[Measurement of contrast vision: mesopic or photopic vision? : Comparison of different methods for measuring contrast sensitivity within the framework of driving licence regulations].

L C E Bergmann1, S Darius1, S Kropf2, I Böckelmann3.   

Abstract

BACKGROUND: Since 2011 the regulations for occupational driving licences make the examination of contrast vision sensitivity obligatory; however, apart from mesopic procedures no uniform regulations are available for methods and minimum requirements.
OBJECTIVE: By comparing different mesopic and photopic contrast sensitivity tests this study analyzed whether these could be equivalently used and lead to the same results.
MATERIAL AND METHODS: Contrast vision sensitivity was determined in 150 subjects with emmetropia using five different methods, i.e. the mesopic test device Mesotest II as the reference method, Rodatest 302 and Optovist as photopic test devices and two photopic test charts (Vistech chart and Mars charts). The results of passing the tests were compared and Cohens κ was determined to quantify the conformity between the tests.
RESULTS: Poor agreement was found between Mesotest II and Optovist as well as between Mesotest II and the Vistech chart. There was no agreement between Rodatest 302 or Mars charts and Mesotest II; nevertheless, the contrast vision sensitivity measured with Optovist, Rodatest 302 and the Vistech chart showed good correlation (0.46 ≤ r ≤ 0.69).
CONCLUSION: Apart from a few limitations, the reference method Mesotest II as well as Optovist and the Vistech chart are suitable for testing contrast vision sensitivity, whereas Rodatest 302 and Mars charts cannot be recommended based on the current criteria for minimum requirements. The minimum requirements must be urgently adapted and strict regulations for measurement must be formulated. In addition, due to the poor agreement between the methods an amendment of the driving licence regulations should be considered, which requires examination of both mesopic and photopic contrast vision or alternatively mesopic contrast vision alone.

Keywords:  Contrast sensitivity; Contrast vision; Driving licence regulations; Mesopic vision; Mesotest II

Mesh:

Year:  2016        PMID: 27142036     DOI: 10.1007/s00347-016-0261-5

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  13 in total

1.  Changes in the retinocortical evoked potentials in subjects 75 years of age and older.

Authors:  L Justino; H Kergoat; M J Kergoat
Journal:  Clin Neurophysiol       Date:  2001-07       Impact factor: 3.708

2.  [Experience in contrast vision testing].

Authors:  H Wilhelm; D Hofmann; S Roelcke; R Quast
Journal:  Klin Monbl Augenheilkd       Date:  2011-11-09       Impact factor: 0.700

3.  Improving the design of the letter contrast sensitivity test.

Authors:  Aries Arditi
Journal:  Invest Ophthalmol Vis Sci       Date:  2005-06       Impact factor: 4.799

Review 4.  [Photopic contrast sensitivity. Local contrast perception].

Authors:  M Bach; W Wesemann; G Kolling; J Bühren; H Krastel; U Schiefer
Journal:  Ophthalmologe       Date:  2008-01       Impact factor: 1.059

5.  [Assessment of mesopic and contrast vision for driving licences: which cut-off values, which methods are appropriate?].

Authors:  H Wilhelm; T Peters; W Durst; S Roelcke; R Quast; M Hütten; B Wilhelm
Journal:  Klin Monbl Augenheilkd       Date:  2013-11-04       Impact factor: 0.700

6.  Influence of chronic exposure to organic solvent mixtures on contrast sensitivity in silk-screen printers: matched-pair analysis.

Authors:  Irina Böckelmann; Eberhard Alexander Pfister; Brigitte Peters
Journal:  Environ Toxicol Pharmacol       Date:  2005-01-26       Impact factor: 4.860

7.  Marked dissociation of photopic and mesopic contrast sensitivity even in normal observers.

Authors:  Hannah Hertenstein; Michael Bach; Nikolai Johannes Gross; Flemming Beisse
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-04-29       Impact factor: 3.117

8.  [The examination on fitness for driving at darkness with the mesoptometer].

Authors:  E Aulhorn; H Harms
Journal:  Klin Monbl Augenheilkd       Date:  1970-12       Impact factor: 0.700

9.  Characteristics of dynamic processing in the visual field of patients with age-related maculopathy.

Authors:  Werner Eisenbarth; Manfred Mackeben; Dorothe A Poggel; Hans Strasburger
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-09-19       Impact factor: 3.117

10.  Measuring contrast sensitivity in normal subjects with OPTEC 6500: influence of age and glare.

Authors:  Bettina Hohberger; Robert Laemmer; Werner Adler; Anselm G M Juenemann; Folkert K Horn
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-08-11       Impact factor: 3.117

View more
  3 in total

Review 1.  [Contrast vision-definitions, conversions, and equivalence tables].

Authors:  M Bach; M B Hoffmann; H Jägle; S P Heinrich; U Schiefer; W Wesemann
Journal:  Ophthalmologe       Date:  2017-04       Impact factor: 1.059

2.  Development of a Method to Potentially Substitute Direct Evaluation of Mesopic Visual Acuity in Drivers.

Authors:  Marta Garcia-Rojo; Cristina Bonnin-Arias; Eva Chamorro; Cristina Alvarez-Peregrina; Celia Sanchez-Ramos
Journal:  Int J Environ Res Public Health       Date:  2021-04-29       Impact factor: 3.390

3.  The Influence of Protective Headgear on the Visual Field of Recreational-Level Skiers.

Authors:  Mateja Očić; Ivan Bon; Lana Ružić; Vjekoslav Cigrovski; Tomislav Rupčić
Journal:  Int J Environ Res Public Health       Date:  2022-08-25       Impact factor: 4.614

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.