Alois Philipp1, Christoph Wiesenack, Renate Behr, Franz X Schmid, Dietrich E Birnbaum.
Abstract
In cardiac surgery with the aid of extracorporeal circulation (ECC), inhalation anaesthetics can be administered via the oxygenator. Until the recent advent of a new type of diffusion membrane oxygenator, we routinely added the inhalation agent, isoflurane, to the gas flow of a microporous capillary membrane-type oxygenator. Applying this procedure to the diffusion-type oxygenators, the depth of anaesthesia appeared to be affected, which manifested itself through unusually high intraoperative perfusion pressures. This observation led to a prospective randomized study comprising 60 patients and two models of a microporous capillary membrane oxygenator, as well as two models of a diffusion membrane oxygenator. Simultaneous isoflurane concentration measurements at both the gas inlet and outlet ports of the oxygenators showed that, whereas in the microporous capillary-type oxygenators the isoflurane administered was reduced by about 50% during the passage of gas through the device, there was only a minimal transfer of isoflurane in the diffusion-type membrane oxygenators.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12017384 DOI: 10.1191/0267659102pf566oa
Source DB: PubMed Journal: Perfusion ISSN: 0267-6591 Impact factor: 1.972