Mhd Hassan Albogha1, Toru Kitahara1, Mitsugu Todo2, Hiroto Hyakutake3, Ichiro Takahashi4.
Abstract
Factors responsible for the success or failure of orthodontic mini-implants (OMIs) in clinical settings are unclear. Failure of OMIs was found to be associated with increased maximum principal strain (MaxPN) when assessed using the subject-specific finite element (FE) modeling technique. The purpose of the present study was to identify factors that increase MaxPN and thereby predispose the OMI to failure. Using the FE method, MaxPN was calculated around 28 OMIs placed in orthodontic patients, 6 of which failed during the first 5 months. Sixteen potential risk factors related to patients or to OMI position were measured on computerized tomographic images or calculated in FE models. The impact of these factors on MaxPN was verified using regression analysis. Three factors were found to have significant nonlinear relationships with MaxPN: cortical bone quality, vertical angulation of the OMI, and proximity of the OMI to the tooth in the direction of force. In conclusion, failure of an OMI is a multifactorial problem, and position and angulation of the implant are among the affecting factors. Slight apical inclination and positioning at least 1 mm off the root in the direction of force may significantly decrease failure probability.Entities:
Keywords: Bone screws; Finite element analysis; Orthodontic anchorage procedures; Retrospective study; Risk factors
Mesh:
Substances:
Year: 2016 PMID: 26983844 DOI: 10.1007/s10439-016-1584-8
Source DB: PubMed Journal: Ann Biomed Eng ISSN: 0090-6964 Impact factor: 3.934