Francois Beloncle1,2, Evangelia Akoumianaki3, Nuttapol Rittayamai1,4, Aissam Lyazidi5, Laurent Brochard6,7. 1. Keenan Research Centre and Li Ka Shing Knowledge Institute, St. Michael's Hospital, 30 Bond St, Toronto, ON, M5B 1W8, Canada. 2. Medical Intensive Care Unit, Hospital of Angers, University of Angers, Angers, France. 3. Department of Intensive Care Medicine, University Hospital of Heraklion, Crete, Greece. 4. Division of Respiratory Diseases and Tuberculosis, Department of Medicine, Faculty of Medicine Siriraj Hospital, Bangkok, Thailand. 5. Institut Supérieur des Sciences de la Santé, Université Hassan 1er, Settat, Morocco. 6. Keenan Research Centre and Li Ka Shing Knowledge Institute, St. Michael's Hospital, 30 Bond St, Toronto, ON, M5B 1W8, Canada. brochardl@smh.ca. 7. Interdepartmental Division of Critical Care Medicine, University of Toronto, Toronto, ON, Canada. brochardl@smh.ca.
Abstract
BACKGROUND: Proportional assist ventilation+ (PAV+) delivers airway pressure (P aw) in proportion to patient effort (P mus) by using the equation of motion of the respiratory system. PAV+ calculates automatically respiratory mechanics (elastance and resistance); the work of breathing (WOB) is estimated by the ventilator. The accuracy of P mus estimation and hence accuracy of the delivered P aw and WOB calculation have not been assessed. This study aimed at assessing the accuracy of delivered P aw and calculated WOB by PAV+ and examining the factors influencing this accuracy. METHODS: Using an active lung model with different respiratory mechanics, we compared (1) the actual delivered P aw by the ventilator to the theoretical P aw as defined by the equation of motion and (2) the WOB value displayed by the ventilator to the WOB measured from a Campbell diagram. RESULTS: Irrespective of respiratory mechanics and gain, the ventilator provided a P aw approximately 25 % lower than expected. This underassistance was greatest at the beginning of the inspiration. Intrinsic PEEP (PEEPi), associated with an increase in trigger delay, was a major factor affecting PAV+ accuracy. The absolute value of total WOB displayed by the ventilator was underestimated, but the changes in WOB were accurately detected by the ventilator. CONCLUSION: The assistance provided by PAV+ well follows P mus but with a constant underassistance. This is associated with an underestimation by the ventilator of the WOB. PEEPi can be a major factor contributing to PAV+ inaccuracy. Clinical recommendations should include using a high trigger sensitivity and a careful PEEP titration.
BACKGROUND: Proportional assist ventilation+ (PAV+) delivers airway pressure (P aw) in proportion to patient effort (P mus) by using the equation of motion of the respiratory system. PAV+ calculates automatically respiratory mechanics (elastance and resistance); the work of breathing (WOB) is estimated by the ventilator. The accuracy of P mus estimation and hence accuracy of the delivered P aw and WOB calculation have not been assessed. This study aimed at assessing the accuracy of delivered P aw and calculated WOB by PAV+ and examining the factors influencing this accuracy. METHODS: Using an active lung model with different respiratory mechanics, we compared (1) the actual delivered P aw by the ventilator to the theoretical P aw as defined by the equation of motion and (2) the WOB value displayed by the ventilator to the WOB measured from a Campbell diagram. RESULTS: Irrespective of respiratory mechanics and gain, the ventilator provided a P aw approximately 25 % lower than expected. This underassistance was greatest at the beginning of the inspiration. Intrinsic PEEP (PEEPi), associated with an increase in trigger delay, was a major factor affecting PAV+ accuracy. The absolute value of total WOB displayed by the ventilator was underestimated, but the changes in WOB were accurately detected by the ventilator. CONCLUSION: The assistance provided by PAV+ well follows P mus but with a constant underassistance. This is associated with an underestimation by the ventilator of the WOB. PEEPi can be a major factor contributing to PAV+ inaccuracy. Clinical recommendations should include using a high trigger sensitivity and a careful PEEP titration.
