Literature DB >> 27174443

A novel position-sensitive mega-size dosimeter for photoneutrons in high-energy X-ray medical accelerators.

Mehdi Sohrabi1, Amir Hakimi2, Seyed Rabi Mahdavi3.   

Abstract

PURPOSE: A novel position-sensitive mega-size polycarbonate (MSPC) dosimeter is introduced. It provides photoneutron (PN) dose equivalent matrix of positions in and out of a beam of a high energy X-ray medical accelerator under a single exposure.
METHODS: A novel position-sensitive MSPC dosimeter was developed and applied. It has an effective etched area of 50×50cm(2), as used in this study, processed in a mega-size electrochemical etching chamber to amplify PN-induced-recoil tracks to a point viewed by the unaided eyes. Using such dosimeters, PN dose equivalents, dose equivalent profiles and isodose equivalent distribution of positions in and out of beams for different X-ray doses and field sizes were determined in a Siemens ONCOR Linac.
RESULTS: The PN dose equivalent at each position versus X-ray dose was linear up to 20Gy studied. As the field size increased, the PN dose equivalent in the beam was also increased but it remained constant at positions out of the beam up to 20cm away from the beam edge. The jaws and MLCs due to material differences and locations relative to the target produce different PN contributions.
CONCLUSIONS: The MSPC dosimeter introduced in this study is a perfect candidate for PN dosimetry with unique characteristics such as simplicity, efficiency, dose equivalent response, large size, flexibility to be bent, resembling the patient's skin, highly position-sensitive with high spatial resolution, highly insensitive to X-rays, continuity in measurements and need to a single dosimeter to obtain PN dose equivalent matrix data under a single X-ray exposure.
Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  High-energy X-rays; Mega-size; Photoneutron dosimeter; Polycarbonate; Position-sensitive; Radiotherapy

Mesh:

Substances:

Year:  2016        PMID: 27174443     DOI: 10.1016/j.ejmp.2016.05.002

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  Photoneutron spectrometry by novel multi-directional spherical neutron spectrometry system.

Authors:  Mehdi Sohrabi; Amir Hakimi
Journal:  Sci Rep       Date:  2021-02-05       Impact factor: 4.379

2.  Breakthrough whole body energy-specific and tissue-specific photoneutron dosimetry by novel miniature neutron dosimeter/spectrometer.

Authors:  Mehdi Sohrabi; Morteza Ebrahimzadeh Torkamani
Journal:  Sci Rep       Date:  2021-10-15       Impact factor: 4.379

3.  Breakthrough in 4π ion emission mechanism understanding in plasma focus devices.

Authors:  Mehdi Sohrabi; Arefe Zarinshad; Morteza Habibi
Journal:  Sci Rep       Date:  2016-12-12       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.