Literature DB >> 26941338

Combined leadless pacemaker and subcutaneous implantable defibrillator therapy: feasibility, safety, and performance.

F V Y Tjong1, T F Brouwer2, L Smeding2, K M Kooiman2, J R de Groot2, D Ligon3, R Sanghera4, M J Schalij5, A A M Wilde2, R E Knops2.   

Abstract

AIMS: The subcutaneous implantable cardioverter-defibrillator (S-ICD) and leadless pacemaker (LP) are evolving technologies that do not require intracardiac leads. However, interactions between these two devices are unexplored. We investigated the feasibility, safety, and performance of combined LP and S-ICD therapy, considering (i) simultaneous device-programmer communication, (ii) S-ICD rhythm discrimination during LP communication and pacing, and (iii) post-shock LP performance. METHODS AND
RESULTS: The study consists of two parts. Animal experiments: Two sheep were implanted with both an S-ICD and LP (Nanostim, SJM), and the objectives above were tested. Human experience: Follow-up of one S-ICD patient with bilateral subclavian occlusion who received an LP and two LP (all Nanostim, SJM) patients (without S-ICD) who received electrical cardioversion (ECV) are presented. Animal experiments : Simultaneous device-programmer communication was successful, but LP-programmer communication telemetry was temporarily lost (2 ± 2 s) during ventricular fibrillation (VF) induction and 4/54 shocks. Leadless pacemaker communication and pacing did not interfere with S-ICD rhythm discrimination. Additionally, all VF episodes (n = 12/12), including during simultaneous LP pacing, were detected and treated by the S-ICD. Post-shock LP performance was unaltered, and no post-shock device resets or dislodgements were observed (24 S-ICD and 30 external shocks). Human experience : The S-ICD/LP patient showed adequate S-ICD sensing during intrinsic rhythm, nominal, and high-output LP pacing. Two LP patients (without S-ICD) received ECV during follow-up. No impact on performance or LP dislodgements were observed.
CONCLUSION: Combined LP and S-ICD therapy appears feasible in all animal experiments (n = 2) and in one human subject. No interference in sensing and pacing during intrinsic and paced rhythm was noted in both animal and human subjects. However, induced arrhythmia testing was not performed in the patient. Defibrillation therapy did not seem to affect LP function. More data on safety and performance are needed. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author 2016. For permissions please email: journals.permissions@oup.com.

Entities:  

Keywords:  Animal experiments; Human; Leadless pacing; Pacemaker; Subcutaneous ICD

Mesh:

Year:  2016        PMID: 26941338     DOI: 10.1093/europace/euv457

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  10 in total

Review 1.  [Current impact of cardiac implantable electronic devices].

Authors:  J Kuschyk; B Rudic; M Borggrefe; I Akin
Journal:  Herz       Date:  2017-04       Impact factor: 1.443

Review 2.  Lead or be led: an update on leadless cardiac devices for general physicians.

Authors:  Benedict M Wiles; Paul R Roberts
Journal:  Clin Med (Lond)       Date:  2017-02       Impact factor: 2.659

Review 3.  [Leadless pacemakers and subcutaneously implantable cardioverter defibrillators].

Authors:  C Stellbrink; B Hansky; D Meyer Zu Vilsendorf
Journal:  Internist (Berl)       Date:  2018-10       Impact factor: 0.743

4.  Analysis of Screening Electrocardiogram for the Subcutaneous Defibrillator in Adults with Congenital Heart Disease.

Authors:  Vincent C Thomas; Mark Peterson; Martin McDaniel; Humberto Restrepo; Abraham Rothman; Amit Jain
Journal:  Pediatr Cardiol       Date:  2017-05-22       Impact factor: 1.655

Review 5.  [New developments in leadless pacing systems].

Authors:  Florian Doldi; Benedikt Biller; Florian Reinke; Lars Eckardt
Journal:  Herz       Date:  2021-10-22       Impact factor: 1.443

Review 6.  Pacing Without Wires: Leadless Cardiac Pacing.

Authors:  Michael L Bernard
Journal:  Ochsner J       Date:  2016

Review 7.  Innovations in Clinical Cardiac Electrophysiology: Challenges and Upcoming Solutions in 2018 and Beyond.

Authors:  Vaibhav R Vaidya; Alan Sugure; Samuel J Asirvatham
Journal:  J Innov Card Rhythm Manag       Date:  2017-12-15

8.  Artefacts in 1.5 Tesla and 3 Tesla cardiovascular magnetic resonance imaging in patients with leadless cardiac pacemakers.

Authors:  Daniel Kiblboeck; Christian Reiter; Juergen Kammler; Pierre Schmit; Hermann Blessberger; Joerg Kellermair; Franz Fellner; Clemens Steinwender
Journal:  J Cardiovasc Magn Reson       Date:  2018-07-05       Impact factor: 5.364

Review 9.  Non-traditional implantable cardioverter-defibrillator configurations and insertion techniques: a review of contemporary options.

Authors:  Johanna B Tonko; Christopher A Rinaldi
Journal:  Europace       Date:  2022-02-02       Impact factor: 5.214

10.  Long-term results of combined cardiac contractility modulation and subcutaneous defibrillator therapy in patients with heart failure and reduced ejection fraction.

Authors:  Susanne Röger; Boris Rudic; Ibrahim Akin; Tetyana Shchetynska-Marinova; Fabian Fastenrath; Erol Tülümen; Volker Liebe; Ibrahim El-Battrawy; Stefan Baumann; Jürgen Kuschyk; Martin Borggrefe
Journal:  Clin Cardiol       Date:  2018-04-26       Impact factor: 2.882

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.