Proportional assist ventilation (PAV), described by Younes [1], was the first ventilator mode that introduced the concept of ‘patient-driven’ ventilation: a patient’s effort could influence not only the timing but also the magnitude of the ventilator assistance. This two-way interaction aimed to bypass numerous disadvantages linked to conventional assisted ventilation: patient–ventilator asynchrony, lack of adaptability to changing ventilator demands and loss of normal breathing variability. In critically ill mechanically ventilated patients, PAV has been shown to preserve breathing variability, improve comfort, decrease work of breathing (WOB) and patient–ventilator interaction with the potential to reduce the duration of controlled mechanical ventilation [2-6].Through automatic calculation of the respiratory system elastance and resistance and using the respiratory system equation of motion, PAV with load adjustable gain factors (PAV+) is able to partition the ventilator (Paw) and the patient (Pmus) contribution to total pressure of the respiratory system (Ptot) [7-9]. PAV gain, selected by the physician, determines the partition between Pmus and Ptot. Thus, PAV+ is the only ventilator mode that calculates noninvasively, respiratory mechanics and WOB. This information would be extremely useful to evaluate the efficacy of assistance, to adjust ventilator settings and to assess patient’s respiratory status and take therapeutic decisions [10]. Carteaux et al. recently published a study where they used their own calculations of Pmus [10]. It was based on the assumption that Paw and Pmus were complementary according to the equation of motion and therefore that the ventilator accurately delivered PAV. Second, whether the values of WOB displayed by the ventilator could be used, instead of calculating Pmus, was not known. Third, we wanted to assess the effect of intrinsic PEEP in case of obstructive lung disorders.PAV+ estimations of respiratory system mechanics are indirect and hence, based on a number of assumptions which, if not entirely fulfilled, could affect the accuracy and the reliability of Pmus and WOB calculation and, eventually, the delivered Paw. The aim of this study was to assess the accuracy of delivered Paw and calculated WOB by PAV+ under different conditions in terms of respiratory system mechanics and patient breathing pattern and to examine the factors influencing this accuracy. In the first step, we assessed the accuracy of PAV+ by comparing the actual delivered Paw by the ventilator () to the theoretical Paw as defined by the equation of motion (). Since Paw provided by the ventilator is always a proportion of Pmus, errors in Paw would reflect errors in Pmus and, hence, errors in WOB calculation. Therefore, in the second step, we assessed the accuracy of PAV+ WOB calculation by comparing the value displayed by the ventilator to the WOB, estimated by the Campbell diagram.
Methods
A PB840 ventilator (Puritan-Bennett 840; Covidien, Boulder, USA), which is currently the only ventilator able to deliver PAV+, was used. The ventilator was connected to the lung simulator by a conventional circuit with separate inspiratory and expiratory limbs. The circuit was connected to the lung simulator with the use of a heat–moisture exchanger (HME). Prior to each experiment, the ventilator was calibrated and tested for leaks. All experiments took place using room air [fraction of inspired oxygen (FiO2) 21 %].
Lung model
An Active Servo Lung 5000 (ASL 5000; Ingmar Medical, Pittsburg, PA, USA) was used as described in previous studies [11, 12]. The ASL 5000 is a digitally controlled real-time breathing computerized simulator consisting of a piston moving inside a cylinder. To control the piston’s movement, a microprocessor is programmed with a script driver, which uses a mathematical model of the equation of motion. Instantaneous flow (V′) and Paw were measured by flow and pressure sensors at the entrance of the piston, and volume (V) is obtained by the integration of V′ over time. The spontaneous breathing pattern of the lung simulator was determined by the lung model parameters and the inspiratory effort (Pmus). The two main lung model parameters were the compliance CRS and the resistance RRS. The inspiratory effort embodied the breathing rate, effort amplitude, effort slope and inhaled percentage. Thus, a range of respiratory mechanics and inspiratory efforts could be simulated.
