlikely to generate
such intense magnetic fields, being more realistically near to the superparamagnetic
limit. Conversely, materials with magnetic properties similar to those of the tip
cannot still be analyzed with our technique. Indeed, even if the magnetic field they
produce cannot polarize the demagnetized tip, the demagnetization/magnetization
procedure could modify the orientation of their magnetic domains. This would not
represent a limit if in-field measurement are required, but prevents the possibility
of investigating their ‘pristine’ magnetic state.
Case study: analysis of superparamagnetic nanoparticles
The effectiveness and the potentialities of the CM-MFM approach have been
demonstrated through the study of superparamagnetic NPs, which probably represent
the most challenging kind of sample for the sensitivity of standard MFM. Indeed,
their nanometer size comparable with the MFM lateral resolution and their low
magnetic moments make the magnetic tip-sample forces comparable with the
corresponding electrostatic interactions23. Furthermore, because of
their superparamagnetic character, if the MFM measurements are carried without
applying any external magnetic field, the magnetization of the NPs is only due to
the magnetic field induced by the probe during the scan and, thus, is always
oriented along the same direction of the probe magnetization. The SM mode is thus
inapplicable, while the ZM mode of CM-MFM can be used to decouple electrostatic
artifacts from MFM images.The test sample was prepared depositing a colloidal solution of commercial
Fe3O4 NPs (Sigma Aldrich) with nominal diameter of
20 nm on a clean monocrystalline Si substrate. The analysis was carried
out using a MFM tip with a standard momentum (MESP, Bruker Inc.). During the
measurement session, both topographic and MFM images of the investigated areas were
continuously acquired. Subsequent images of the same area could allow us to monitor
possible gradual demagnetization of the tip as well as the occurrence of abrupt
phenomena, e.g., destructive tip-sample contacts or snatching of NPs, which however
have never occurred during the whole experiment. The calibration of the tip was
performed following the previously illustrated procedure, revealing a remanent
saturation field
H = 440 Oe
at which deg is measured. At the beginning of the
experiment, the probe was magnetized through the application of a magnetic field
with intensity H. The magnetic field was
switched off and ‘standard MFM’ images of NPs were recorded,
the sample being magnetized only through the magnetic field induced by the tip.
Then, a magnetic field −H was
applied (and switched off) in order to annul the tip magnetization. Electrostatic
images of the same sample area were then recorded with the demagnetized probe.
CM-MFM images were obtained by subtracting the electrostatic images to the standard
MFM images. An example of this CM-MFM characterization in ZM mode is reported in
Fig. 4. In particular, Fig. 4a,b
show the topography and the standard MFM image acquired at
Δz = 20 nm, respectively, of
an area of the sample where the biggest NP with diameter
d = 30 nm is surrounded by smaller NPs
with diameters d < 20 nm. In the
image recorded with the magnetized tip (Fig. 4b) a slight
negative contrast was detected in correspondence of the biggest NP, indicating a
weak attractive tip-NP interaction. Conversely, a positive contrast was detected in
correspondence of smaller NPs, indicating a repulsive tip-sample interaction.
Because of their superparamagnetic character, in absence of an external magnetic
field, NPs can be magnetized only by the tip magnetic stray field and, thus, only an
attractive magnetic tip-sample interaction can occur. Consequently, the observed
positive contrast cannot be attributable to magnetic phenomena. This result
indicates the presence of a significant electrostatic contribution in standard MFM
images, which results to be even higher than the magnetic one in correspondence of
small NPs. In the phase image acquired with the demagnetized tip (Fig. 4c) a positive contrast is observed in correspondence of all the
NPs, i.e., also in the bigger one which exhibited a negative contrast in the images
acquired with the magnetized tip. This confirms the presence of a significant
positive electrostatic contribution in correspondence of all the analyzed NPs.
Although the positive phase shift in correspondence of the NPs may indicate a
difference between their electric properties and those of the substrate, it is more
likely attributable to a topography-induced artifact due to the capacitive
tip-sample coupling, i.e., to the reduction of the tip-sample attractive forces in
correspondence of the NPs produced by the increasing of the average tip-sample
distance23. Finally, Fig. 4d shows the
image resulting form the subtraction of Fig. 4c (electrostatic
contribution detected with the demagnetized tip) to Fig. 4b
(standard MFM image acquired with the magnetized tip which is affected by both the
electrostatic and the magnetic contribution). It represents the magnetic
contribution, depurated of the electrostatic effects. A larger negative contrast is
observable in correspondence of the NP which exhibits a negative contrast in
standard MFM images (d = 30 nm).
