| Literature DB >> 27052063 |
Caetano Nigro Neto, Francesco De Simone, Luigi Cassara, Carlos Gustavo Dos Santos Silva, Thiago Augusto Azevedo Marãnhao Cardoso, Francesco Carco, Alberto Zangrillo, Giovanni Landoni1.
Abstract
BACKGROUND: Recently, evidence of cardio-protection and reduction in mortality due to the use of volatile agents during cardiac surgery led to an increase in their use during cardiopulmonary bypass (CPB). These findings seem to be enhanced when the volatile agents are used during all the surgical procedure, including the CPB period. AIMS: Since the administration of volatile agents through CPB can be beneficial to the patients, we decided to review the use of volatile agents vaporized in the CPB circuit and to summarize some tricks and tips of this technique using our 10-year experience of Brazilian and Italian centers with a large volume of cardiac surgeries. STUDYEntities:
Mesh:
Substances:
Year: 2016 PMID: 27052063 PMCID: PMC4900340 DOI: 10.4103/0971-9784.179592
Source DB: PubMed Journal: Ann Card Anaesth ISSN: 0971-9784
Tricks and tips for the adapted vaporize system using volatile agents during cardiopulmonary bypass based on the 10 years’ experience of two teaching hospitals (one in Italy and one in Brazil)
| Problem | Cause | Solution |
|---|---|---|
| 1. Insufficient blood oxygenation | Vaporizer not correctly put on the stirrup | Reposition the vaporizer correctly on the stirrup in other to avoid gas flux leakage |
| Breath in fraction Luer Lock not correctly closed | Connect the breathe in reading line or a little plug in order to avoid gas leak | |
| 2. No inspired fraction reading gas | a. Reading line not positioned | Place a reading line |
| b. Broken reading line | Place a new reading line | |
| c. Tap (if present) not correctly turned | Turn the tap in the direction of the monitoring | |
| d. Empty vaporizer | Reload the vaporizer | |
| e. Switched off vaporizer | The gas vaporizer is an electric supply and it must be connected to the net or to a new light socket that works properly | |
| f. Vaporizer not correctly positioned on the stirrup | Correctly reposition the vaporizer on the stirrup | |
| g. Nonelectronic vaporizer run out of gas even though the sentinel column is sufficient | Reload the vaporizer (see problem 9) | |
| 3. Malfunctioning of the vaporizer alarm | The electronic vaporizer was put on a stirrup where the gas input and output were inverted | Invert the gas input and output (air/oxygen) following the indication on the stirrup |
| 4. The vaporizer does not function and the lights are flashing simultaneously | The electronic vaporizer did not pass the initial check because it was positioned on a stirrup not in axis, but turned forward | While waiting for the stirrup maintenance the vaporizer can be put on a horizontal plain. After passing the check, put it on the stirrup |
| 5. The vaporizer flashes the heating | The electronic vaporizer (desflurane) did not reach the proper temperature | Wait the proper time to reach the optimal temperature. Inform the anesthesiologist and administer intravenous agents while waiting |
| 6. On the vaporizer there is the gas reserve alarm but the sentinel column is full enough | The electronic vaporizer is on a stirrup that is not in axis, and the reading level on the sentinel column is altered | While waiting for the stirrup maintenance, the vaporizer must be reloaded independently of the sentinel column level |
| 7. Alarm during the operation of the electronic vaporizer (desflurane) knob | Electronic vaporizer not connected to the electric current | Connect it to the electric current trough a functioning socket |
| 8. Vaporizer knob is blocked | The nonelectronic vaporizer gas run out the gas | Reload the vaporizer |
| 9. Sentinel column of the nonelectronic vaporizer is sufficient but the inhale fraction is zero | The nonelectronic vaporizer is empty but is positioned on a bar that is not in axis and this alters the reading level on the sentinel | While waiting for the stirrup maintenance, the vaporizer must be reloaded independently of the sentinel column level |
| 10. EtCO2 curve present but the reading of the exhale halogenates fraction is zero | The vaporizer is turned off | Turn it on according to the protocol |
| 11. EtCO2 curve present but the reading of the exhale halogenates fraction is zero even though the vaporizer is turned on | The vaporizer is not working | Monitor the inhale gas fraction, if it does not detect any value of halogenate see problem point-2d, point-2e, point-2g, and point-9 |
| 12. Exhale gas leakage (halogenates included) from the additional discharge doors | The bayonet socket of venturi was not correctly placed in the discharge door on the wall | Place it in the proper way and make sure that the light-emitted diode, if present, is switched on |
| The gas waste pipe (between the gas exit and the venturi entry) is too long | Shorten the length of the waste pipe in order to reduce the resistance to the disposal | |
| The gas waste pipe (between the gas exit and the venturi entry) has a section area too small | Replace it with a tube with a proper section area that does not offer too many resistances to the gas disposal | |
| Luer lock the exhale fraction not correctly closed | Connect the line to reading the exhale fraction or put a plug in order to avoid the discharge gas leakage | |
| Air/oxygen fluxes too high (higher than 2 L/min) | Reduce the fluxes if possible or, Reduce the dead space between the gas oxygen exit and the Venturi entry or Replace the gas evacuator with another one equipped with a better aspiration | |
| The gas reading line is positioned inside the additional oxygen discharge door | Throw it back until it is positioned in a few millimeters from the additional discharge door in order to avoid the reading (of the gas) still inside the oxygen | |
| 13. Venturi manometer on the zero position | The venturi bayonet not correctly placed in the wall socked | Place it properly and make sure that the light-emitting diode (if present) is turned on |
| The discharge door is not working | IF pushing manually the valve of the discharge door on the wall you cannot feel any suction, stop the halogenate protocol | |
| Obstruction of the rubber tube that runs from the wall discharge door to the Venturi | Remove the obstacle | |
| The venturi could have been connected to the secondary discharge door of the oxygen-gas | Check |
EtCO2: End-tidal carbon dioxide
Figure 1Schematic representation of the use of volatile agents adapted to cardiopulmonary bypass machine
Difference in the hollow-fiber membrane oxygenators and their variability in transmembrane passage of volatile agents
| Hollow-fiber membrane oxygenators | Transmembrane passage | Examples |
|---|---|---|
| Polypropylene | Yes | Standard CPB |
| Poly-(4-methyl-1-pentene) | No | ECMO |
ECMO: Extracorporeal membrane oxygenation, CPB: Cardiopulmonary bypass