| Literature DB >> 27066203 |
Jae-Hang Shim1, Woo Jae Jeon1, Sang Yun Cho1, Gyu Ho Choe1.
Abstract
BACKGROUND: A "difficult airway" can be simulated with an extrication collar, which restricts cervical motion and mouth opening. The purpose of this study is to compare the efficacy of the GlideScope and the McGrath in difficult airway simulation.Entities:
Keywords: Airway management; Intratracheal intubation; Laryngoscopes
Year: 2016 PMID: 27066203 PMCID: PMC4823407 DOI: 10.4097/kjae.2016.69.2.133
Source DB: PubMed Journal: Korean J Anesthesiol ISSN: 2005-6419
Fig. 1Study flow diagram.
Characteristics and Airway Data of Patients Undergoing Tracheal Intubation Using the GlideScope or the McGrath Method with a Tube Exchanger and Forceps
| GlideScope | McGrath | |
|---|---|---|
| ASA physical status (I/II/III) | 11/18/0 | 18/11/1 |
| Age (yr) | 47.2 ± 12.6 | 44.5 ± 12.9 |
| Sex (M/F) | 14/15 | 14/16 |
| Weight (kg) | 67.6 ± 9.8 | 67.2 ± 12.1 |
| Height (cm) | 164.9 ± 8.2 | 165.7 ± 9.1 |
| Body mass index (kg/m2) | 24.8 ± 2.5 | 24.4 ± 3.3 |
| Thyromental distance (cm) | 6.6 ± 0.8 | 6.9 ± 0.6 |
| Neck circumference (cm) | 38.1 ± 4.1 | 37.9 ± 4.0 |
| Mouth opening without a cervical collar (cm) | 3.9 ± 0.5 | 3.8 ± 0.4 |
Values are expressed as raw numbers or means ± SD. ASA: American Society of Anesthesiologists.
Characteristics of Intubation with the GlideScope and McGrath with a Tube Exchanger and Forceps and the Associated Complications
| GlideScope | McGrath | P value | |
|---|---|---|---|
| Mouth opening with collar (cm) | 0.9 ± 0.3 | 1.1 ± 0.4 | 0.30 |
| Laryngoscopic view (grade 1/2/3/4) | 0/0/15/14 | 0/0/16/14 | 0.90 |
| Videoscopic view (grade 1/2/3/4) | 21/8/0/0 | 19/10/1/0 | 0.52 |
| Total intubation time (seconds) | 73.0 ± 25.3 | 72.3 ± 20.9 | 0.92 |
| Intubation attempts (1/2/3) | 24/4/2 | 25/3/2 | 0.79 |
| First attempt success rate | 24/29 (82.8%) | 25/30 (83.3%) | 0.95 |
| Sore throat in the recovery room | 3/29 (10.3%) | 4/30 (13.3%) | 0.72 |
| Sore throat at 24 h | 5/29 (17.2%) | 2/30 (6.7%) | 0.21 |
| Dental injury | 0 | 0 | |
| Blood on the tube | 5/29 (17.2%) | 2/30 (6.7%) | 0.21 |
Values are expressed as raw numbers, proportions, or means ± SD.
Fig. 2Kaplan-Meier plot showing the percentage of patients intubated as time progressed. Total intubation times was not significantly different between two groups.