| Literature DB >> 27104163 |
Young Suk Kim1, Jaegi Lee2, Jong In Park2, Wonmo Sung2, Sol Min Lee1, Gwi Eon Kim1.
Abstract
PURPOSE: Radiotherapy of the neck is known to cause carotid artery stenosis. We compared the carotid artery dose received between volumetric modulated arc therapy (VMAT) and conventional fixed-field intensity-modulated radiotherapy (IMRT) plans in patients with early glottic cancer.Entities:
Keywords: Cancer; Carotid arteries; Glottis; Intensity-modulated radiotherapy; Volumetric modulated arc therapy
Year: 2016 PMID: 27104163 PMCID: PMC4831965 DOI: 10.3857/roj.2016.34.1.18
Source DB: PubMed Journal: Radiat Oncol J ISSN: 2234-1900
The dose constraint for IMRT and VMAT
IMRT, intensity-modulated radiotherapy; VMAT, volumetric modulated arc therapy; PTV, planning target volume; PRV, planning organ at risk volume.
a)A 0.3-cm isotropic margin of the PTV was subtracted from the PRV of the carotid arteries. b)A 0.3-cm isotropic margin of the PTV was subtracted from the thyroid gland.
Comparison of average dosimetric values by planning technique (n = 21)
VMAT, volumetric modulated arc therapy; IMRT, intensity-modulated radiotherapy; 3DCRT, 3-dimensional conformal radiotherapy; LPRT, lateral parallel-opposed photon field radiotherapy; PTV, planning target volume; D5, dose received by at least 5% volume of the planning target volume; D95, dose received by at least 95% volume of the planning target volume; Dmean, the mean dose to the planning target volume; V35Gy, volume irradiated at least 35 Gy on the carotid arteries; V50Gy, volume irradiated at least 50 Gy on the carotid arteries.
a)Homogeneity index = (D5 – D95) / Dmean. b)The values are presented as percent volume of the parameter.
Fig. 1Box and Whisker plot of the mean dose to both carotid arteries for each treatment planning technique. The horizontal line indicates the median, the box covers the 25%ile to 75%ile, and the upper and lower bars are the maximum and minimum values, respectively. VMAT, volumetric modulated arc therapy; IMRT, intensity-modulated radiotherapy; 3DCRT, 3-dimensional conformal radiotherapy; LPRT, lateral parallel-opposed photon field radiotherapy.
Fig. 2Isodose curves on an axial slice for a representative case planned with (A) volumetric modulated arc therapy, (B) intensity-modulated radiotherapy, (C) 3-dimensional conformal radiotherapy, and (D) lateral parallel-opposed photon field radiotherapy. The carotid arteries were delineated in yellow line.
Fig. 3Dose-volume histogram of the planning target volume (PTV) and the carotid arteries (CA) for the four treatment modalities in a representative case. VMAT, volumetric modulated arc therapy; IMRT, intensity-modulated radiotherapy; 3DCRT, 3-dimensional conformal radiotherapy; LPRT, lateral parallel-opposed photon field radiotherapy.