Comment on "Argon I Lines Produced in a Hollow Cathode Source, 332 nm to 5865 nm".

Craig J Sansonetti1.   

Abstract

Recent observations at the National Institute of Standards and Technology indicate that the Ar I wavenumbers reported by Whaling et al. [J. Res. Natl. Inst. Stand. Technol. 107, 149 (2002)] are systematically too large. To investigate the source of this problem, selected lines of Ar I and Ar II were remeasured in the same spectra used by Whaling et al. The measurements show that the Ar I wavenumbers of Whaling et al. are systematically shifted with respect to the Ar II wavenumbers previously reported by Whaling et al. [J. Quant. Spectrosc. Radiat. Transfer 53, 1 (1995)] based on the same spectra. The Ar I wavenumbers can be corrected by a constant multiplicative correction of 0.999 999 933.

Entities:  

Keywords:  Fourier transform spectroscopy; argon; atomic spectroscopy; wavelengths

Year:  2007        PMID: 27110473      PMCID: PMC4655922          DOI: 10.6028/jres.112.023

Source DB:  PubMed          Journal:  J Res Natl Inst Stand Technol        ISSN: 1044-677X


1. Introduction

In Ref. [1], Whaling, Anderson, Carle, Brault, and Zarem presented a comprehensive list of Ar I lines in the region 332 nm to 5865 nm as emitted from a hollow-cathode discharge with Ar buffer gas. The measurements were based on spectra retrieved from the archives of the 1 m Fourier transform spectrometer (FTS) of the National Solar Observatory (Kitt Peak). These spectra had previously been used by the same authors to measure lines of Ar II in this spectral region [2]. In recent work at the National Institute of Standards and Technology (NIST), 17 lines of Ar I ranging from 22163 cm−1 to 25315 cm−1 were observed in a Hg/Ar electrodeless discharge lamp (edl) with an FTS optimized for the ultraviolet [3]. The spectra were calibrated using precisely known lines of 198Hg [4]. For all 17 lines the NIST wavenumbers were smaller than the wavenumbers of Whaling et al. [1]. Pressure shifts due to differences in the Ar carrier gas pressure were investigated as a possible cause of the discrepancy but were not able to explain it. In a separate set of experiments, the infrared spectrum of a Th/Ar hollow cathode lamp was observed with the NIST 2 m Fourier transform spectrometer [5]. The spectra were calibrated by Th lines measured by laser optogalvanic spectroscopy [6]. Approximately 600 lines of Ar I were measured. When these results were compared to Ref. [1], it was apparent that they also were smaller on average than the wavenumbers of Whaling et al. [1]. For lines associated with energy levels of high angular momentum or high excitation, the deviations were large and scattered. For several hundred low excitation lines, however, the NIST observations were smaller by a consistent shift of approximately 1 part in 107. Based on these two sets of measurements in widely separated spectral regions, it appeared possible that the results of [1] might be high due to a systematic calibration error. We have used Ar II wavenumbers from Ref. [2] in several experiments and have found them to be consistent with other generally accepted wavelength standards. It seemed interesting, therefore, to try to determine whether the Ar I wavenumbers from Ref. [1] and Ar II wavenumbers from Ref. [2] were consistent with each other. To investigate this question we retrieved three of the spectra used for the Ar measurements — Ti #7 (8/15/92), Cu #2 (4/17/83), and Ti #2 (11/19/85) — from the Kitt Peak Archives [7]. The spectra were analyzed using the interactive program Xgremlin [8]. Each line of interest was fit with a Voigt profile convoluted with the FTS instrumental function to obtain the wavenumber. Ti #7 (8/15/92) covers the region of the NIST Hg/Ar edl observations. In this spectrum we measured the 28 Ar II lines recommended by Learner and Thorne for use as calibration lines [9]. From these lines we calculated a correction factor for the spectrum with an uncertainty of 2 parts in 109, as shown in Table 1. We then measured the 17 Ar I lines that appeared in the NIST Hg/Ar observations and corrected their wavenumbers using the factor from Table 1. The results are shown in Table 2, where they are compared with the values from Whaling et al. [1]. The values from [1] are larger on average by the multiplicative constant 1.000 000 065 7(24). The combined standard uncertainty is calculated as the sum in quadrature of the standard error of the mean, representing the random scatter of the data, and a systematic component that propagates directly from the standard error in the calibration factor.
Table 1

Determination of correction factor for the spectrum Ti #7 (8/15/92) from 28 Ar II lines.

