| Literature DB >> 27042646 |
Eduardo da Silva Guarilha1, Paulo Roberto de Andrade Fígaro Caldeira1, Ozorio de Almeida Lira Neto2, Marcelo Schmidt Navarro3, Antonio Milani4, Mario Carneiro Filho5.
Abstract
OBJECTIVE: This study had the objective of prospectively comparing transverse fixation (Cross-Pin™) with extracortical fixation (EZLoc™) for the femur, in surgical reconstruction of the anterior cruciate ligament, from a clinical, biomechanical and functional point of view.Entities:
Keywords: Anterior Cruciate Ligament; Arthroscopy; Prospective Studies
Year: 2015 PMID: 27042646 PMCID: PMC4799398 DOI: 10.1016/S2255-4971(15)30111-7
Source DB: PubMed Journal: Rev Bras Ortop ISSN: 2255-4971
Characterization of the sample.
| Variables | n = 50 |
|---|---|
| Age (years) – mean (sd) | 34.1 (11.03) |
| minimum – maximum | 13 – 57 |
| Female | 4 (8) |
| Male | 46 (92) |
| Right | 27 (54.0) |
| Left | 23 (46.0) |
| None | 33 (66) |
| With meniscal lesion | 14 (28) |
| With meniscal and chondral lesions | 3 (6) |
| Soccer | 31 (62) |
| Martial arts | 12 (24) |
| Volleyball | 4 (8) |
| Others | 3 (6) |
Figure 1Cross-Pin™ transfemoral fixation.
Figure 2EZLocTM transfemoral fixation.
Lysholm scale.
| Mean | sd | Median | Minimum | Maximum | |
|---|---|---|---|---|---|
| Before operation – Total | 66.90 | 66.90 | 70 | 73.50 | 23 |
| After operation – Total | 92.80 | 92.80 | 93 | 94.00 | 71 |
| Group 1 – Before operation | 72.48 | 16.16 | 75 | 75.00 | 28.00 |
| Group 1 – After operation | 92.52 | 6.69 | 93 | 93.00 | 76.00 |
| Group 2 – Before operation | 61.32 | 21.75 | 63 | 63.00 | 23.00 |
| Group 2 – After operation | 93.08 | 9.04 | 99 | 99.00 | 71.00 |
sd – standard deviation; group 1 – Cross-Pin™; group 2 – EZLoc™.
KT-1000™ arthrometer.
| Variable | Fixation | N | Mean | sd | Median | Minimum | Maximum | P |
|---|---|---|---|---|---|---|---|---|
| Difference | Cross-Pin™ | 25 | 2.16 | 1.95 | 2.00 | −2.00 | 5.00 | 0.703 |
| EZLoc™ | 25 | 2.10 | 2.63 | 1.00 | −3.00 | 7.00 | ||
Results table with application of nonparametric Mann-Whitney test.
KT-1000 arthrometer.
| Fixation | |||||
|---|---|---|---|---|---|
| Cross-Pin™ | EZLoc™ | p | |||
| Good | N | 18 | 17 | ||
| % | 72.0% | 68.0% | 1.000 | ||
| Poor | N | 7 | 8 | ||
| % | 28.0% | 32.0% | |||
Chi-square test was used for comparisons between the two techniques.
Figure 3Graph showing data from SF-36 quality-of-life questionnaire.