Formulas
Equation of motion of the respiratory systemwhere Ptot is the total pressure applied to the respiratory system, RRS and ERS the respiratory system resistance and elastance, respectively, V′ and V the instantaneous flow and volume and PEEPt the total positive end-expiratory pressure.PAV gain represents the proportion which balances the ventilator (Paw) and the patient (Pmus) contribution to total pressure of the respiratory system (Ptot):
Design of the experiment
Accuracy of Paw delivered during PAV+
The ASL 5000 was set at single-compartment model to resemble four respiratory mechanics conditions: normal (RRS = 10 cmH2O/L/s and CRS = 60 mL/cmH2O), obstructive (RRS = 20 cmH2O/L/s and CRS = 60 mL/cmH2O), restrictive (RRS = 10 cmH2O/L/s and CRS = 30 mL/cmH2O) and mixed obstructive and restrictive (RRS = 20 cmH2O/L/s and CRS = 30 mL/cmH2O). A semi-sinusoidal inspiratory waveform was selected. The inspiratory waveform had a rise time of 30 %, inspiratory holding time of 0 % and releasing time of 15 %. PAV+ was tested under different respiratory mechanics (as described above), gains (30 and 60 %), inspiratory trigger (0.8, 5, and 15 L/min), Pmus (10 and 15 cmH2O) and PEEP levels (0 and 5 cmH2O). Various respiratory rates (RR) (10, 15, 20, 25 and 30 breath/min) were examined during obstructive respiratory mechanics to assess the impact of intrinsic PEEP (PEEPi) on delivered Paw. In total, 24 different conditions were tested.
Accuracy of WOB calculation
The ASL 5000 was set as a single-compartment model to resemble three respiratory mechanics: normal (RRS = 5 cmH2O/L/s, CRS = 60 mL/cmH2O), restrictive (RRS = 5 cmH2O/L/s, CRS = 20 mL/cmH2O) and obstructive (RRS = 20 cmH2O/L/s, CRS = 60 mL/cmH2O). A semi-sinusoidal inspiratory waveform was selected with a rise time of 25 %, inspiratory holding time of 0 % and releasing time of 15 %. Each condition was tested under 2 RR (20 and 30 breaths/min) and across 2 levels of Pmus (5 and 10 cmH2O). The flow triggering was 0.8 L/min and PEEP was 5 cmH2O.Moreover, we also used a real esophageal pressure signal, derived from our patients’ database, to drive the simulator and test PAV+. This signal was tested under normal, restrictive and obstructive conditions.The aforementioned scenarios were examined at two PAV+ gain levels (30 and 60 %). Therefore, 15 scenarios at two levels of Gain were assessed (30 conditions in total).
Data collection
Each test was recorded for 5 min, which was the time period for stabilization of the system. Then the last minute of the recording was selected and analyzed offline.Data acquisition from ASL 5000 was performed at 128 Hz and stored in a laptop computer for subsequent analysis with AcqKnowledge software (Biopac Systems, Goleta, CA, USA). V′, Pmus and curves over time were provided by the ASL 5000. V was derived from V′ integration over time.PEEPi was estimated as the pressure difference between the Pmus at the onset of inspiration (defined as the first point where Pmus started to decrease at end-expiration) and Pmus at the start of inspiratory V′.
Variables were expressed as medians (25–75th interquartile range, IQR) or means (±standard deviation, SD). Data were analyzed using nonparametric tests. The relationships between WOBdispl and WOBreal, between %Δi and PEEPi and between %error C and %error R and PEEPi were evaluated using Pearson’s correlation. Bland–Altman analysis was used to compare the absolute values of WOBdispl with those of WOBreal which was regarded as the gold standard.Statistical significance was defined at p value <0.05. The statistical analysis was performed using Prism (GraphPad Software, La Jolla, CA, USA).
Results
Accuracy of Paw delivered during PAV+
Effect of different respiratory mechanics and gains
Irrespective of respiratory mechanics and gain, imeas was always lower than iTh (Δi and %Δi were −2.9 ± 0.9 cmH2O and −25.4 ± 4.6 %, respectively), indicating a lower assistance provided by the ventilator than expected from the equation of motion (Table 1). The magnitude of underassistance was greater at the beginning of the inspiratory cycle and decreased toward the end of inspiration (Fig. 1).