Ar II Uncorrected (cm−1)Whaling Ar II [2] (cm−1)Correction Factor
19429.7645219429.76941.000 000 251
19749.3814919749.38731.000 000 294
19957.1607019957.16621.000 000 275
20106.3685020106.37431.000 000 288
20135.0409520135.04661.000 000 281
20265.1193720265.12491.000 000 273
20448.1894920448.19491.000 000 264
20486.6500520486.65561.000 000 271
20622.1064720622.11211.000 000 273
20801.4173320801.42301.000 000 273
20981.0838220981.08951.000 000 271
21109.3759321109.38171.000 000 273
21149.7350221149.74091.000 000 278
21462.8834921462.88921.000 000 266
21831.0400721831.04581.000 000 262
21995.7780021995.78381.000 000 264
22561.9488722561.95511.000 000 276
22566.0605722566.06671.000 000 272
22587.4116422587.41761.000 000 264
22715.7939322715.80011.000 000 272
22720.3865922720.39281.000 000 273
22826.3687222826.37511.000 000 279
22992.2746122992.28081.000 000 269
23081.7976023081.80431.000 000 290
23431.6686123431.67441.000 000 247
23644.2977623644.30371.000 000 251
25447.0014225447.00761.000 000 243
26806.9946526807.00201.000 000 274
Average1.000 000 270
Standard Error0.000 000 002
Table 2

Calibrated wavenumbers for selected lines of Ar I from the spectrum Ti #7 (8/15/92).

Ar I Corrected Ti #7 (8/15/92) (cm−1)Whaling Ar I [1] (cm−1)Deviation (cm−1)Ratio Whaling/Ti #7
22163.128822163.13030.00151.000 000 0689
23007.602123007.60350.00151.000 000 0630
23059.770323059.77180.00151.000 000 0637
23069.225423069.22690.00151.000 000 0650
23248.729523248.73080.00131.000 000 0559
23400.730223400.73180.00161.000 000 0667
23432.995223432.99670.00151.000 000 0657
23471.089523471.09110.00161.000 000 0675
23798.996323798.99790.00161.000 000 0661
23812.359723812.36120.00151.000 000 0647
23853.766223853.76770.00151.000 000 0650
23855.566123855.56770.00161.000 000 0667
23905.933223905.93480.00161.000 000 0655
24007.567424007.56910.00171.000 000 0693
24039.832124039.83370.00161.000 000 0672
24718.454924718.45660.00171.000 000 0695
25315.838625315.84030.00171.000 000 0673
Average1.000 000 0657
Standard Error0.000 000 0008
Uncertainty0.000 000 0024
Cu #2 (4/17/83) covers a portion of the infrared in which many Ar I lines appear in the NIST Th/Ar hollow-cathode observations. In this spectrum we chose 20 lines of Ar II with high signal-to-noise ratio and good line symmetry for use in determining the wavenumber correction factor. The results are shown in Table 3. The scatter is significantly larger than for the recommended lines used in the previous spectrum, and two of the lines produce correction factors that are clear outliers. We excluded these lines, shown in italics in Table 3, in calculating the average. We then measured 20 Ar I lines that had high signal-to-noise ratio, good symmetry, and were associated with states of low excitation and low angular momentum. The wavenumbers for these lines were corrected with the factor from Table 3. The results are shown in Table 4. Again the results of Whaling et al. [1] are systematically larger. Neither the correction factor nor the ratio in Table 4 is determined as precisely for this spectrum as for Ti #7 because the number of significant figures given in Ref. [2] becomes a limiting factor; however, to a good approximation the results of [1] are again larger by a constant factor.
Table 3

Determination of correction factor for the spectrum Cu #2 (4/17/83) from 20 Ar II lines. The two lines shown in italics were excluded from the average.

