Woo Seok Lee1, Jun Kim1, Chang-Hee Kwon1, Jin Hee Choi1, Uk Jo1, Yoo Ri Kim1, Gi-Byoung Nam1, Kee-Joon Choi1, You-Ho Kim1.
Abstract
BACKGROUND AND OBJECTIVES: Implantable cardioverter-defibrillators (ICDs) are indicated in patients with Brugada syndrome (BS), early repolarization syndrome (ERS), or idiopathic ventricular fibrillation (IVF) who are at high risk for sudden cardiac death. The optimal ICD programming for reducing inappropriate shocks in these patients remains to be determined. We investigated the difference in the mean cycle length of tachyarrhythmias that activated either appropriate or inappropriate ICD shocks in these three patient groups to determine the optimal ventricular fibrillation (VF) zone for minimizing inappropriate ICD shocks. SUBJECTS AND METHODS: We selected 41 patients (35 men) (mean age±standard deviation=42.6±13.0 year) who received ICD shocks between April 1996 and April 2014 to treat BS (n=24), ERS (n=9), or IVF (n=8). Clinical and ICD interrogation data were retrospectively collected and analyzed for all events with ICD shocks.Entities:
Keywords: Brugada syndrome; Cardioverter defibrillator, implantable; Ventricular fibrillation
Year: 2016 PMID: 27014348 PMCID: PMC4805562 DOI: 10.4070/kcj.2016.46.2.179
Source DB: PubMed Journal: Korean Circ J ISSN: 1738-5520 Impact factor: 3.243
Baseline characteristics of patients with ICDs
| Total population | BS | ERS | IVF | p | |
|---|---|---|---|---|---|
| Age at implantation (year) | 42.6±13.0 | 43.3±12.9 | 39.7±13.1 | 43.6±14.3 | 0.755 |
| Male | 35 (85.4) | 24 (100.0) | 5 (55.6) | 6 (75.0) | 0.017 |
| Body mass index (kg/m2) | 22.8±2.8 | 22.3±2.4 | 23.3±2.1 | 23.5±2.3 | 0.330 |
| LVEF (%) | 61.4±5.6 | 63.0±5.9 | 61.9±3.6 | 56.4±4.6 | 0.011 |
| LVEDD (mm) | 50.4±3.6 | 49.7±3.5 | 50.8±3.6 | 52.1 3.4 | 0.228 |
| Clinical presentation | |||||
| Aborted cardiac arrest | 33 | 16 | 9 | 8 | N/A |
| Syncope | 3 | 3 | 0 | 0 | N/A |
| Inducible VF | 6 | 5 | 0 | 0 | N/A |
| Family history of sudden cardiac death | 11 | 7 | 2 | 2 | N/A |
| ICD shock during follow-up | |||||
| Appropriate shock | 29 | 18 | 7 | 4 | N/A |
| Inappropriate shock | 23 | 12 | 6 | 5 | N/A |
Values are means±standard deviation or number. ICD: implantable cardioverter–defibrillator, BS: Brugada syndrome, ERS: early repolarization syndrome, LVEF: left ventricular ejection fraction, LVEDD: left ventricular end-diastolic dimension, VF: ventricular fibrillation
Fig. 1Representative electrograms of the episodes obtained upon device interrogation. (A) Appropriate ICD shock due to VF, (B) inappropriate ICD shock due to atrial fibrillation with rapid ventricular response. ICD: implantable cardioverter-defibrillator, VF: ventricular fibrillation, AMP: amplitude, HV: high voltage shock, AS: atrial sensing, STIM: stimulation (antitachycardia pacing).
Causes of inappropriate ICD shocks
| Total episodes | Single-chamber ICDs | Dual-chamber ICDs | |
|---|---|---|---|
| Supraventricular tachyarrhythmia | 59 (92.2) | 42 (89.4) | 17 (100) |
| Abnormal sensing | |||
| Noise oversensing | 4 (6.3) | 4 (8.5) | 0 (0) |
| T-wave oversensing | 1 (1.6) | 1 (2.1) | 0 (0) |
Values are the number of episodes (%). ICD: implantable cardioverter–defibrillator
Fig. 2Distribution of the tachyarrhythmia cycle lengths of appropriate or inappropriate ICD shocks. ICD: implantable cardioverter-defibrillator.
Fig. 3Mean cycle lengths of the tachyarrhythmias causing appropriate or inappropriate ICD shocks in all patients. ICD: implantable cardioverter-defibrillator.
Mean cycle length of tachyarrhythmias causing appropriate or inappropriate shocks in all study patients and by diagnosis
| Total | BS | ERS | IVF | |
|---|---|---|---|---|
| Appropriate shock (ms) | 178.9±28.7* | 178.4±30.4† | 181.1±29.0† | 176.1±22.5† |
| Inappropriate shock (ms) | 284.8±24.4* | 291.5±20.1‡ | 292.6±22.3‡ | 274.8±25.0‡ |
Values are means±standard deviation. *p<0.001; †p=0.002; ‡p=0.013. BS: Brugada syndrome, ERS: early repolarization syndrome, IVF: idiopathic ventricular fibrillation
Fig. 4Best cutoff value for discriminating between appropriate and inappropriate ICD shocks (235 ms; sensitivity: 98.4%, specificity: 95.6%). ICD: implantable cardioverter-defibrillator