Table 1
Measured and theoretical mean airway pressure during inspiration (i
meas and i
Th) in different respiratory mechanics
Gain (%)
Mechanics
imeas (cmH2O)
iTh (cmH2O)
Δi (cmH2O)
%Δi (%)
30
Normal
6.6
9.4
−2.8
−29.8
Obstructive
8.4
10.8
−2.4
−22.2
Restrictive
7.0
8.5
−1.5
−17.6
Mixed
6.7
8.8
−2.1
−23.9
60
Normal
9.6
13.7
−4.2
−29.9
Obstructive
9.5
13.6
−4.1
−30.1
Restrictive
10.2
13.0
−2.8
−21.5
Mixed
10.3
13.5
−3.2
−23.7
All conditions
8.5 ± 1.6
11.4 ± 2.3
−2.9 ± 0.9
−25.4 ± 4.6
Difference and percentage of difference between i
meas and i
Th were calculated as follows: Δi = i
meas − i
Th and %Δi = (i
meas − i
Th)/i
Th × 100. Inspiratory trigger = 5 L/min; muscular pressure = 10 cmH2O; PEEP = 5 cmH2O; respiratory rate = 20/min. Respiratory system mechanics, normal: resistance (R) = 10 cmH2O/L/s and compliance (C) = 60 mL/cmH2O; obstructive: R = 20 cmH2O/L/s and C = 60 mL/cmH2O; restrictive: R = 10 cmH2O/L/s and C = 30 mL/cmH2O; and mixed: R = 20 cmH2O/L/s and C = 30 mL/cmH2O
Fig. 1
Percentage of difference between measured airway pressure () and theoretical airway pressure () (%ΔP
aw) at 25, 50, 75 and 100 % of inspiration with different lung mechanics with gain 30 % (a) and 60 % (b). %ΔP
aw is expressed in percentage of (%ΔP
aw = ( − )/ × 100). Representative tracing of and in 4 respiratory mechanics with gain 30 % (c) and 60 % (d). Black lines
waveforms; blue lines
waveforms. Inspiratory trigger = 5 L/min; muscular pressure = 10 cmH2O; PEEP = 5 cmH2O; respiratory rate = 20/min. Respiratory mechanics; normal: resistance (R) = 10 cmH2O/L/s and compliance (C) = 60 mL/cmH2O; obstructive: R = 20 cmH2O/L/s and C = 60 mL/cmH2O; restrictive: R = 10 cmH2O/L/s and C = 30 mL/cmH2O; and mixed: R = 20 cmH2O/L/s and C = 30 mL/cmH2O
Measured and theoretical mean airway pressure during inspiration (i
meas and i
Th) in different respiratory mechanicsDifference and percentage of difference between i
meas and i
Th were calculated as follows: Δi = i
meas − i
Th and %Δi = (i
meas − i
Th)/i
Th × 100. Inspiratory trigger = 5 L/min; muscular pressure = 10 cmH2O; PEEP = 5 cmH2O; respiratory rate = 20/min. Respiratory system mechanics, normal: resistance (R) = 10 cmH2O/L/s and compliance (C) = 60 mL/cmH2O; obstructive: R = 20 cmH2O/L/s and C = 60 mL/cmH2O; restrictive: R = 10 cmH2O/L/s and C = 30 mL/cmH2O; and mixed: R = 20 cmH2O/L/s and C = 30 mL/cmH2OPercentage of difference between measured airway pressure () and theoretical airway pressure () (%ΔP
aw) at 25, 50, 75 and 100 % of inspiration with different lung mechanics with gain 30 % (a) and 60 % (b). %ΔP
aw is expressed in percentage of (%ΔP
aw = ( − )/ × 100). Representative tracing of and in 4 respiratory mechanics with gain 30 % (c) and 60 % (d). Black lines
waveforms; blue lines
waveforms. Inspiratory trigger = 5 L/min; muscular pressure = 10 cmH2O; PEEP = 5 cmH2O; respiratory rate = 20/min. Respiratory mechanics; normal: resistance (R) = 10 cmH2O/L/s and compliance (C) = 60 mL/cmH2O; obstructive: R = 20 cmH2O/L/s and C = 60 mL/cmH2O; restrictive: R = 10 cmH2O/L/s and C = 30 mL/cmH2O; and mixed: R = 20 cmH2O/L/s and C = 30 mL/cmH2O
Effect of different triggers, Pmus and PEEP
The underassistance of PAV+ was also highlighted under different trigger, Pmus or PEEP settings in normal respiratory mechanics (Fig. 2; Additional file 2: Tables S1, Additional file 3: Table S2, Additional file 4: Table S3). A higher trigger value (lower sensitivity) led to greater underassistance at the end of inspiration versus a lower trigger (Fig. 2a). A high Pmus was associated with a greater underassistance during the entire inspiration versus a low Pmus (Fig. 2b). A decrease in PEEP was associated with a major underassistance at the start of the inspiration (Fig. 2c). These findings were replicated under different trigger, Pmus or PEEP settings in obstructive and restrictive respiratory mechanics (Additional file 5: Figure S2). Of note, with obstructive respiratory mechanics and trigger = 15 L/min, the ventilator was unable to estimate initial values for Rvent and Rvent and the PAV+ mode did not operate.