Ar II Uncorrected (cm−1)Whaling Ar II (cm−1)Correction Factor
3312.752063312.74640.999 998 293
3336.552913336.54740.999 998 347
3405.899683405.89450.999 998 478
3423.091923423.08620.999 998 330
3441.050693441.04520.999 998 404
3469.599593469.59400.999 998 389
3495.533823495.52800.999 998 336
3519.861943519.85610.999 998 340
3562.073843562.06780.999 998 304
3580.417553580.41130.999 998 256
3612.335883612.33000.999 998 373
3617.295403617.28930.999 998 314
3620.500533620.49460.999 998 362
3695.224063695.21810.999 998 386
3765.745833765.73870.999 998 107
3807.757993807.75170.999 998 348
3878.808193878.80180.999 998 354
3906.182813906.17610.999 998 282
4030.029284030.02010.999 997 723
4031.869394031.86270.999 998 340
Average0.999 998 346
Standard Error0.000 000 012
Table 4

Calibrated wavenumbers for selected lines of Ar I from the spectrum Cu #2 (4/17/83).

Ar I Corrected Cu #2 (4/17/83) (cm−1)Whaling Ar I [1] (cm−1)Deviation (cm−1)Ratio Whaling/Cu #2
3334.49693334.49710.00021.000 000 060
3415.22533415.22550.00021.000 000 059
3417.29653417.29670.00021.000 000 059
3435.42203435.42220.00021.000 000 058
3467.03723467.03730.00011.000 000 029
3504.05233504.05250.00021.000 000 057
3525.41353525.41370.00021.000 000 057
3573.36343573.36360.00021.000 000 056
3602.29583602.29600.00021.000 000 056
3619.68733619.68750.00021.000 000 055
3647.11373647.11390.00021.000 000 055
3663.91783663.91800.00021.000 000 055
3725.36413725.36420.00011.000 000 027
3757.62873757.62890.00021.000 000 053
3766.43873766.43890.00021.000 000 053
3828.06073828.06100.00031.000 000 078
3922.40073922.40090.00021.000 000 051
3978.97133978.97160.00031.000 000 075
4012.58954012.58970.00021.000 000 050
4171.34974171.34990.00021.000 000 048
Average1.000 000 0054
Standard Error0.000 000 0003
Uncertainty0.000 000 0012
Ti #2 (11/19/85) covers the near ultraviolet to near infrared region between the two sets of unpublished NIST observations. In this region 18 Ar II lines are recommended for use as wavelength standards [2]. One of these lines was too weak to measure, but the remaining 17 were measured and used with the published wavenumbers from [2] to determine values of the correction factor, as shown in Table 5. The two longest wavelength lines produced correction factors that are in poor agreement with the other lines, and the overall scatter of the data is larger than in the other spectra. Since some of the lines recommended as standards were weak, an additional 20 lines were selected and measured based solely on their high signal-to-noise ratio and good symmetry. These lines produced correction factors with about the same average but a standard deviation 1.6 times larger than the original set; consequently, the average correction factor was calculated from the recommended lines, omitting the two outliers from the average. We then measured 20 lines of Ar I that were selected using the same criteria as for Cu #2 (4/17/83). The wavenumbers were corrected using the correction factor from Table 5. The results are presented in Table 6. As in the other two spectra the results from [1] are systematically larger than the new measurements.
Table 5

Determination of correction factor for the spectrum Ti #2 (11/19/85) from 20 Ar II lines. The two lines shown in italics were excluded from the average.

Ar II Uncorrected (cm−1)Whaling Ar II (cm−1)Correction Factor
4776.240044776.23720.999 999 406
5833.125955833.12340.999 999 562
7167.215297167.21410.999 999 834
7540.049747540.04870.999 999 862
8123.068948123.06860.999 999 959
8204.995348204.99450.999 999 898
8494.170328494.16900.999 999 844
8918.397418918.39500.999 999 730
9152.116559152.11520.999 999 852
9231.542879231.54341.000 000 058
9245.683919245.68290.999 999 890
9551.056839551.05510.999 999 819
9887.832329887.83230.999 999 998
10091.7274810091.72580.999 999 834
10550.9191510550.91760.999 999 853
10773.2459210773.24390.999 999 812
11086.3944211086.39230.999 999 808
Average0.999 999 870
Standard Error0.000 000 021
Table 6

Calibrated wavenumbers for selected lines of Ar I from the spectrum Ti #2 (11/19/85).

