Falsely low values of oxygen saturation measured by pulse oximetry in a boy treated with Chinese herb tea.

Agnes S Meidert1, Alexander Lang2, Georg Hennig2, Patricia Bernasconi3, Aurelia Peraud4, Josef Briegel3, Tanija K Hüttl3.   

Abstract

An 8-year-old boy suffering from progressive glioblastoma was scheduled for neurosurgery. Prior to induction of anaesthesia pulse oximetry measured 64 % saturation of oxygen (SpO2). Arterial blood gas analysis revealed normal oxygen saturation and normal oxygen partial pressure. After having ruled out technical problems of pulse oximetry the neurosurgical procedure was halted. Meticulous examination of the child's history and medication did not explain a possible interaction of drugs with pulse oximetry. A Chinese herb tea had been given to the child, but was then stopped on the day of admission. The surgical procedure took place the next day without any complications. During the subsequent inpatient stay, repeated blood gas analyses showed normal oxygenation, but pulse oximetry measured initially SpO2 values of 64 %, gradually increasing over 7 days up to 91 % by the time of discharge from hospital. Blood samples were taken and analysed. Absorption spectroscopy from the patient's blood showed an uncommon absorption maximum at 684 nm besides the normal maxima. The normalisation of SpO2 values after stopping Chinese herb tea administration leads to the conclusion that one of its ingredients caused the distorted pulse oximetry measurement.

Entities:  

Keywords:  Absorption spectroscopy; Blood gas analysis; Chinese herb tea; Diagnostic error; Low oxygen saturation; Pulse oximetry

Mesh:

Substances:

Year:  2016        PMID: 27013078     DOI: 10.1007/s10877-016-9865-1

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  5 in total

Review 1.  Pulse oximetry.

Authors:  Roger Cn McMorrow; Michael G Mythen
Journal:  Curr Opin Crit Care       Date:  2006-06       Impact factor: 3.687

2.  Measurement of carboxyhemoglobin and methemoglobin by pulse oximetry: a human volunteer study.

Authors:  Steven J Barker; Jeremy Curry; Daniel Redford; Scott Morgan
Journal:  Anesthesiology       Date:  2006-11       Impact factor: 7.892

3.  A novel hemoglobin, Bonn, causes falsely decreased oxygen saturation measurements in pulse oximetry.

Authors:  Berndt Zur; Andreas Hornung; Johannes Breuer; Ulrike Doll; Christine Bernhardt; Michael Ludwig; Birgit Stoffel-Wagner
Journal:  Clin Chem       Date:  2008-03       Impact factor: 8.327

4.  International Standards for a Safe Practice of Anesthesia 2010.

Authors:  Alan F Merry; Jeffrey B Cooper; Olaitan Soyannwo; Iain H Wilson; John H Eichhorn
Journal:  Can J Anaesth       Date:  2010-09-21       Impact factor: 5.063

Review 5.  Pulse oximetry.

Authors:  Amal Jubran
Journal:  Crit Care       Date:  2015-07-16       Impact factor: 9.097

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.