Impact of New-generation Hybrid Imaging Technology on Radiation Dose during Percutaneous Coronary Interventions and Trans-femoral Aortic Valve Implantations: A comparison with conventional flat-plate angiography.

John E Boland1, Louis W Wang2, Bernard J Love1, Marino Christofi1, David W M Muller3.   

Abstract

BACKGROUND: Technological advancements in newer-generation catheterisation laboratories may reduce patient and occupational radiation exposure.
METHODS: We compared fluoroscopy time and dose-area product (DAP) between a Philips Allura X-PER FD20 and Siemens Artis Zeego Hybrid systems for 47 single-vessel percutaneous coronary interventions (PCI) and 35 transcatheter aortic valve implantations (21 Corevalve, 14 Edwards Sapien TAVI) using the FD20, versus 30 PCI and 28 TAVI (15 Corevalve, 13 Sapien) with the Zeego over a 24-month period.
RESULTS: Multivariate analysis revealed that, adjusting for patient weight and fluoroscopy time, DAP (median, interquartile range) was 26% lower for PCI with the Zeego than the FD20 [55.6 (27.0-91.5) vs 77.6 (51.2-129.1) Gy.cm(2), P=0.03)] and using tomographic imaging with the Zeego did not increase DAP for TAVI procedures [98.1 (65.9-136.6) vs 112.4 (64.9-156.2) Gy.cm(2) (P=NS). Although fluoroscopy times were longer for TAVI procedures than PCI with both systems (23.5-24.4 vs 7.3-9.2mins, p<0.0001), there was a significant difference in DAP between PCI and combined TAVI with the Zeego (55.6 vs 112.4Gy.cm(2), P<0.006) but not with the FD20 (77.6 vs 98.1Gy.cm(2), P=NS).
CONCLUSION: Specific dose-reducing features of the new-generation system reduced DAP more for PCI than TAVI, as valve replacement procedures use additional cine-acquisition not necessary for PCI.
Copyright © 2016 Australian and New Zealand Society of Cardiac and Thoracic Surgeons (ANZSCTS) and the Cardiac Society of Australia and New Zealand (CSANZ). Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Cardiac catheterisation; Coronary angiography; Percutaneous coronary intervention; Radiation; Structural heart intervention; Transcatheter aortic valve replacement

Mesh:

Year:  2016        PMID: 26906283     DOI: 10.1016/j.hlc.2015.08.023

Source DB:  PubMed          Journal:  Heart Lung Circ        ISSN: 1443-9506            Impact factor:   2.975


  4 in total

1.  Accuracy of a new intraoperative cone beam CT imaging technique (Artis zeego II) compared to postoperative CT scan for assessment of pedicle screws placement and breaches detection.

Authors:  Virginie Cordemans; Ludovic Kaminski; Xavier Banse; Bernard G Francq; Olivier Cartiaux
Journal:  Eur Spine J       Date:  2017-05-20       Impact factor: 3.134

2.  Radiation doses during cardiac catheterisation procedures in India: a multicentre study: Radiation dose study.

Authors:  Vijayakumar Subban; Sophie Amelot; Suma M Victor; Anil Potdar; Vishawanath Yadav; Tejas Patel; Sanjay Shah; Thomas Alexander; Balakumaran Jeyakumaran; Juno Angel; Mullasari S Ajit
Journal:  AsiaIntervention       Date:  2020-07-20

3.  Comparison of pedicle screw placement accuracy between two types of imaging support (Artis Zeego versus two-dimensional fluoroscopy): a cross-sectional observational study.

Authors:  Akira Matsuoka; Tomoaki Toyone; Ichiro Okano; Yoshifumi Kudo; Koji Ishikawa; Hiroshi Maruyama; Tomoyuki Ozawa; Toshiyuki Shirahata; Katsunori Inagaki
Journal:  BMC Musculoskelet Disord       Date:  2022-07-05       Impact factor: 2.562

4.  Curved periacetabular osteotomy using intraoperative real-time 3-dimensional computed tomography with a robotic C-arm system: A case report.

Authors:  Ariha Goshi; Shigeo Fukunishi; Shohei Okahisa; Taishi Okada; Shinichi Yoshiya
Journal:  Medicine (Baltimore)       Date:  2018-11       Impact factor: 1.817

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.