Conversely, only a slight negative contrast is observed in correspondence of small
NPs (d < 20 nm) which can be
hardly distinguished from the noise. This can be ascribed to the small volume of the
NPs and, thus, to their magnetic stray field, lower than the probe sensitivity.
Figure 4
CM-MFM characterization of superparamagnetic NPs.
(a) Topography of an area where some NPs are visible and (b)
corresponding standard MFM phase image acquired with the magnetized tip.
(c) Phase image acquired with the demagnetized tip. (d)
CM-MFM phase image obtained by subtracting (c) from (b).
In order to further confirm the magnetic nature of the tip-sample interaction
detected in CM-MFM, we carried out an analysis of the phase contrast as a function
of the tip-sample distance (i.e., lift height Δz) and of the NPs
diameter d. Several images of the same sample area with different
Δz have been recorded with both the magnetized and the
demagnetized probe. For each image, the absolute value of the phase shift detected
in correspondence of the NPs has been determined as the difference between the mean
values of the phase measured inside the NP and in correspondence of an adjacent
region of substrate. The uncertainty in the measured values has been determined
combining the statistics on a small area corresponding to the top of the NP and on
the selected region of substrate. Figure 5a shows an example
of standard MFM phase contrast
Δφ as a function of the
lift height Δz obtained with the magnetized probe on a NP with
d = 30 nm. In correspondence of small
tip-sample separations
(Δz < 50 nm), an
attractive interaction with the NP is experienced by the tip (negative phase shift),
which can be ascribed to the predominance of magnetic interactions in this region.
Increasing the tip sample distance, the repulsive force decreases until becoming
null and then attractive
(Δz > 50 nm),
indicating the predominance of electrostatic forces at large tip-sample separations.
This is congruent with the faster distance-decay of magnetic forces (expected to be
proportional to z−4) with respect to the
electrostatic ones (expected to be proportional to
z−2)47. Nevertheless, although
the general trend of the data is not difficult to justify, the actual
Δφ(Δz)
data reported in Fig. 5a can be hardly rationalized through
simple analytical models and their not monotonic behavior is apparently ascribable
only to the occasional presence of a random bias in different MFM images. An
analogous
Δφ(Δz)
curve on the same NP obtained with the demagnetized probe is reported in Fig. 5b, which represents the electrostatic signal. As expected,
it is always positive, indicating a reduced tip-sample attractive force, and
decreases with the increasing of the distance. Also in this case, data can be hardly
rationalized through simple analytical models. Nevertheless, by subtracting the data
acquired with the demagnetized probe (Fig. 5b) to those
acquired with the magnetized probe (Fig. 5a) the CM-MFM signal
Δφ is
obtained (Fig. 5c). It turned out that a negative phase shift
indicating an attractive (i.e., magnetic) tip-NP interaction exists also at
distances larger than 50 nm. This was not detectable in standard MFM
images because of the preponderance, at large tip-sample distances, of the
electrostatic contribution with respect to the magnetic one. Thus, in absence of any
external magnetic field, the NP is magnetized by the tip stray field also when the
probe is at distance larger than 50 nm. This produces an attractive
magnetic interaction which, nevertheless, is not detectable by standard MFM
measurements due to the predominant electrostatic contribution, but which can be
revealed using CM-MFM in ZM mode. It is interesting to notice that while both the
standard MFM data (Fig. 5a) and the electrostatic ones (Fig. 5b) cannot be described by simple models and seem affected
by a remarkable uncertainty, the magnetic data (Fig. 5c) show
a monotonic trend. Moreover, the fit of
Δφ(Δz)
data reported in Fig. 5c using Eq. (2)
demonstrates that they are very well described by the simple model in which both the
tip and the NP are represented with punctiform magnetic dipoles. In particular,
being performed the images with set-point amplitude
A = 28 nm, from the
fit an experimental value of
δ = 39 ± 4 nm
is determined. This result indicates that, when the MFM signal is depurated of the
effect of electrostatic tip-sample forces, a MFM tip acts as a single-point dipole.
By averaging the results of an analogous analysis of data obtained on three NPs with
different diameters, a value of
δ = 45 ± 7 nm
has been obtained. Fig. 5d reports the
Δφ(d)
data measured at Δz = 20 nm
together with the corresponding fit which has been obtained using Eq.
(2) assuming
δ = 45 nm.
