| Literature DB >> 27040110 |
Foroozan Hasanpour1, Hassan Hadadzadeh2, Masoumeh Taei3, Mohsen Nekouei3, Elmira Mozafari3.
Abstract
Analytical performance of conventional spectrophotometer was developed by coupling of effective dispersive liquid-liquid micro-extraction method with spectrophotometric determination for ultra-trace determination of cobalt. The method was based on the formation of Co(II)-alpha-benzoin oxime complex and its extraction using a dispersive liquid-liquid micro-extraction technique. During the present work, several important variables such as pH, ligand concentration, amount and type of dispersive, and extracting solvent were optimized. It was found that the crucial factor for the Co(II)-alpha benzoin oxime complex formation is the pH of the alkaline alcoholic medium. Under the optimized condition, the calibration graph was linear in the ranges of 1.0-110 μg L(-1) with the detection limit (S/N = 3) of 0.5 μg L(-1). The preconcentration operation of 25 mL of sample gave enhancement factor of 75. The proposed method was applied for determination of Co(II) in soil samples.Entities:
Keywords: Alpha-benzoin oxime; Cobalt(II); Dispersive liquid-liquid micro-extraction; Spectrophotometric determination
Mesh:
Substances:
Year: 2016 PMID: 27040110 DOI: 10.1007/s10661-016-5263-x
Source DB: PubMed Journal: Environ Monit Assess ISSN: 0167-6369 Impact factor: 2.513