| Literature DB >> 27048639 |
Christina A Sutherland1, David P Nicolau2.
Abstract
Herein, we report the development and validation of an HPLC method to analyze ceftolozane and tazobactam simultaneously in human plasma, human serum, swine serum and saline matrixes. A reversed-phase column was used with a UV detector set at 260 nm and switched to 218 nm. The mobile phase consisted of methanol and sodium phosphate buffer at a flow rate of 1.1 mL/min. Cefepime was used as the internal standard. The standard curves were linear over a range of 0.4-50 μg/mL. This methodology represents a simple, reproducible approach to the determination of drug concentrations with sufficient accuracy and precision for pharmacokinetic studies undertaken with this recently FDA-approved antimicrobial therapy.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27048639 DOI: 10.1093/chromsci/bmw047