Visual grading evaluation of commercially available metal artefact reduction techniques in hip prosthesis computed tomography.

Karin M Andersson1,2, Eva Norrman1, Håkan Geijer2,3, Wolfgang Krauss2,3, Yang Cao4,5, Johan Jendeberg2,3, Mats Geijer3,6, Mats Lidén3, Per Thunberg1,2.   

Abstract

OBJECTIVE: To evaluate metal artefact reduction (MAR) techniques from four CT vendors in hip prosthesis imaging.
METHODS: Bilateral hip prosthesis phantom images, obtained by using MAR algorithms for single-energy CT data or dual-energy CT (DECT) data and by monoenergetic reconstructions of DECT data, were visually graded by five radiologists using 10 image quality criteria. Comparisons between the MAR images and a reference image were performed for each scanner separately. Ordinal probit regression analysis was used.
RESULTS: The MAR algorithms in general improved the image quality based on the majority of the criteria (up to between 8/10 and 10/10) with a statistical improvement in overall image quality (p < 0.001). However, degradation of image quality, such as new artefacts, was seen in some cases. A few monoenergetic reconstruction series improved the image quality (p < 0.004) for one of the DECT scanners, but it was only improved for some of the criteria (up to 5/10). Monoenergetic reconstructions resulted in worse image quality for the majority of the criteria (up to 7/10) for the other DECT scanner.
CONCLUSION: The MAR algorithms improved the image quality of the hip prosthesis CT images. However, since additional artefacts and degradation of image quality were seen in some cases, all algorithms should be carefully evaluated for every clinical situation. Monoenergetic reconstructions were in general concluded to be insufficient for reducing metal artefacts. ADVANCES IN KNOWLEDGE: Qualitative evaluation of the usefulness of several MAR techniques from different vendors in CT imaging of hip prosthesis.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27123700     DOI: 10.1259/bjr.20150993

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  6 in total

1.  Metal Artifact Reduction in Head CT Performed for Patients with Deep Brain Stimulation Devices: Effectiveness of a Single-Energy Metal Artifact Reduction Algorithm.

Authors:  Y Nagayama; S Tanoue; S Oda; D Sakabe; T Emoto; M Kidoh; H Uetani; A Sasao; T Nakaura; O Ikeda; K Yamada; Y Yamashita
Journal:  AJNR Am J Neuroradiol       Date:  2019-12-26       Impact factor: 3.825

2.  Metal implants on abdominal CT: does split-filter dual-energy CT provide additional value over iterative metal artifact reduction?

Authors:  Hildegard M Wichtmann; Kai R Laukamp; Sebastian Manneck; Konrad Appelt; Bram Stieltjes; Daniel T Boll; Matthias R Benz; Markus M Obmann
Journal:  Abdom Radiol (NY)       Date:  2022-09-30

3.  Single-energy metal artifact reduction technique for reducing metallic coil artifacts on post-interventional cerebral CT and CT angiography.

Authors:  Masaki Katsura; Jiro Sato; Masaaki Akahane; Taku Tajima; Toshihiro Furuta; Harushi Mori; Osamu Abe
Journal:  Neuroradiology       Date:  2018-08-24       Impact factor: 2.804

4.  Comparison of quantitative measurements of four manufacturer's metal artifact reduction techniques for CT imaging with a self-made acrylic phantom.

Authors:  Ryan Chou; Hung-Yi Chi; Yi-Hung Lin; Liu-Kuo Ying; Yu-Ju Chao; Cheng-Hsun Lin
Journal:  Technol Health Care       Date:  2020       Impact factor: 1.285

5.  Reduction of metal artifacts from knee tumor prostheses on CT images: value of the single energy metal artifact reduction (SEMAR) algorithm.

Authors:  Fang-Ling Zhang; Ruo-Cheng Li; Xiao-Ling Zhang; Zhao-Hui Zhang; Ling Ma; Lei Ding
Journal:  BMC Cancer       Date:  2021-12-02       Impact factor: 4.430

Review 6.  Improving radiation physics, tumor visualisation, and treatment quantification in radiotherapy with spectral or dual-energy CT.

Authors:  Matthijs Ferdinand Kruis
Journal:  J Appl Clin Med Phys       Date:  2021-11-07       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.