EVALUATION OF DOSE REDUCTION POTENTIALS OF A NOVEL SCATTER CORRECTION SOFTWARE FOR BEDSIDE CHEST X-RAY IMAGING.

Bernhard Renger1, Carina Brieskorn2, Vivien Toth3, Detlef Mentrup4, Sascha Jockel4, Fabian Lohöfer3, Martin Schwarz3, Ernst J Rummeny3, Peter B Noël3.   

Abstract

Bedside chest X-rays (CXR) for catheter position control may add up to a considerable radiation dose for patients in the intensive care unit (ICU). In this study, image quality and dose reduction potentials of a novel X-ray scatter correction software (SkyFlow, Philips Healthcare, Hamburg, Germany) were evaluated. CXRs of a 'LUNGMAN' (Kyoto Kagaku Co., LTD, Kyoto, Japan) thoracic phantom with a portacath system, a central venous line and a dialysis catheter were performed in an experimental set-up with multiple tube voltage and tube current settings without and with an antiscatter grid. Images with diagnostic exposure index (EI) 250-500 were evaluated for the difference in applied mAs with and without antiscatter grid. Three radiologists subjectively assessed the diagnostic image quality of grid and non-grid images. Compared with a non-grid image, usage of an antiscatter grid implied twice as high mAs in order to reach diagnostic EI. SkyFlow significantly improved the image quality of images acquired without grid. CXR with grid provided better image contrast than grid-less imaging with scatter correction.
© The Author 2016. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Year:  2016        PMID: 26977074     DOI: 10.1093/rpd/ncw031

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  4 in total

1.  Characterization and validation of the thorax phantom Lungman for dose assessment in chest radiography optimization studies.

Authors:  Sunay Rodríguez Pérez; Nicholas William Marshall; Lara Struelens; Hilde Bosmans
Journal:  J Med Imaging (Bellingham)       Date:  2018-02-06

2.  Application of Deconvolution Algorithm of Point Spread Function in Improving Image Quality: An Observer Preference Study on Chest Radiography.

Authors:  Kum Ju Chae; Jin Mo Goo; Su Yeon Ahn; Jin Young Yoo; Soon Ho Yoon
Journal:  Korean J Radiol       Date:  2018-01-02       Impact factor: 3.500

3.  The Potential Role of Grid-Like Software in Bedside Chest Radiography in Improving Image Quality and Dose Reduction: An Observer Preference Study.

Authors:  Su Yeon Ahn; Kum Ju Chae; Jin Mo Goo
Journal:  Korean J Radiol       Date:  2018-04-06       Impact factor: 3.500

4.  Air gap technique is recommended in axiolateral hip radiographs.

Authors:  Susanne Kivistö; Antti Kotiaho; Anja Henner; Terhi Nevala; Jaakko Niinimäki; Miika T Nieminen; Matti Hanni
Journal:  J Appl Clin Med Phys       Date:  2020-09-21       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.