Also in this case, the experimental curve is well described by Eq.
(2) although the relatively narrow distribution of NP diameters does not
allow us to present more efficacious results. Considering that in these experimental
conditions the minimum value of phase shift we can reveal is
10−2 deg, we can evaluate that with the
present settings and equipment our technique could allow the study of NPs with
diameter not smaller than 10 nm. Interestingly, our results demonstrate
that the weak magnetic field generated by the tip is sufficient to completely orient
the magnetic domains of superparamagnetic NPs, which is a debated issue in the
scientific community of MFM users2948. Moreover, as a result of the
removal of electrostatic signal in MFM images through CM-MFM, the tip-sample
interaction is found to be describable with the simple one dimensional model of two
interacting magnetic dipoles.
Figure 5
Analysis of images collected using CM-MFM technique in ZM mode.
(a) Standard MFM phase contrast () as a
function of the lift height () measured on a
NP with diameter d = 30 nm using
a magnetized probe, which is affected by both electrostatic and magnetic
tip-sample interactions. (b) Phase contrast on the same NP as a
function of the lift height acquired with the demagnetized tip (), which is affected only by the electrostatic
tip-sample interactions. (c) CM-MFM phase contrast () as a function of the lift height obtained by
subtracting data in (b) from those in (a) with the
corresponding fit using the simple model of two magnetic dipoles in Eq. (2). (d) CM-MFM phase contrast () as a function of the NP diameter obtained
analyzing five NPs with the corresponding fit using the simple model of two
magnetic dipoles in Eq. (2).
Current limits and future perspectives
The results reported in this work demonstrate that, in principle, CM-MFM may
represent a powerful technique to delete electrostatic artifacts resulting from
tip-sample capacitive coupling in MFM images. Thus, CM-MFM images can be used to
deduce information on local magnetic properties of materials, e.g., magnetic
momentum or magnetization, with nanometer lateral resolution. Despite the
potentiality and the correctness of its working principle, however, we must point
out some current limitations of our technique, mainly due to practical issues
related to the experimental setup which basically lead to the incomplete
demagnetization of the probe. Understanding and solving these limitations represent
the main challenges of our current work of improvement of CM-MFM.Obviously, being a two-pass technique, the correctness of topographic images is an
essential prerequisite for the accuracy of CM-MFM. Artifacts in the reconstruction
of the topography, e.g., due to incorrect choice of instrumental parameters like
set-point, scan rate, or feedback gain, result in artifacts in the magnetic images
which cannot be corrected. This problem is somewhat more severe in CM-MFM as two
subsequent images of the same area have to be acquired, with the magnetized and the
demagnetized tip, respectively.These issues affect CM-MFM as well as any other two-pass technique. In addition,
since the MFM phase shift depends on the instrumental parameters (e.g., drive
frequency and amplitude, set-point amplitude), the same parameters must be used in
the calibration on the reference sample (e.g., the floppy) and in the analysis of
the investigated sample (e.g., the NPs).Other severe limitations can occur due to the incomplete demagnetization of the tip.
Depending on the sensitivity of the power supply, indeed, the experimental setup is
characterized by a minimum step allowed between two values of the applied magnetic
field. In our case, for instance, the minimum allowed step between two values of
magnetic field is 15 Oe. Apart from particular and occasional cases in
which −H coincides with one of
applicable values of magnetic field, this demonstrates that with our experimental
setup we cannot reach the complete demagnetized state of the tip. Indeed, referring
to Fig. 1a, we determine
−H = −278.0 ± 7.5 Oe
(which is in agreement with the (in-field) coercivity of about 400 Oe
reported by the manufacturer) while the closest value of magnetic field we can apply
is Oe. In correspondence of we observe a contrast in the images equal to deg, which is 10% of the saturation contrast deg. As Δφ is
proportional to the tip magnetization, the signal detected with the (partially)
demagnetized tip includes not only the electrostatic contribution but also a part of
the magnetic one. Therefore, when subtracting the two signals obtained with the
magnetized and (partially) demagnetized tip, respectively, the signal reflecting the
magnetic tip-sample interaction is also modified. The extent of this effect can be
evaluated, and the effect itself can be corrected, in the simple case of a sample
the magnetization of which is independent from that of the tip. In this case, Eq. (5) can be rewritten aswhere ε is a demagnetization ratio given by . Notably, the value of ε can be
determined from the MFM contrast on the floppy since on such a sample not only are
electrostatic effects negligible with respect to the magnetic ones, but also because
they are uniform over the surface. Therefore they are deleted when subtracting the
maximum and minimum phase shifts in the MFM images to evaluate the phase
contrast32. In this case, instead of by Eq.
