T Togawa1.
Abstract
A method of estimating skin emissivity based on reflectance measurement upon transient stepwise change in the ambient radiation temperature was proposed. To effect this change, two shades at different temperatures were switched mechanically, and the change in radiation from the skin surface was recorded through an aperture for each shade by a high-resolution, fast-response radiometer having a sensitivity within the 8-14 microns range. Measurements were made on the forehead, forearm, palm and back of the hand in 10 male and 10 female subjects. No significant differences in emissivity were observed among sites and between sexes. The overall average of the skin emissivity obtained was 0.971 +/- 0.005 (SD). This result is inconsistent with most reported skin emissivity values. However, as the former studies had many inherent inadequacies, both theoretical and experimental, it is considered that most of these reported skin emissivities are unacceptable. The method proposed in this study has the following advantages: (1) relative calibration between instruments is unnecessary, (2) non-contact measurement can be achieved, and (3) each measurement can be made within one minute.Mesh:
Year: 1989 PMID: 2714059 DOI: 10.1088/0143-0815/10/1/004
Source DB: PubMed Journal: Clin Phys Physiol Meas ISSN: 0143-0815