Literature DB >> 27143725

Corrosion at the head-neck interface of current designs of modular femoral components: essential questions and answers relating to corrosion in modular head-neck junctions.

K Osman1, A P Panagiotidou1, M Khan2, G Blunn1, F S Haddad3.   

Abstract

There is increasing global awareness of adverse reactions to metal debris and elevated serum metal ion concentrations following the use of second generation metal-on-metal total hip arthroplasties. The high incidence of these complications can be largely attributed to corrosion at the head-neck interface. Severe corrosion of the taper is identified most commonly in association with larger diameter femoral heads. However, there is emerging evidence of varying levels of corrosion observed in retrieved components with smaller diameter femoral heads. This same mechanism of galvanic and mechanically-assisted crevice corrosion has been observed in metal-on-polyethylene and ceramic components, suggesting an inherent biomechanical problem with current designs of the head-neck interface. We provide a review of the fundamental questions and answers clinicians and researchers must understand regarding corrosion of the taper, and its relevance to current orthopaedic practice. Cite this article: Bone Joint J 2016;98-B:579-84. ©2016 The British Editorial Society of Bone & Joint Surgery.

Entities:  

Keywords:  Total hip arthroplasty; corrosion; crevice; failure; fretting; head-neck junction; modular; taper; trunnion; wear

Mesh:

Year:  2016        PMID: 27143725     DOI: 10.1302/0301-620X.98B5.35592

Source DB:  PubMed          Journal:  Bone Joint J        ISSN: 2049-4394            Impact factor:   5.082


  20 in total

1.  Outcomes of neck modularity in total hip arthroplasty: an Italian perspective.

Authors:  R Papalia; A Di Martino; A Caldaria; B Zampogna; V Denaro
Journal:  Musculoskelet Surg       Date:  2019-01-23

2.  Head-stem trunnion dissociation due to corrosion in total hip arthroplasty.

Authors:  Michael R Whitehouse
Journal:  Ann Transl Med       Date:  2017-09

Review 3.  Regenerative Medicine Strategies in Biomedical Implants.

Authors:  Divya Rani Bijukumar; Clay McGeehan; Mathew T Mathew
Journal:  Curr Osteoporos Rep       Date:  2018-06       Impact factor: 5.096

Review 4.  Corrosion of the Head-Stem Taper Junction-Are We on the Verge of an Epidemic?: Review Article.

Authors:  Michael Morlock; Dennis Bünte; Julian Gührs; Nicholas Bishop
Journal:  HSS J       Date:  2016-09-20

5.  Ceramic Bearings with Titanium Adapter Sleeves Implanted During Revision Hip Arthroplasty Show Minimal Fretting or Corrosion: a Retrieval Analysis.

Authors:  Chelsea N Koch; Mark Figgie; Mark P Figgie; Marcella E Elpers; Timothy M Wright; Douglas E Padgett
Journal:  HSS J       Date:  2017-06-22

6.  Acute trunnion failure of a TMZF alloy stem with large diameter femoral heads.

Authors:  E J Fahey; E Pomeroy; F E Rowan
Journal:  J Orthop       Date:  2020-01-10

7.  Are Damage Modes Related to Microstructure and Material Loss in Severely Damaged CoCrMo Femoral Heads?

Authors:  Stephanie M McCarthy; Deborah J Hall; Mathew T Mathew; Joshua J Jacobs; Hannah J Lundberg; Robin Pourzal
Journal:  Clin Orthop Relat Res       Date:  2021-09-01       Impact factor: 4.755

8.  Effect of Femoral Stem Modular Neck's Material on Metal Ion Release.

Authors:  Janie Barry; Marc-Olivier Kiss; Vincent Massé; Martin Lavigne; Jihad Matta; Pascal-Andre Vendittoli
Journal:  Open Orthop J       Date:  2017-11-29

9.  Exchangeable Femoral Neck (Dual-Modular) THA Prostheses Have Poorer Survivorship Than Other Designs: A Nationwide Cohort of 324,108 Patients.

Authors:  Sandrine Colas; Assia Allalou; Antoine Poichotte; Philippe Piriou; Rosemary Dray-Spira; Mahmoud Zureik
Journal:  Clin Orthop Relat Res       Date:  2017-02-13       Impact factor: 4.176

10.  Comparison of metal ion concentrations and implant survival after total hip arthroplasty with metal-on-metal versus metal-on-polyethylene articulations.

Authors:  Henrik Dahlstrand; André Stark; Marius C Wick; Lucas Anissian; Nils P Hailer; Rüdiger J Weiss
Journal:  Acta Orthop       Date:  2017-07-12       Impact factor: 3.717

View more