(6), the CM-MFM signal is given byThus, despite the not complete demagnetization of the tip, the electrostatic
contribution is completely deleted in
Δφ,
but the magnetic signal is underestimated or overestimated. This effect can be
corrected since the magnetic signal can be obtained asAs ε can be easily determined from the MFM images of the floppy
reference sample, our technique allows us to obtain the correct intensity of the
magnetic signal depurated from the electrostatic contribution when the sample
magnetization is independent of that of the probe. It is worth noting that the same
result can be obtained even if is not chosen as close
as possible to H providing the corresponding
value of ε is determined. Minimizing the residual magnetization
of the tip (and thus ε), however, ensures the highest
signal-to-noise ratio. The consistency of this correction method is demonstrated by
data reported in Fig. 6. From the calibration of the probe on
the floppy (Fig. 6a), the value of
H = 510 Oe
is obtained at which the remanent magnetization of the tip can be considered
saturated. Two values of magnetic field are determined at which the tip can be
considered nearly demagnetized, i.e., − Oe
and − Oe, respectively. These correspond
to not null phase contrast values equal to deg and
deg (note that the the contrast is reversed
in correspondence of −), respectively. The
values of phase contrast were determined from statistics on lines corresponding to
maximum and minimum of the phase of the floppy domains in a certain area. Being
deg, in correspondence of these two nearly
demagnetized states of the tip the values
ε1 = 0.23 ± 0.01
and
ε2 = −0.36 ± 0.01
are determined, respectively, where the negative sign of
ε2 is due to the reversal of the contrast in
correspondence of −. In a different area
of the sample, the contrast between two selected points was used to determine the
Δφ signal as
a function of the lift height Δz in the two cases of nearly
demagnetized tip. As clearly shown in Fig. 6b, although
Δφ is
supposed to represent the ‘magnetic only’ signal, two
completely different curves are obtained. This is due to the fact that in the case
characterized by ε1 a fraction of the magnetic signal
is actually subtracted and thus the latter is underestimated. Conversely, in the
case characterized by ε2 a fraction of the magnetic
signal is added (due to reversal in sign of the phase contrast) and thus the latter
is overestimated. After correction using Eq. (9), however, the
same values of Δφ are obtained
(Fig. 5c), which confirms the consistency of the method.
The correction of CM-MFM data can effectively compensate the incomplete
demagnetization of the tip in case of samples the magnetization of which does not
depend on that of the tip, and can be applied not only to data obtained on selected
points but on the whole CM-MFM images.
Figure 6
(a) Tip calibration on the floppy, from which the remanent saturation
magnetic field H and two values near
the remanent coercive field are determined. (b) CM-MFM signal
() as a function of the lift height
() measured on the floppy in two cases
of nearly demagnetized tip, characterized by two values of the
demagnetization factor ε. (c) Corrected magnetic
signals () as a function of the lift height
(), obtained using Eq.
(9).
If the spin of the sample depends on that of the tip, like in superparamagnetic NPs,
the incorrectly subtracted fraction of
Δφ depends also on the
magnetization states of the NP when the tip is saturated and nearly demagnetized.
For instance, if the magnetic momentum of the NP is saturated neither when the tip
is saturated nor nearly demagnetized, the magnetization of the NP is proportional to
the magnetic field generated by the tip. This in turns is proportional to the tip
magnetization. Therefore, Eq. (7) can be rewritten asand thus the CM-MFM signal is given byThus, dividing
Δφ by
(1−ε2), the corrected
Δφ versus
Δz curves can be obtained. Depending on the field generated
by the tip, however, the magnetization of the NP can be saturated by the magnetized
tip but not by the nearly demagnetized one. Moreover, depending on the residual
magnetization of the tip, the magnetization state of the NP varies with the lift
height. In this case, the ratio between the spin of the NP when the tip is fully
magnetized and demagnetized is no longer proportional to ε. Thus,
the correction factor cannot be easily estimated. In order to demonstrate the
effectiveness of this correction procedure in the case of the analyzed NPs, we
acquired three sets of CM-MFM phase versus lift height curves using three different
values of the demagnetization coefficient, i.e.,
ε1 = 0.17,
ε2 = 0.48, and
ε3 = 0.60, determined on
the reference floppy as previously described. In Fig. 7a,
three of these curves, obtained on a NP with diameter
d = 25 nm, are reported. The curves
clearly show that the bigger ε the lower the
Δφ
signal, which is obviously due to the bigger fraction of the subtracted magnetic
signal a result of a bigger residual magnetization of the (nearly) demagnetized tip.
