A novel hemoglobin, Bonn, causes falsely decreased oxygen saturation measurements in pulse oximetry.

Berndt Zur1, Andreas Hornung, Johannes Breuer, Ulrike Doll, Christine Bernhardt, Michael Ludwig, Birgit Stoffel-Wagner.   

Abstract

BACKGROUND: A 4-year-old boy and his father exhibited low oxygen saturation measured transcutaneously by pulse oximetry, a finding that could not be confirmed by arterial blood gas analysis. Both patients exhibited slight hemolysis in their blood, and the boy had a microcytic anemia. There was no evidence of hypoxemia or methemoglobinemia. Despite the normal results from the arterial blood gas analysis, a right-to-left-shunt was assumed in the boy until a cardiology examination excluded this diagnosis. Sleep apnea syndrome was suspected in the father and treated with nocturnal positive pressure respiration based on the low oxygen saturation values obtained with pulse oximetry. Only after consultation with our laboratory was a hemoglobin variant suspected and investigated.
METHODS: We performed hemoglobin protein analysis by HPLC, electrophoretic separation, and spectrophotometry and DNA sequence analysis of the alpha-globin gene.
RESULTS: Both HPLC chromatographic separation and alkaline electrophoresis revealed a unique hemoglobin peak. In both patients, alpha-globin gene sequencing revealed a mutation resulting in a histidine-to-aspartatic acid substitution at position alpha87. The low oxygen saturation measurement by pulse oximetry was due to hemoglobin Bonn oxyhemoglobin having an absorption peak at 668 nm, near the 660 nm measured by pulse oximeters.
CONCLUSION: Hemoglobin Bonn is a novel hemoglobin variant of the proximal alpha-globin that results in falsely low oxygen saturation measurements with pulse oximetry.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18310146     DOI: 10.1373/clinchem.2007.095158

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  4 in total

1.  First reported case of hemoglobin lansing in Asia detected by false low oxygen saturation on pulse oximetry.

Authors:  Kenji Ishitsuka; Junji Uchino; Junko Kato; Mikihiro Ikuta; Kentaro Watanabe; Akira Matsunaga; Kazuo Tamura
Journal:  Int J Hematol       Date:  2012-05-26       Impact factor: 2.490

2.  A novel haemoglobin variant mimicking cyanotic congenital heart disease.

Authors:  Francisco Abecasis; Inês Marques; Celeste Bento; Anabela Ferrão
Journal:  BMJ Case Rep       Date:  2016-01-28

3.  Case-control studies of novel hemoglobin anomalies as differential diagnosis in sleep apnea syndrome.

Authors:  Berndt Zur; Michael Ludwig; Birgit Stoffel-Wagner
Journal:  Sleep Breath       Date:  2012-05-13       Impact factor: 2.816

4.  Falsely low values of oxygen saturation measured by pulse oximetry in a boy treated with Chinese herb tea.

Authors:  Agnes S Meidert; Alexander Lang; Georg Hennig; Patricia Bernasconi; Aurelia Peraud; Josef Briegel; Tanija K Hüttl
Journal:  J Clin Monit Comput       Date:  2016-03-25       Impact factor: 2.502

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.