Reduce Driveline Trauma Through Stabilization and Exit Site Management: 30 Days Feasibility Results from the Multicenter RESIST Study.

Marcia Stahovich1, Kartik S Sundareswaran, Sarah Fox, William Hallinan, Peggy Blood, Leway Chen, Salpy V Pamboukian, Raymond Chinn, David J Farrar, Francis D Pagani, Laura Blue.   

Abstract

The percutaneous lead management kit (PLMK) was developed for the HeartMate 2 (HM2) left ventricular assist device (LVAD) to reduce trauma at the exit site and to maintain a clean environment. REduce Driveline Trauma through StabIlization and Exit Site ManagemenT (RESIST) was a multicenter, prospective, nonrandomized study designed to evaluate the feasibility of the PLMK for managing the HM2 driveline exit site. Fifty patients were enrolled at five sites at a median of 495 days post-HM2 implant; 92% (46 of 50) of patients used the PLMK for a minimum of 30 days. At 30 days, more patients found the PLMK to be extremely comfortable (80% vs. 37%, p < 0.001) and extremely effective at stabilizing the driveline (82% vs. 40%, p < 0.001) compared with each center's standard of care. Frequency of dressing changes was 6-7 days or higher for 85% of the patients with PLMK. Three patients developed driveline infection while on PLMK (6%, 0.15 events per patient year), and 35 patients continued to use the PLMK after 6 months. The PLMK is easy to use, increases patient comfort, and increases driveline stability with a dressing change frequency of 6-7 days.

Entities:  

Mesh:

Year:  2016        PMID: 27111738     DOI: 10.1097/MAT.0000000000000374

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  4 in total

Review 1.  Non-patient factors associated with infections in LVAD recipients: A scoping review.

Authors:  Supriya Shore; Michael J Pienta; Tessa M F Watt; Gardner Yost; Whitney A Townsend; Lourdes Cabrera; Michael D Fetters; Carol Chenoweth; Keith D Aaronson; Francis D Pagani; Donald S Likosky
Journal:  J Heart Lung Transplant       Date:  2021-10-22       Impact factor: 13.569

Review 2.  Wound care of the driveline exit site in patients with a ventricular assist device: A systematic review.

Authors:  Zeliha Özdemir; Sevilay Şenol Çelik
Journal:  Turk Gogus Kalp Damar Cerrahisi Derg       Date:  2018-04-30       Impact factor: 0.332

3.  Driveline Features as Risk Factor for Infection in Left Ventricular Assist Devices: Meta-Analysis and Experimental Tests.

Authors:  Melanie Kranzl; Martin Stoiber; Anne-Kristin Schaefer; Julia Riebandt; Dominik Wiedemann; Christiane Marko; Günther Laufer; Daniel Zimpfer; Heinrich Schima; Thomas Schlöglhofer
Journal:  Front Cardiovasc Med       Date:  2021-12-16

4.  Driveline exit-site care protocols in patients with left ventricular assist devices: a systematic review.

Authors:  Zeliha Ozdemir Koken; Yunus C Yalcin; Diana van Netten; Chantal C de Bakker; Maaike van der Graaf; Umit Kervan; Nelianne J Verkaik; Kadir Caliskan
Journal:  Eur J Cardiothorac Surg       Date:  2021-09-11       Impact factor: 4.191