Fig. 2
Percentage of difference between measured airway pressure and theoretical airway pressure (%ΔP
aw) at 25, 50, 75 and 100 % of inspiration with different inspiratory trigger (IT) (a), muscular pressure (P
mus) (b) and positive end-expiratory pressure (PEEP) (c) under normal respiratory mechanics. Difference between and is expressed in percentage of (%ΔP
aw = ( − )/ × 100). Resistance = 10 cmH2O/L/s; compliance = 60 mL/cmH2O; gain = 60 %; and respiratory rate = 20/min; a different IT at 0.8, 5 and 15 L/min; P
mus = 10 cmH2O; PEEP = 5 cmH2O. b Different P
mus at 10 and 15 cmH2O; IT 5 L/min; PEEP = 5 cmH2O; c different PEEP at 0 and 5 cmH2O; IT 5 L/min; P
mus 10 cmH2O
Percentage of difference between measured airway pressure and theoretical airway pressure (%ΔP
aw) at 25, 50, 75 and 100 % of inspiration with different inspiratory trigger (IT) (a), muscular pressure (P
mus) (b) and positive end-expiratory pressure (PEEP) (c) under normal respiratory mechanics. Difference between and is expressed in percentage of (%ΔP
aw = ( − )/ × 100). Resistance = 10 cmH2O/L/s; compliance = 60 mL/cmH2O; gain = 60 %; and respiratory rate = 20/min; a different IT at 0.8, 5 and 15 L/min; P
mus = 10 cmH2O; PEEP = 5 cmH2O. b Different P
mus at 10 and 15 cmH2O; IT 5 L/min; PEEP = 5 cmH2O; c different PEEP at 0 and 5 cmH2O; IT 5 L/min; P
mus 10 cmH2O
Effect of PEEPi
To assess the effect of PEEPi on the accuracy of Paw delivered by PAV+, the same experiments were replicated under obstructive respiratory mechanics with increasing RR. An increase in RR leading to an increase in PEEPi resulted in a higher %ΔPaw during the entire cycle, showing that PEEPi which is associated with an increase in trigger delay affected PAV+ accuracy (Fig. 3; Additional file 6: Table S4). Combining the data from all conditions, PEEPi was correlated with %Δi (Additional file 7: Figure S3). The higher the PEEPi, the lower the pressure assistance the ventilator provided.
