Evaluation of sealed vacuum extraction method (Seditainer) for measurement of erythrocyte sedimentation rate.

W N Patton1, P J Meyer, J Stuart.   

Abstract

A sealed vacuum extraction method (Seditainer) for determining the erythrocyte sedimentation rate (ESR) was compared with the standard Westergren ESR technique. The Seditainer method was particularly easy to use, showed acceptable precision, reduced the biohazard risk to laboratory staff and, on storage of sealed blood specimens for 24 hours at 4 degrees C, gave ESR values that had decreased by a mean of only 1.9% (95% CI + 0.2 to -4.0%). Seditainer tubes are shorter (100 mm) than Westergren tubes (200 mm) and this reduced test sensitivity at ESR values above 55 mm/first hour. After application of the manufacturer's correction formula to adjust for tube length there was improved correlation (n = 150, r = 0.936, p less than 0.001), but still considerable scatter, between the Seditainer and Westergren results. Seditainer ESR values above 55 mm/first hour should therefore be regarded as semiquantitative. This vacuum extraction method offers a simple and safe technique for measuring the ESR and specimens can be stored overnight at 4 degrees C.

Mesh:

Year:  1989        PMID: 2703548      PMCID: PMC1141875          DOI: 10.1136/jcp.42.3.313

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  1 in total

1.  Guidelines on selection of laboratory tests for monitoring the acute phase response. International Committee for Standardization in Haematology (expert panel on blood rheology).

Authors: 
Journal:  J Clin Pathol       Date:  1988-11       Impact factor: 3.411

  1 in total
  5 in total

1.  Evaluation of a 200 mm long vacuum aspiration tube for measurement of erythrocyte sedimentation rate.

Authors:  M Caswell; J Stuart
Journal:  J Clin Pathol       Date:  1991-05       Impact factor: 3.411

2.  Assessment of Diesse Ves-matic automated system for measuring erythrocyte sedimentation rate.

Authors:  M Caswell; J Stuart
Journal:  J Clin Pathol       Date:  1991-11       Impact factor: 3.411

3.  A Comparison Between the StaRRsed Auto-Compact Erythrocyte Sedimentation Rate Instrument and the Westergren Method.

Authors:  Juha Horsti; Riikka Rontu; Auni Collings
Journal:  J Clin Med Res       Date:  2010-12-11

4.  Elevated C-reactive protein is associated with lower increase in knee muscle strength in patients with knee osteoarthritis: a 2-year follow-up study in the Amsterdam Osteoarthritis (AMS-OA) cohort.

Authors:  Diana C Sanchez-Ramirez; Marike van der Leeden; Martin van der Esch; Leo D Roorda; Sabine Verschueren; Jaap H van Dieën; Joost Dekker; Willem F Lems
Journal:  Arthritis Res Ther       Date:  2014-06-13       Impact factor: 5.156

5.  Imaging Erythrocyte Sedimentation in Whole Blood.

Authors:  Alexis Darras; Hans Georg Breunig; Thomas John; Renping Zhao; Johannes Koch; Carsten Kummerow; Karsten König; Christian Wagner; Lars Kaestner
Journal:  Front Physiol       Date:  2022-01-28       Impact factor: 4.566

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.