Assuming that the magnetic stray field generated by the saturated tip is not much
bigger than the saturation field of the NP, the CM-MFM signal can be corrected and
the magnetic signal can be obtained as . Thus, the
curves shown in Fig. 7b are obtained, which demonstrate that
the corrected values retrieved using the three
different values of ε coincide within the experimental
uncertainty. Although potentially capable to correct CM-MFM data, the described
procedures are admittedly a bit intricate and lengthen the whole CM-MFM procedure
introducing additional sources of uncertainty. Also, the sample may contain
different types of magnetic nanomaterials and the magnetic field generated by the
tip may not uniformly affect them. In this case the correction procedure cannot be
applied to the whole image but only to data collected in selected points. Also, even
at fixed lift height the correction factor may be not constant over the sample
surface. In addition, we note that the uncertainty in the value of the applied
magnetic field results in an uncertainty in the values of the correction factors
(which is included in the error bars in the graphics in Figs
6c and 7b). As the demagnetization is performed only at
the beginning of the experiment and, thus, all the points in the curves are obtained
with the very same value of ε, its uncertainty acts as a
coefficient which multiplies all the points and therefore it must be considered when
comparing curves obtained in different measurement sessions, i.e., after different
demagnetization steps. Thus, not with standing the possibility of correcting CM-MFM
data, the incomplete demagnetization of the tip represents a serious limitation to
the accuracy of the technique. In order to overcome this limitation, if one wants to
use the approach described in this work, an improved setup should be considered.
Enhanced sensitivity of the power supply can ensure a smaller residual magnetization
of the tip in its (nearly) demagnetized state. Also, a more effective
demagnetization procedure could be selected, e.g., through the use of damped
oscillating magnetic fields.
Figure 7
(a) CM-MFM signal () as a function of
the lift height () measured on a NP with
diameter d = 25 nm in three cases
of nearly demagnetized tip, characterized by three values of the
demagnetization factor ε. (b) Corrected magnetic
signals () as a function of the lift height
(), obtained using Eq.
(11).
Another current limit of CM-MFM is that the tip calibration procedure for the
determination of the coercive field is performed using the floppy reference sample
at a certain distance along the axis of the coil. The field generated by the coil,
nevertheless, decreases as the distance from the coil along its axis increases.
Therefore, if the sample is placed on the top of the coil, variations in the sample
thickness result in variations in the distance between the tip and the coil and,
thus, in the height at which the tip demagnetization procedure is performed. On
samples much thinner or thicker than the floppy (including possible additional
substrates), the demagnetization step is performed at height from the coil different
from that at which calibration was performed. This leads to an incorrect
demagnetization during the experiment with a residual magnetization of the probe
significantly bigger than that estimated, dramatically affecting the accuracy of the
measurement. In this work, great attention was paid to perform the tip
demagnetization at the same height in both the calibration step on the floppy and in
the analysis of the NPs. This strategy, however, may be hardly applicable in some
specific cases, e.g., if the sample to be analyzed is particularly thick. In this
case, no correction can be carried out. A solution to this limitation could consist
in the ex situ demagnetization of the tip, using a
‘demagnetization station’. Although modern AFM setups ensure
a relatively accurate positioning even after macroscopic displacements of tip and
sample (e.g., to shift between ‘sample load’ and
‘sample scan’ positions), this procedure may lead to drifts
in the imaged area. This may result in more time-consuming experimental session
especially when small areas and high resolutions are required. Another solution
could be the use of a different demagnetization procedure, i.e., using oscillating
damped magnetic field. Depending on the initial values of magnetic field, this
procedure could allow a certain margin of variation of the tip-coil distance without
compromising the accuracy of the tip demagnetization. Finally, another solution
could be the use of a coil integral with the AFM head and, thus, with the tip. This
would allow one to perform the demagnetization procedure at fixed and thus correct
coil-tip distance. This however could not be implemented on every AFM setup due to
the volume and weight of the coil. In particular, it could be hardly used in
‘tip scan’ AFM systems.Not with standing all the aforementioned limitations, CM-MFM has however the
potential, which we have not fully explored yet, for accurate characterization of
magnetic parameters of isolated nanomaterials. For example, the curves reported in
Fig. 7b can be fitted using Eq. (2)
to roughly estimate the mass magnetization saturation of the analyzed NP. Indeed:
k = 5 N/m and
Q = 170 have been determined
for the used cantilever;
m = 1 × 10−13 EMU
can be assumed, although actually only its order of magnitude is indicated by the
producer and thus we cannot be more accurate without its independent measurement;
the values of the other parameters can be determined by the fit itself, even if the
absence of proper statistics dramatically affects the accuracy of these values. With
all these assumptions,
M = 20 EMU/g is
determined which, considering the uncertainty in m, is
compatible with but smaller than the values (about
60–70 EMU/g) obtained on macroscopic
Fe3O4 NPs samples with similar diameters measured at room
temperature by conventional techniques reported in literature49.