Fig. 3
Percentage of difference between measured airway pressure () and theoretical airway pressure () (%ΔP
aw) at 25, 50, 75 and 100 % of inspiration with different respiratory rates in obstructive respiratory mechanics (a). Difference between and is expressed in percentage of (% ΔP
aw = ( − )/ × 100). Representative tracing of and (b). Black lines
waveforms. Blue lines
waveforms. Resistance = 20 cmH2O/L/s; compliance = 60 mL/cmH2O; gain = 60 %; inspiratory trigger = 5 L/min; PEEP = 0 cmH2O; and P
mus = 10 cmH2O
Percentage of difference between measured airway pressure () and theoretical airway pressure () (%ΔP
aw) at 25, 50, 75 and 100 % of inspiration with different respiratory rates in obstructive respiratory mechanics (a). Difference between and is expressed in percentage of (% ΔP
aw = ( − )/ × 100). Representative tracing of and (b). Black lines
waveforms. Blue lines
waveforms. Resistance = 20 cmH2O/L/s; compliance = 60 mL/cmH2O; gain = 60 %; inspiratory trigger = 5 L/min; PEEP = 0 cmH2O; and P
mus = 10 cmH2O
Measurements of CRS and RRS
In comparison with CRS, Cvent was globally slightly overestimated for the low value of compliance (30 mL/cmH2O) [34 (IQR 34–36) mL/cmH2O] and slightly underestimated for the high value of compliance (60 mL/cmH2O) [59 (IQR 63–65) mL/cmH2O] (Additional file 8: Figure S4). In comparison with RRS, Rvent was always underestimated, irrespective of RRS [8.8 (IQR 8.2–8.9) and 15 (IQR 15–16) cmH2O/L/s for RRS = 10 and 20 cmH2O/L/s, respectively] (Additional file 8: Figure S4).We found a strong correlation between the percentage of error in the measurement of compliance (%error C) and PEEPi (r2 = 0.68, p < 0.001, Additional file 9: Figure S5). The association between the percentage of error in the measurement of resistance (%error R) and PEEPi was weaker (r2 = 0.27, p = 0.007). This underestimation of compliance in case of high PEEPi should lead to an increase in the assistance delivered by the ventilator in comparison with (calculated with the actual compliance of the simulator) and thus counteract in part the underassistance observed in PAV+ when PEEPi is high.
Accuracy of WOB measurements during PAV+
There was a strong linear correlation between total WOB calculated by the ventilator and total WOB based on the Campbell diagram (r2 = 0.93, p < 0.001, Fig. 4). The Bland–Altman plot, performed to evaluate the accuracy of the absolute values of total WOB calculation, revealed a mean bias of 0.27 J/L, indicating an underestimation of the , with a limit of agreement of 0.6 to −0.11 J/L (Fig. 5). The changes in total WOB were accurately detected by the ventilator, but the absolute values of total WOB displayed by the ventilator were underestimated. The linear correlation between the patient’s WOB as calculated by the ventilator and as computed by the Campbell diagram was significant but much weaker than for WOBtot (r2 = 0.63, p < 0.001).
Fig. 4
Correlation between the total work of breathing calculated by the ventilator () and the corresponding calculated by Campbell ()
Fig. 5
Bland–Altman plot of total inspiratory work of breathing measurements, expressed in J/L, between the two methods compared (Campbell and Ventilator). , inspiratory work of breathing calculated by the ventilator; , inspiratory work of breathing calculated by the Campbell diagram
Correlation between the total work of breathing calculated by the ventilator () and the corresponding calculated by Campbell ()Bland–Altman plot of total inspiratory work of breathing measurements, expressed in J/L, between the two methods compared (Campbell and Ventilator). , inspiratory work of breathing calculated by the ventilator; , inspiratory work of breathing calculated by the Campbell diagram
Discussion
This bench study showed that although the Paw delivered by PAV+ reasonably follows the muscular pressure when compared to the theoretical Paw (i.e., the pressure that the ventilator should ideally deliver according to the equation of motion of the respiratory system), the ventilator provides a 25 % underassistance irrespective of the respiratory mechanics or ventilator settings. This underassistance is particularly marked when PEEPi is high. Of note even slight PEEPi values lead to dramatic increases in this underassistance (around 40 % for PEEPi values around 4 cmH2O).This inaccuracy in compared with reflects the inaccurate estimation of Pmus and is thus associated with an underestimation of WOB. Despite the fact that the absolute values of WOB displayed on the ventilator bar graph underestimated the actual WOB, changes in its values accurately reflect measured changes in WOB.Trigger delay and PEEPi play a pivotal role in this underassistance and in this relative inaccuracy of WOB measurements in PAV+ mode. PEEPi is associated with an increase in trigger delay. On PAV+ mode, once the ventilator