An X-ray fluorescence microscopic analysis of the tissue surrounding the multi-channel cochlear implant electrode array.

Kathryn Spiers1, Tina Cardamone2, John B Furness2,3, Jonathan C M Clark4, James F Patrick5, Graeme M Clark6.   

Abstract

OBJECTIVES: The aim of this study was to analyse the tissue surrounding the University of Melbourne's (UOMs) multi-channel cochlear implant electrode array and cochlear limited replacements, after long-term implantations. In particular, it aimed to identify the particulate material in the fibrous tissue capsule of the arrays implanted in 1978, 1983, and 1998, by using the Australian Synchrotron for X-ray fluorescence microscopy (XFM) to reveal the characteristic spectrum of metal, in particular platinum. This also helped to determine its format and chemical state. Tissue was retrieved following the recipient's death in 2007.
METHODS: Tissue was fixed and sections taken across the UOM and Cochlear Corporation (CI-22 and CI-24) electrode tracks. These were stained with Masson's trichrome. The Australian Synchrotron enabled XFM to accurately identify platinum from its characteristic fluorescence spectrum.
RESULTS: There was a fibrous tissue capsule (about 100-µm thick) and small regions of calcification around the UOM and CI-22 arrays, but a thinner capsule (40-60-µm thick) around CI-24, and a greater degree of calcification. Dark particulate matter was observed within macrophages and especially in fibrous tissue in proximity to the UOM and CI-22 arrays. This was identified as platinum using X-ray fluorescence. There was also diffusion of platinum into the tissue surrounding the UOM and CI-22 electrodes and fine particles had penetrated the spiral ligament. DISCUSSION: The larger particulate matter in the tissue around the UOM and CI-22 arrays suggested that it had flaked off in the manufacturing of the UOM electrodes. The more diffuse spread of platinum in the tissue around the UOM and CI-22 electrodes was likely due to electrolysis, probably from charge imbalance with the bipolar pulses from the UOM implant. This did not occur with the Cochlear CI-24 device. Furthermore, the widespread fine particles of platinum could have also been due to corrosion, especially from the UOM electrodes.

Entities:  

Keywords:  Cochlear implants; Electrodes; Histopathology; Platinum

Mesh:

Substances:

Year:  2016        PMID: 27078517     DOI: 10.1080/14670100.2016.1157943

Source DB:  PubMed          Journal:  Cochlear Implants Int        ISSN: 1467-0100


  9 in total

1.  A Materials Roadmap to Functional Neural Interface Design.

Authors:  Steven M Wellman; James R Eles; Kip A Ludwig; John P Seymour; Nicholas J Michelson; William E McFadden; Alberto L Vazquez; Takashi D Y Kozai
Journal:  Adv Funct Mater       Date:  2017-07-19       Impact factor: 18.808

2.  The development of neural stimulators: a review of preclinical safety and efficacy studies.

Authors:  Robert K Shepherd; Joel Villalobos; Owen Burns; David A X Nayagam
Journal:  J Neural Eng       Date:  2018-05-14       Impact factor: 5.379

3.  Electrochemistry of a Robust Neural Interface.

Authors:  Pavel A Takmakov
Journal:  Electrochem Soc Interface       Date:  2017

4.  Chronic intracochlear electrical stimulation at high charge densities results in platinum dissolution but not neural loss or functional changes in vivo.

Authors:  Robert K Shepherd; Paul M Carter; Ya Lang Enke; Andrew K Wise; James B Fallon
Journal:  J Neural Eng       Date:  2018-12-05       Impact factor: 5.379

5.  Platinum dissolution and tissue response following long-term electrical stimulation at high charge densities.

Authors:  Robert K Shepherd; Paul M Carter; Ashley N Dalrymple; Ya Lang Enke; Andrew K Wise; Trung Nguyen; James Firth; Alex Thompson; James B Fallon
Journal:  J Neural Eng       Date:  2021-03-17       Impact factor: 5.379

6.  Non-clinical and Pre-clinical Testing to Demonstrate Safety of the Barostim Neo Electrode for Activation of Carotid Baroreceptors in Chronic Human Implants.

Authors:  Seth J Wilks; Seth A Hara; Erika K Ross; Evan N Nicolai; Paul A Pignato; Adam W Cates; Kip A Ludwig
Journal:  Front Neurosci       Date:  2017-08-02       Impact factor: 4.677

Review 7.  Spectroscopic Methods Used in Implant Material Studies.

Authors:  Sławomir Lach; Przemysław Jurczak; Natalia Karska; Agnieszka Kubiś; Aneta Szymańska; Sylwia Rodziewicz-Motowidło
Journal:  Molecules       Date:  2020-01-29       Impact factor: 4.411

8.  Chronic intracochlear electrical stimulation at high charge densities: reducing platinum dissolution.

Authors:  Robert K Shepherd; Paul M Carter; Ya Lang Enke; Alex Thompson; Brianna Flynn; Ella P Trang; Ashley N Dalrymple; James B Fallon
Journal:  J Neural Eng       Date:  2020-10-08       Impact factor: 5.379

9.  Platinum corrosion products from electrode contacts of human cochlear implants induce cell death in cell culture models.

Authors:  Kirsten Wissel; Gudrun Brandes; Nils Pütz; Gian Luigi Angrisani; Jan Thieleke; Thomas Lenarz; Martin Durisin
Journal:  PLoS One       Date:  2018-05-15       Impact factor: 3.240

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.