Comparison of Aerodynamic Particle Size Distribution Between a Next Generation Impactor and a Cascade Impactor at a Range of Flow Rates.

Hiroyuki Yoshida1, Akemi Kuwana2, Hiroko Shibata2, Ken-Ichi Izutsu2, Yukihiro Goda2.   

Abstract

Wide variation in respiratory flow rates between patients emphasizes the importance of evaluating the aerodynamic particle size distribution (APSD) of dry powder inhaler (DPI) using a multi-stage impactor at different flow rates. US Pharmacopeia recently listed modified configurations of the Andersen cascade impactor (ACI) and new sets of cut-off diameter specifications for the operation at flow rates of 60 and 90 L/min. The purpose of this study was to clarify the effect of these changes on the APSD of DPI products at varied flow rates. We obtained APSD profiles of four DPIs and device combinations, Relenza®-Diskhaler® (GlaxoSmithKline Co.), Seebri®-Breezhaler® (Novartis Pharma Co.), Pulmicort®-Turbuhaler® (Astrazeneca Co.), and Spiriva®-Handihaler® (Nippon Boehringer Ingelheim Co.) using Next Generation Impactors (NGIs) and ACIs at flow rates from 28.3 to 90 L/min to evaluate the difference in the use of previous and new sets of cut-off diameter specifications. Processing the data using the new specifications for ACI apparently reduced large differences in APSD obtained by NGI and ACI with the previous specifications at low and high flow rates in all the DPIs. Selecting the appropriate configuration of ACI corresponding to the flow rate provided comparable APSD profiles of Pulmicort®-Turbuhaler® to those using NGIs at varied flow rates. The results confirmed the relevance of the current US Pharmacopeia specifications for ACI analysis in obtaining APSD profiles of DPI products at wide flow rates.

Entities:  

Keywords:  Andersen cascade impactor; Next Generation Impactor; aerodynamic particle size distribution; cut-off diameter; dry powder inhaler

Mesh:

Substances:

Year:  2016        PMID: 27173989     DOI: 10.1208/s12249-016-0544-9

Source DB:  PubMed          Journal:  AAPS PharmSciTech        ISSN: 1530-9932            Impact factor:   3.246


  3 in total

1.  Glycoside scutellarin enhanced CD-MOF anchoring for laryngeal delivery.

Authors:  Kena Zhao; Tao Guo; Caifen Wang; Yong Zhou; Ting Xiong; Li Wu; Xue Li; Priyanka Mittal; Senlin Shi; Ruxandra Gref; Jiwen Zhang
Journal:  Acta Pharm Sin B       Date:  2020-05-07       Impact factor: 11.413

2.  Wind tunnel-based testing of a photoelectrochemical oxidative filter-based air purification unit in coronavirus and influenza aerosol removal and inactivation.

Authors:  Yuechen Qiao; My Yang; Ian A Marabella; Devin A J McGee; Bernard A Olson; Montserrat Torremorell; Christopher J Hogan
Journal:  Indoor Air       Date:  2021-05-07       Impact factor: 6.554

3.  Deposition Pattern of Polydisperse Dry Powders in Andersen Cascade Impactor - Aerodynamic Assessment for Inhalation Experimentally and In Silico.

Authors:  Janwit Dechraksa; Tan Suwandecha; Teerapol Srichana
Journal:  Turk J Pharm Sci       Date:  2020-02-19
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.