Evaluation of tablet ultrasound for routine abdominal interventional procedures.

Anna Maria Ierardi1, Federico Fontana1, Francesca Giorlando1, Giuseppe De Marchi1, Antonio Pinto2, Alessandro Radaelli3, Stephanie Schampaert3, Massimo Tonolini4, Raffaele Novario5, Gianpaolo Carrafiello6.   

Abstract

AIM: The aim of the study was to establish if a novel tablet ultrasound (US) may replace a high-end US machine during routine interventional radiology activities.
MATERIALS AND METHODS: Thirty consecutive patients were evaluated by two operators comparing the performance of the new US tablet system (VISIQ, Philips Healthcare) against a high-end US system (iU22, Philips Healthcare) using a curved probe (C5-2). A structured questionnaire was used to rank on a 4-point scale the ability of each system to locate a target as detected by previous examinations and visualize needles and path during an interventional procedure. Necessity for conversion from the tablet US to the high-end US system was registered; body mass index (BMI) was annotated for each patient.
RESULTS: Agreement between the operators was found for every patient. Mean patient BMI was 25 (range 17-34). Image quality of the tablet US was considered insufficient in 1 case to visualize the target and in another case to visualize the needle. Mean target image quality was superior with the high-end US system, while needle visibility scored higher with the tablet US. Conversion to the high-end system was registered in 40 % of cases: in most cases (66.6 %) the decision was due to the absence of a dedicated needle guide for the tablet US.
CONCLUSIONS: The novel tablet US was found to provide sufficient image quality for the majority of routine interventional procedures. Dedicated accessories and additional experience with this new generation US device may be needed to replace bulky high-end US systems.

Entities:  

Keywords:  Interventional radiology; Tablet; Ultrasound

Mesh:

Year:  2016        PMID: 27138011     DOI: 10.1007/s11547-016-0641-6

Source DB:  PubMed          Journal:  Radiol Med        ISSN: 0033-8362            Impact factor:   3.469


  16 in total

1.  Real-time B-mode ultrasound quality control test procedures. Report of AAPM Ultrasound Task Group No. 1.

Authors:  M M Goodsitt; P L Carson; S Witt; D L Hykes; J M Kofler
Journal:  Med Phys       Date:  1998-08       Impact factor: 4.071

2.  Detection of therapeutically relevant diagnoses made by sonography of the upper abdomen: portable versus high-end sonographic units -- a prospective study.

Authors:  C M Ziegler; K Seitz; U Leicht-Biener; M Mauch
Journal:  Ultraschall Med       Date:  2004-12       Impact factor: 6.548

Review 3.  Pocket-sized versus standard ultrasound machines in abdominal imaging.

Authors:  K H Tse; W H Luk; M C Lam
Journal:  Singapore Med J       Date:  2014-06       Impact factor: 1.858

4.  A concordance correlation coefficient to evaluate reproducibility.

Authors:  L I Lin
Journal:  Biometrics       Date:  1989-03       Impact factor: 2.571

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Comparison of a pocket-size ultrasound device with a premium ultrasound machine: diagnostic value and time required in bedside ultrasound examination.

Authors:  Konrad Friedrich Stock; Bettina Klein; Dominik Steubl; Christian Lersch; Uwe Heemann; Stefan Wagenpfeil; Florian Eyer; Dir-Andre Clevert
Journal:  Abdom Imaging       Date:  2015-10

7.  Improving Echo-Guided Procedures Using an Ultrasound-CT Image Fusion System.

Authors:  Michele Diana; Peter Halvax; Damien Mertz; Andras Legner; Jean-Marcel Brulé; Eric Robinet; Didier Mutter; Patrick Pessaux; Jacques Marescaux
Journal:  Surg Innov       Date:  2015-03-22       Impact factor: 2.058

8.  Transluminal coil embolization of an inferior gluteal artery aneurysm by ultrasound-guided direct puncture of the target vessel.

Authors:  A Kuzuya; K Fujimoto; S Iyomasa; M Matsuda
Journal:  Eur J Vasc Endovasc Surg       Date:  2005-08       Impact factor: 7.069

9.  Outcomes after arterial endovascular procedures performed in patients with an elevated international normalized ratio.

Authors:  Joshua A Wilensky; Ahsan T Ali; Mohammed M Moursi; Guillermo A Escobar; Matthew R Smeds
Journal:  Ann Vasc Surg       Date:  2014-06-12       Impact factor: 1.466

Review 10.  New applications of ultrasonography: interventional ultrasound.

Authors:  L Solbiati
Journal:  Eur J Radiol       Date:  1998-05       Impact factor: 3.528

View more
  1 in total

Review 1.  Traumatic fractures in adults: missed diagnosis on plain radiographs in the Emergency Department.

Authors:  Antonio Pinto; Daniela Berritto; Anna Russo; Federica Riccitiello; Martina Caruso; Maria Paola Belfiore; Vito Roberto Papapietro; Marina Carotti; Fabio Pinto; Andrea Giovagnoni; Luigia Romano; Roberto Grassi
Journal:  Acta Biomed       Date:  2018-01-19
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.