Electromagnetic guided couch and multileaf collimator tracking on a TrueBeam accelerator.

Rune Hansen1, Thomas Ravkilde1, Esben Schjødt Worm1, Jakob Toftegaard2, Cai Grau2, Kristijan Macek3, Per Rugaard Poulsen2.   

Abstract

PURPOSE: Couch and MLC tracking are two promising methods for real-time motion compensation during radiation therapy. So far, couch and MLC tracking experiments have mainly been performed by different research groups, and no direct comparison of couch and MLC tracking of volumetric modulated arc therapy (VMAT) plans has been published. The Varian TrueBeam 2.0 accelerator includes a prototype tracking system with selectable couch or MLC compensation. This study provides a direct comparison of the two tracking types with an otherwise identical setup.
METHODS: Several experiments were performed to characterize the geometric and dosimetric performance of electromagnetic guided couch and MLC tracking on a TrueBeam accelerator equipped with a Millennium MLC. The tracking system latency was determined without motion prediction as the time lag between sinusoidal target motion and the compensating motion of the couch or MLC as recorded by continuous MV portal imaging. The geometric and dosimetric tracking accuracies were measured in tracking experiments with motion phantoms that reproduced four prostate and four lung tumor trajectories. The geometric tracking error in beam's eye view was determined as the distance between an embedded gold marker and a circular MLC aperture in continuous MV images. The dosimetric tracking error was quantified as the measured 2%/2 mm gamma failure rate of a low and a high modulation VMAT plan delivered with the eight motion trajectories using a static dose distribution as reference.
RESULTS: The MLC tracking latency was approximately 146 ms for all sinusoidal period lengths while the couch tracking latency increased from 187 to 246 ms with decreasing period length due to limitations in the couch acceleration. The mean root-mean-square geometric error was 0.80 mm (couch tracking), 0.52 mm (MLC tracking), and 2.75 mm (no tracking) parallel to the MLC leaves and 0.66 mm (couch), 1.14 mm (MLC), and 2.41 mm (no tracking) perpendicular to the leaves. The motion-induced gamma failure rate was in mean 0.1% (couch tracking), 8.1% (MLC tracking), and 30.4% (no tracking) for prostate motion and 2.9% (couch), 2.4% (MLC), and 41.2% (no tracking) for lung tumor motion. The residual tracking errors were mainly caused by inadequate adaptation to fast lung tumor motion for couch tracking and to prostate motion perpendicular to the MLC leaves for MLC tracking.
CONCLUSIONS: Couch and MLC tracking markedly improved the geometric and dosimetric accuracies of VMAT delivery. However, the two tracking types have different strengths and weaknesses. While couch tracking can correct perfectly for slowly moving targets such as the prostate, MLC tracking may have considerably larger dose errors for persistent target shift perpendicular to the MLC leaves. Advantages of MLC tracking include faster dynamics with better adaptation to fast moving targets, the avoidance of moving the patient, and the potential to track target rotations and deformations.

Entities:  

Mesh:

Year:  2016        PMID: 27147350     DOI: 10.1118/1.4946815

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Respiratory signal prediction based on adaptive boosting and multi-layer perceptron neural network.

Authors:  W Z Sun; M Y Jiang; L Ren; J Dang; T You; F-F Yin
Journal:  Phys Med Biol       Date:  2017-08-03       Impact factor: 3.609

2.  Technical Note: In silico and experimental evaluation of two leaf-fitting algorithms for MLC tracking based on exposure error and plan complexity.

Authors:  Vincent Caillet; Ricky O'Brien; Douglas Moore; Per Poulsen; Tobias Pommer; Emma Colvill; Amit Sawant; Jeremy Booth; Paul Keall
Journal:  Med Phys       Date:  2019-03-04       Impact factor: 4.071

3.  Scatter imaging during lung stereotactic body radiation therapy characterized with phantom studies.

Authors:  Kevin C Jones; Julius Turian; Gage Redler; Gizem Cifter; John Strologas; Alistair Templeton; Damian Bernard; James C H Chu
Journal:  Phys Med Biol       Date:  2020-08-10       Impact factor: 3.609

4.  Predicting real-time 3D deformation field maps (DFM) based on volumetric cine MRI (VC-MRI) and artificial neural networks for on-board 4D target tracking: a feasibility study.

Authors:  Jonathan Pham; Wendy Harris; Wenzheng Sun; Zi Yang; Fang-Fang Yin; Lei Ren
Journal:  Phys Med Biol       Date:  2019-08-21       Impact factor: 3.609

5.  Couch and multileaf collimator tracking: A clinical feasibility study for pancreas and liver treatment.

Authors:  Lei Zhang; Thomas LoSasso; Pengpeng Zhang; Margie Hunt; Gig Mageras; Grace Tang
Journal:  Med Phys       Date:  2020-09-11       Impact factor: 4.071

6.  Tumor motion tracking based on a four-dimensional computed tomography respiratory motion model driven by an ultrasound tracking technique.

Authors:  Lai-Lei Ting; Ho-Chiao Chuang; Ai-Ho Liao; Chia-Chun Kuo; Hsiao-Wei Yu; Hsin-Chuan Tsai; Der-Chi Tien; Shiu-Chen Jeng; Jeng-Fong Chiou
Journal:  Quant Imaging Med Surg       Date:  2020-01

7.  Robust optimization of VMAT for lung cancer: Dosimetric implications of motion compensation techniques.

Authors:  Ben R Archibald-Heeren; Mikel V Byrne; Yunfei Hu; Meng Cai; Yang Wang
Journal:  J Appl Clin Med Phys       Date:  2017-08-08       Impact factor: 2.102

8.  The potential of Gantry beamline large momentum acceptance for real time tumour tracking in pencil beam scanning proton therapy.

Authors:  Giovanni Fattori; Ye Zhang; David Meer; Damien Charles Weber; Antony John Lomax; Sairos Safai
Journal:  Sci Rep       Date:  2020-09-18       Impact factor: 4.379

9.  Experimental investigation of dynamic real-time rotation-including dose reconstruction during prostate tracking radiotherapy.

Authors:  Casper Gammelmark Muurholm; Thomas Ravkilde; Robin De Roover; Simon Skouboe; Rune Hansen; Wouter Crijns; Tom Depuydt; Per R Poulsen
Journal:  Med Phys       Date:  2022-04-25       Impact factor: 4.506

10.  Fast Fourier transform combined with phase leading compensator for respiratory motion compensation system.

Authors:  Chia-Chun Kuo; Ho-Chiao Chuang; Ai-Ho Liao; Hsiao-Wei Yu; Syue-Ru Cai; Der-Chi Tien; Shiu-Chen Jeng; Jeng-Fong Chiou
Journal:  Quant Imaging Med Surg       Date:  2020-05
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.