Thiago Saads Carvalho1, Tommy Baumann1, Adrian Lussi1.
Abstract
In the present study, the surface reflection intensity (SRI) was measured from enamel with different induced erosion degrees using a hand-held pen-size reflectometer (Hand-Held) and a Table-Top reflectometer. To validate the Hand-Held reflectometer, we correlated its optical signals with the change of surface microhardness (SMH), and amount of calcium released from the enamel samples during erosion. We used 124 tooth enamel specimens that were exposed to an erosive challenge of either 1, 2, 4, 6, 8, or 10 minutes. SRI and SMH were measured before and after the erosive challenge and we also measured the amount of calcium released to the citric acid. Relative SRI loss (rSRIloss) and relative SMH loss (rSMHloss) were calculated. rSRIloss from the Hand-Held and the Table-Top reflectometers were similar and significantly correlated to rSMHloss and calcium release. The regression analyses showed a significant association between rSRIloss from both reflectometers and rSMHloss and calcium, showing that both reflectometers can be used to measure erosive demineralization of enamel. The Hand-Held reflectometer is capable of assessing in vitro erosion, correlating to other commonly used methods. It is small, easy to handle and provides fast measurement, being a possible candidate to measure erosion in clinical studies.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27121129 PMCID: PMC4848505 DOI: 10.1038/srep25259
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Summary of the Spearman’s correlation coefficient values (rs) and their respective significances (p-values), for the association between the different methods used to measure erosion.
| rSRIloss Table-Top | rSMHloss | Calcium Release | |
|---|---|---|---|
| rSRIloss Hand Held | 0.81 p < 0.001 | 0.79 p < 0.001 | 0.70 p < 0.001 |
| rSMHloss | 0.77 p < 0.001 | – | – |
| Calcium Release | 0.69 p < 0.001 | – | – |
Figure 1Relative surface reflection intensity loss (rSRI) from the Table-Top (1a) and Hand-Held (1b) devices after different erosion times.
Figure 2Relative surface microhardness loss (rSMH; 2a) and calcium released from the specimens (2b) after different erosion times.
Figure 3Association between the relative surface reflection intensity loss (rSRI) from the Table-Top (◻) and Hand-Held (△) devices with relative surface microhardness loss (3a) and calcium (after loge-transformation) released from the specimens (3b).
Solid line: regression for Table-Top device; dashed line: regression for Hand-Held device.