is triggered delivered Paw is continuously proportional to Pmus, but during the triggering phase no assistance is provided by the ventilator. A delay in the onset of pressurization by the ventilator reduces the correctly assisted fraction of neural inspiratory time and thus the global assistance by the ventilator. Thus, the further increase in trigger delay due to PEEPi leads to a global underassistance. In critically ill patients ventilated in PAV+ mode, Kondili et al. showed that an increase in PEEPi from 0.8 to 3.2 cmH2O due to an increase in respiratory workload by chest and abdominal wall compression led to a decrease in the portion of supported inspiratory effort from 86 to 66 % [5]. The conventional pneumatic triggering used in PAV+ ventilation appears as an important limitation especially when compared to the other proportional mode of ventilation (neurally adjusted ventilatory assist), in which the triggering by the electrical activity of the diaphragm (Eadi) is not affected by PEEPi [14, 15]. However, a high PEEPi may also lead to a greater underestimation of CRS. To estimate the compliance, the ventilator applies a 300-ms pause maneuver at the end of inspiration at random intervals of four to 10 breaths [7]. Paw at the end of the occlusion (Paw, occl) is measured and CRS is calculated by the equation of motion (C = V/(Paw, occl-totalPEEP)). However, as PB840 cannot detect the actual PEEPi value, this calculated value of CRS may be underestimated in case of dynamic hyperinflation [9]. Thus, the assistance provided by the ventilator calculated by using the equation of motion of the respiratory system is increased as a result of this underestimation of CRS. Overall, PEEPi has two effects on PAV+ accuracy: It is associated with an increase in trigger delay leading to an underassistance, but this effect is in part counterbalanced by the effect on compliance estimation. Of note, this underassistance delivered by PAV+ may prevent the occurrence of runaway phenomena [16].One of the major advantages of PAV+ is to allow the clinician to assess noninvasively the WOB. We show in this study that the absolute value of total WOB is underestimated by the ventilator. This finding is of particular importance when we consider the way to adjust Gain with a WOB range target [10]. Importantly for clinical practice, the changes in total WOB in a patient are accurately detected by the ventilator.The main limitation of our study is that it is a bench study. Even if the lung model that we used was set to imitate human spontaneous breathing in different normal and pathological conditions, this does not reproduce the complexity of the control of breathing. Regarding the specific question addressed, however, this does not invalidate our findings, and we simply cannot use these data to comment on the clinical consequences of this.This study suggests that, in clinical practice, because of the major role of PEEPi in PAV+ inaccuracy, recommendations should include a careful external PEEP titration when PEEPi is suspected. In addition, using a high trigger sensitivity is recommended to reduce the underassistance by PAV+. Following these recommendations, the underassistance has probably a modest clinical impact, whereas WOB values displayed by the ventilator may not be accurate enough to be used to monitor effect of PAV+. However, a clinical study is needed to support these recommendations.
Conclusion
The PAV+ assistance reasonably well follows Pmus but provides a constant underassistance of around 25 % on average, especially at the beginning of inspiration. This underassistance is logically associated with an underestimation by the ventilator of the actual total WOB. PEEPi leading to increased trigger delay is a major factor contributing to PAV+ inaccuracy. Clinical recommendations should include using a high trigger sensitivity and a careful PEEP titration when PEEPi is suspected.
Authors: Stéphanie Delaere; Jean Roeseler; William D'hoore; Pascal Matte; Marc Reynaert; Philippe Jolliet; Thierry Sottiaux; Giuseppe Liistro Journal: Intensive Care Med Date: 2003-03-27 Impact factor: 17.440
Authors: Nektaria Xirouchaki; Eumorfia Kondili; Katerina Vaporidi; George Xirouchakis; Maria Klimathianaki; George Gavriilidis; Evi Alexandopoulou; Maria Plataki; Christina Alexopoulou; Dimitris Georgopoulos Journal: Intensive Care Med Date: 2008-07-08 Impact factor: 17.440
Authors: Martin Bellgardt; Dominik Drees; Vladimir Vinnikov; Livia Procopiuc; Andreas Meiser; Hagen Bomberg; Philipp Gude; Heike Vogelsang; Thomas Peter Weber; Jennifer Herzog-Niescery Journal: J Clin Monit Comput Date: 2018-04-09 Impact factor: 2.502
Authors: Annemijn H Jonkman; Michela Rauseo; Guillaume Carteaux; Irene Telias; Michael C Sklar; Leo Heunks; Laurent J Brochard Journal: Intensive Care Med Date: 2020-08-11 Impact factor: 17.440