Indeed, we must also add to the previously indicated sources of uncertainty the
assumption that the NP is perfectly spherical and that its magnetization is uniform
in the volume ignoring possible near-surface effects.
Conclusions
In conclusion, we have addressed the issue of the effect of electrostatic tip-sample
interactions in MFM, which generally prevent the acquisition of magnetic images and
thus limit the accuracy of magnetic measurements at the nanometer scale. We
developed a MFM-based approach in which the two subsequent images of the same area
are collected, one with the magnetized and one with the quasi-demagnetized probe,
which is possible after the determination of the remanent saturation and remanent
coercivity magnetic fields of the actually used probe performed using a reference
sample with periodically patterned magnetic domains.The effectiveness of our technique is demonstrated through a challenging case study,
i.e., the characterization of superparamagnetic NPs in absence of any applied
external magnetic field. Images reflecting the magnetic properties of the sample
have been obtained subtracting images acquired with the demagnetized tip to the
standard MFM images, demonstrating the effectiveness of our techniques in the
removal of electrostatic artifacts in MFM maps. Moreover, in addition to the
demonstration of the technique, the analysis of our data demonstrated that the
magnetic field produced by the magnetized tip is sufficient to completely orient the
magnetic domains of superparamagnetic NPs even in absence of any applied external
magnetic field. Once the electrostatic artifacts are removed, the tip-sample
interaction is well described by that of two single-point magnetic dipoles. The need
for an accurate control of the instrumental parameters, the effect of the sample
thickness, and the incomplete demagnetization of the probe still represent serious
limitations to CM-MFM which must be overcome to increase the accuracy of the
technique. Overall, our controlled magnetization MFM technique has been demonstrated
to allow us to perform real magnetic characterizations at the nanometer scale.
Additional Information
How to cite this article: Angeloni, L. et al. Removal of electrostatic
artifacts in magnetic force microscopy by controlled magnetization of the tip:
application to superparamagnetic nanoparticles. Sci. Rep.
6, 26293; doi: 10.1038/srep26293 (2016).
Authors: Sharon Schreiber; Mayur Savla; Denis V Pelekhov; Daniel F Iscru; Camelia Selcu; P Chris Hammel; Gunjan Agarwal Journal: Small Date: 2008-02 Impact factor: 13.281
Authors: Sibylle Sievers; Kai-Felix Braun; Dietmar Eberbeck; Stefan Gustafsson; Eva Olsson; Hans Werner Schumacher; Uwe Siegner Journal: Small Date: 2012-06-22 Impact factor: 13.281
Authors: Miriam Jaafar; Oscar Iglesias-Freire; Luis Serrano-Ramón; Manuel Ricardo Ibarra; Jose Maria de Teresa; Agustina Asenjo Journal: Beilstein J Nanotechnol Date: 2011-09-07 Impact factor: 3.649
Authors: Maksim V Sapozhnikov; Yuri V Petrov; Nikita S Gusev; Alexey G Temiryazev; Olga L Ermolaeva; Victor L Mironov; Oleg G Udalov Journal: Materials (Basel) Date: 2019-12-24 Impact factor: 3.623
Authors: Marc Fuhrmann; Anna Musyanovych; Ronald Thoelen; Sibylle von Bomhard; Hildegard Möbius Journal: Nanomaterials (Basel) Date: 2020-12-11 Impact factor: 5.076
Authors: Elis Helena de Campos Pinto Sinnecker; José Miguel García-Martín; Dora Altbir; José D'Albuquerque E Castro; João Paulo Sinnecker Journal: Materials (Basel) Date: 2021-03-23 Impact factor: 3.623