The Accuracy of 6 Inexpensive Pulse Oximeters Not Cleared by the Food and Drug Administration: The Possible Global Public Health Implications.

Michael S Lipnick1, John R Feiner, Paul Au, Michael Bernstein, Philip E Bickler.   

Abstract

BACKGROUND: Universal access to pulse oximetry worldwide is often limited by cost and has substantial public health consequences. Low-cost pulse oximeters have become increasingly available with limited regulatory agency oversight. The accuracy of these devices often has not been validated, raising questions about performance.
METHODS: The accuracy of 6 low-cost finger pulse oximeters during stable arterial oxygen saturations (SaO2) between 70% and 100% was evaluated in 22 healthy subjects. Oximeters tested were the Contec CMS50DL, Beijing Choice C20, Beijing Choice MD300C23, Starhealth SH-A3, Jumper FPD-500A, and Atlantean SB100 II. Inspired oxygen, nitrogen, and carbon dioxide partial pressures were monitored and adjusted via a partial rebreathing circuit to achieve 10 to 12 stable target SaO2 plateaus between 70% and 100% and PaCO2 values of 35 to 45 mm Hg. Comparisons of pulse oximeter readings (SpO2) with arterial SaO2 (by Radiometer ABL90 and OSM3) were used to calculate bias (SpO2 - SaO2) mean, precision (SD of the bias), and root mean square error (ARMS).
RESULTS: Pulse oximeter readings corresponding to 536 blood samples were analyzed. Four of the 6 oximeters tested showed large errors (up to -6.30% mean bias, precision 4.30%, 7.53 ARMS) in estimating saturation when SaO2 was reduced <80%, and half of the oximeters demonstrated large errors when estimating saturations between 80% and 90%. Two of the pulse oximeters tested (Contec CMS50DL and Beijing Choice C20) demonstrated ARMS of <3% at SaO2 between 70% and 100%, thereby meeting International Organization for Standardization (ISO) criteria for accuracy.
CONCLUSIONS: Many low-cost pulse oximeters sold to consumers demonstrate highly inaccurate readings. Unexpectedly, the accuracy of some low-cost pulse oximeters tested here performed similarly to more expensive, ISO-cleared units when measuring hypoxia in healthy subjects. None of those tested here met World Federation of Societies of Anaesthesiologists standards, and the ideal testing conditions do not necessarily translate these findings to the clinical setting. Nonetheless, further development of accurate, low-cost oximeters for use in clinical practice is feasible and, if pursued, could improve access to safe care, especially in low-income countries.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27089002     DOI: 10.1213/ANE.0000000000001300

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  29 in total

1.  Continuous monitoring of cerebrovascular reactivity through pulse transit time and intracranial pressure.

Authors:  Xiuyun Liu; Kais Gadhoumi; Ran Xiao; Nate Tran; Peter Smielewski; Marek Czosnyka; Steven W Hetts; Nerissa Ko; Xiao Hu
Journal:  Physiol Meas       Date:  2019-01-23       Impact factor: 2.833

2.  Clinical Interpretation of Peripheral Pulse Oximeters Labeled "Not for Medical Use".

Authors:  Arlene J Hudson; John Benjamin; Timothy Jardeleza; Curt Bergstrom; William Cronin; Mario Mendoza; Lex Schultheis
Journal:  Ann Fam Med       Date:  2018-11       Impact factor: 5.166

3.  Wrist-Sensor Pulse Oximeter Enables Prolonged Patient Monitoring in Chronic Lung Diseases.

Authors:  Alexander Guber; Gali Epstein Shochet; Sarah Kohn; David Shitrit
Journal:  J Med Syst       Date:  2019-06-14       Impact factor: 4.460

4.  Pulse oximetry in paediatric primary care in low-income and middle-income countries.

Authors:  Eric D McCollum; Carina King; Tim Colbourn; Hamish Graham; Mike Bernstein; Iain H Wilson; William Checkley
Journal:  Lancet Respir Med       Date:  2019-10-17       Impact factor: 30.700

5.  Pulse oximetry in primary care: factors affecting accuracy and interpretation.

Authors:  Paul Silverston; Marco Ferrari; Valentina Quaresima
Journal:  Br J Gen Pract       Date:  2022-02-24       Impact factor: 5.386

6.  Heart-rate tuned comb filters for processing photoplethysmogram (PPG) signals in pulse oximetry.

Authors:  Ludvik Alkhoury; Ji-Won Choi; Chizhong Wang; Arjun Rajasekar; Sayandeep Acharya; Sean Mahoney; Barry S Shender; Leonid Hrebien; Moshe Kam
Journal:  J Clin Monit Comput       Date:  2020-06-17       Impact factor: 2.502

Review 7.  The relationship between intermittent hypoxemia events and neural outcomes in neonates.

Authors:  Juliann M Di Fiore; Thomas M Raffay
Journal:  Exp Neurol       Date:  2021-05-10       Impact factor: 5.620

8.  Effective deployment of technology-supported management of chronic respiratory conditions: a call for stakeholder engagement.

Authors:  Richard W Costello; Alexandra L Dima; Dermot Ryan; R Andrew McIvor; Kay Boycott; Alison Chisholm; David Price; John D Blakey
Journal:  Pragmat Obs Res       Date:  2017-07-05

9.  Improving oxygen therapy for children and neonates in secondary hospitals in Nigeria: study protocol for a stepped-wedge cluster randomised trial.

Authors:  Hamish R Graham; Adejumoke I Ayede; Ayobami A Bakare; Oladapo B Oyewole; David Peel; Amy Gray; Barbara McPake; Eleanor Neal; Shamim Qazi; Rasa Izadnegahdar; Adegoke G Falade; Trevor Duke
Journal:  Trials       Date:  2017-10-27       Impact factor: 2.279

10.  Acute respiratory illness among a prospective cohort of pediatric patients using emergency medical services in India: Demographic and prehospital clinical predictors of mortality.

Authors:  Corey B Bills; Jennifer A Newberry; G V Ramana Rao; Loretta W Matheson; Srinivasa Rao; Swaminatha V Mahadevan; Matthew C Strehlow
Journal:  PLoS One       Date:  2020-04-02       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.