Ar I Corrected Ti #2 (11/19/85) (cm−1)Whaling Ar I [1] (cm−1)Deviation (cm−1)Ratio Whaling/Cu #2
5929.54635929.54670.00041.000 000 065
6294.25426294.25470.00051.000 000 075
6521.65336521.65380.00051.000 000 073
6651.31436651.31470.00041.000 000 064
6858.03086858.03140.00061.000 000 081
7188.41107188.41150.00051.000 000 075
7338.70397338.70450.00061.000 000 084
7554.05907554.05970.00071.000 000 087
7910.17897910.17950.00061.000 000 076
8099.28378099.28430.00061.000 000 070
8403.43988403.44050.00071.000 000 089
8553.73268553.73330.00071.000 000 079
8717.87558717.87620.00071.000 000 084
8887.75928887.75990.00071.000 000 083
9178.23109178.23180.00081.000 000 084
9342.37399342.37450.00061.000 000 065
9541.16019541.16080.00071.000 000 068
9749.59399749.59470.00081.000 000 082
10217.441910217.44250.00061.000 000 057
10687.430510687.43090.00041.000 000 041
Average1.000 000 0074
Standard Error0.000 000 0003
Uncertainty0.000 000 0021
Based on this sample of lines from three different spectra, which cover a wide range of wavelengths from the ultraviolet to the infrared, it is clear that the Ar I wavenumber scale in Ref. [1] is not consistent with the Ar II wavenumber scale in Ref. [2], despite the fact that Ar I and Ar II were measured in the same spectra. For each of the three spectra the scales differ by a multiplicative constant. The constant is defined most precisely for the spectrum Ti #7 (8/15/92), but the values for all three spectra agree within their uncertainties. We have confidence that the wavenumbers for Ar II [2] are reliable. They have been widely used as standards for calibration of FTS spectra and have been shown to be on a consistent scale with independently determined standard lines of 198Hg I at a level of better than 2 parts in 108 [10]. Therefore, we conclude that there must be a systematic calibration error in the Ar I results of Ref. [1]. On an empirical basis the spectra we have investigated suggest that all of the Ar I data can be corrected by reducing the wavenumbers by a single multiplicative factor. This factor is well-determined from the results of Ti #7 (8/15/92) to be 0.999 999 934 3(24). The factors from Cu #2 (4/17/83) and Ti #2 (11/19/85), 0.999 999 945 5(121) and 0.999 999 925 9(213) respectively, have larger uncertainties but are fully consistent with this value. Most of the spectra used for Refs. [1] and [2] were originally calibrated with respect to Ar II lines of Norlén [11] and later shifted to a revised wavenumber scale based on molecular lines of CO [12] as described in [2]. It is striking that the ratio determined in the fourth column of Table 2 is in almost perfect agreement with the ratio between the wavenumber scales of Whaling et al. [2] and Norlén [11], which was determined in [2] to be 1.000 000 067(8). All of the Ar I wavenumbers in [1] are larger by this factor than their values in an unpublished list that was made available to several laboratories in the late 1990s [13]. This leads us to speculate that the systematic error in the published wavenumbers for Ar I [1] resulted from an inadvertent double correction of the wavenumber scale. We have communicated these results to Whaling, but he is not able to verify that a double correction actually occurred [14]. Our speculation is supported by a comparison of the deviations in the third columns of Tables 2, 4, and 6 for each line with the error that would result had the scale correction been applied twice. The average difference of these values for all 57 lines measured is just −0.000 01 cm−1. For 26 of the lines the difference is less than 0.000 05 cm−1, and the largest individual deviations are +0.000 19 cm−1 and −0.000 28 cm−1. These deviations are less than the reported uncertainty for all of the lines; thus the data are entirely consistent with a double correction of the scale.

2. Conclusion

Based on all of the evidence, we propose that the results of [1] be corrected by multiplying all wavenumbers by the factor 0.999 999 933. This factor is the inverse of the Norlén-to-CO scale correction. It is consistent with the corrections determined empirically from the spectra Ti #7 (8/15/92), Cu #2 (4/17/83), and Ti #2 (11/19/85). By making this correction, the Ar I wavenumbers of Ref. [1] will be put on the same scale as the Ar II wavenumbers of Ref. [2], which was derived from CO.
  1 in total

1.  Argon I Lines Produced in a Hollow Cathode Source, 332 nm to 5865 nm.

Authors:  W Whaling; W H C Anderson; M T Carle; J W Brault; H A Zarem
Journal:  J Res Natl Inst Stand Technol       Date:  2002-04-01
  1 in total
  1 in total

1.  Re-Optimized Energy Levels and Ritz Wavelengths of (198)Hg I.

Authors:  Alexander Kramida
Journal:  J Res Natl Inst Stand Technol       Date:  2011-04-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.